In accordance with the USFDA industry guidance for bioanalytical method validation,9 the ideal calibration curve should have at least seven points including a blank. The remaining six points should cover the intended concentration range and be evenly spaced along the x-axis (concentration) to avoid issues with leverage. The blank is included to more accurately assess the detection capability, limit of blank (LoB) and limit of detection (LoD), as this provides a means of determining background relating not only to electronic and chemical noise but also from potential isobaric species inherent to the tissue matrix.10 Therefore, the mimetic tissue model will consist of six levels of spiked homogenate and one non-spiked homogenate for assessment of the detection capability.
Homogenization (10 minutes)
By first intent, the homogenate of the mimetic tissue model should be made from the same tissue type as that which is being quantified. This, however, is not always going to be possible given that certain tissue types do not homogenize well (i.e. skin). Furthermore, appropriate control tissue does not always exist (i.e. tumor or xenograft tissue), and it may not always be possible to obtain control tissue from the same animal species. In these situations, we have found that rat liver tissue can be used as an effective surrogate. While the use of surrogate tissue can mimic the ion suppression and extraction from the tissue being quantified, the potential for unforeseen isobaric species in either the model or the target sample tissue should be noted. If an isobar exists in the sample tissue and not in the model tissue, the limit of detection can be underestimated and vice versa for the inverse situation. When preparing the tissue homogenate for the mimetic model, the use of a bead homogenizer without the addition of solvent is critical to yield a more morphologically relevant homogenate.8, 11
- Determine the minimum amount of homogenate that will be needed.
--1.1. The aim is to have about 300-400 mg of homogenate for each layer.
----1.1.1. Seven layers x 300 mg ≈ minimum 2 g of tissue per model.
- Prepare the tissue for homogenization.
--2.1. If performing a large-scale homogenization (i.e. whole rat liver) choose an appropriate sized tube/vial.
----2.1.1. Whole rat liver can be aliquoted into two 50 mL conical tubes.
--2.2. Cut small enough portions of the tissue that will fit into your tube of choice.
----2.2.1. Portion enough tissue to at least satisfy the minimum amount required.
--2.3. Do not overfill the tube (only fill about ¾ to leave room for beads & homogenization).
--2.4. Add stainless steel homogenizing beads to the tube (MP Metal Bead Lysing Matrix).
----2.4.1. Roughly 15 beads for a ¾ full 50 mL tube will suffice.
--2.5. Wrap the top of the tube with parafilm to prevent loss of tissue.
- Homogenize the tissue using the MP FastPrep 24 or similar bead homogenizer (critical).
--3.1. Homogenize at 4 m/s x 30 s for as many rounds as necessary (usually 2-3 rounds).
--3.2. Let the homogenate settle or spin down at low speed (1000 x g for ~ 30 s) to collect as much as possible from the tube walls.
----3.2.1. The intention of spinning down is to collect the material off the tube walls but it may result in phase separation. If using the homogenate directly, gently stir to reincorporate the mixture.
--3.3. If the tissue was large enough to be spread over several tubes, it is suggested that these homogenates be combined to avoid potential bias in sampling from the tissue.
--3.4. Aliquot the homogenate into smaller single use containers (~2 g each) to avoid multiple freeze/thaw cycles.
See figure in Figures section.Preparing the Mold (5 minutes)
The mold for the mimetic model is a modified 3 mL syringe (BD 3 mL Syringe REF 309657). Regardless of the exact type or brand of syringe, there are a few important attributes to the syringe choice. Given that the volume of the syringe ultimately dictates the amount of homogenate that is needed as well as the height of each layer of the model, the 3 mL syringe has a large enough inner diameter that can accommodate the tip for the positive displacement pipette and is small enough to limit the amount of homogenate required for each layer. Additionally, the plunger seal should be rubber-like material (avoid hard plastic) to limit the penetration of ethanol into the barrel during the freezing step.
- Trim the syringe barrel.
--1.1. Pull back or remove the plunger.
--1.2. Cut the luered tip off the syringe (can use an HPLC tubing cutter).
See figure in Figures section.- Trim the plunger seal.
--2.1. The plunger seal may have a conical shape which will cause a conical distortion in the first layer of the model.
--2.2. Cut the conical tip off the plunger.
--2.3. Re-insert the plunger into the syringe body to roughly the first volume demarcation.
--2.4. Cut the extruding end of the plunger such that the bottom of the plunger is flush with the bottom of the syringe.
----2.4.1. You can save this trimmed end of the plunger to later push the mimetic model out of the syringe.
See figure in Figures section.(Optional)
- Create a handle for dipping the mold into the dry ice-cooled ethanol.
--3.1. Drill or press a small hole through the bottom of the syringe barrel and plunger as shown in the figure below.
--3.2. Press a small diameter L-shaped hexagonal (allen) key through the hole.
--3.3. This serves to lock the plunger in place as well provides a handle to dip the mold into the chilled ethanol.
See figure in Figures section.Standard preparation and homogenate spiking (20 minutes)
To maintain the relevant morphology of the tissue, it is imperative that as little solvent as possible be added when spiking the homogenates. Ideally, the total volume of the spiked standard(s) should be kept below 2-3% (v/v) compared to the homogenate volume. This may, however, be limited by the solubility of the standard. The solvent used to dissolve the standard should be water miscible to ensure efficient mixing with the homogenized tissue. The volume of standard added to each homogenate (mimetic model layer) should be consistent. There is an attached "macro-enabled spreadsheet":http://www.nature.com/protocolexchange/system/uploads/7111/original/Mimetic_Tissue_Model_Calculations.xlsm?1534966473 which can be used to help determine the appropriate volumes and concentrations of standard to add given the weight of the tissue homogenate and the target tissue concentration. It is also possible to add multiple standards to the same mimetic model to allow for the quantification of multiple species. If the model is to contain multiple standards, their concentration gradients should be inverted such that the highest concentration of standard A should be spiked into the same homogenate as the lowest concentration of standard B. When spiking multiple standards, the total spiked volume should be kept below the 2-3% (v/v) cutoff, if possible. Ion suppression resulting from the spiked standard(s) should also be assessed for the impact on the other standard(s). It is not required to use a stable isotopically labeled (SIL) standard unless quantification is to be performed on an endogenous species that is inherently present in the tissue homogenate used for the mimetic tissue model construction. The use of an SIL in this case will allow for deconvolution of the spiked standard level from the endogenous level in the tissue homogenate.
- Determine and assign the intended range and intervals for the final tissue concentration of each level of the mimetic model.
- Determine weight of tissue homogenate for each layer (using the attached spreadsheet).
--2.1. Record the weights of seven ~2 mL screw cap vials (1 vial for each layer of the mimetic model).
--2.2. Using a positive displacement pipette, aliquot approximately 300 mg (~300 µL) of tissue homogenate into the pre-weighed vials.
--2.3. Record the weight of each vial and homogenate to determine the weight of homogenate in each vial.
- Spike each homogenate with standard(s).
--3.1. Determine and prepare an appropriate dilution of stock standard to spike into each homogenate.
----3.1.1. Standard solution volume / tissue homogenate volume should be ≤ 2-3% (v/v).
----3.1.2. Standard should be dissolved in a water miscible solvent to ensure mixing with homogenate.
----3.1.3. The volume of standard spiked into each layer should be roughly equivalent.
--3.2. Add the appropriate volume of standard to each homogenate and mix.
----3.2.1. Once the standard has been added, mix by adding 1-2 metal homogenization beads to each vial and load into the FastPrep 24 for mixing.
------3.2.1.1. Mix at 4 m/s for ~ 30 seconds.
------3.2.1.2. This mixing step is critical to ensure that the standard is evenly distributed throughout the homogenate.
--3.3. The presence of air bubbles in the homogenate will result in voids in the final tissue section which should be avoided where possible.
----3.3.1. Spin down each spiked homogenate vial at a low speed (1000 x g for 30 seconds) to remove air bubbles which may have arisen during the mixing.
------3.3.1.1. Higher speeds can result in a phase separation of the lipophilic components and should be limited or avoided if possible.
------3.3.1.2. Some foaming may be evident on the top of the homogenate after centrifugation.
See figure in Figures section.Preparing the Mimetic Tissue Model plug (15 minutes + > 3 hours to freeze, usually overnight)
The next series of steps involves the construction of the mimetic tissue model. This process is the serial freezing of each of the individual spiked homogenates into a single tissue plug which can be sectioned along with the target sample tissue. To minimize the potential for carryover, the first layer added should be the blank homogenate followed by the lowest concentration spiked homogenate and working up to the last layer which is the highest concentration spiked homogenate.
- Prepare the freezing medium.
--1.1. Set up a small beaker with dry ice-cooled ethanol deep enough to nearly submerge the mold (syringe).
--1.2. As is true with freezing other types of tissue, the speed of freezing can impact the quality of the tissue in terms of the formation of ice crystals and other artifacts. Typically, the faster the freezing the better the quality.
- Add a layer of homogenate to the mold.
--2.1. Using a positive displacement pipette, collect about 250 μL of homogenate.
----2.1.1. This is typically enough to evenly coat the plunger seal or previous layer.
----2.1.2. The exact amount added is not critical but bubbles should be avoided where possible.
--2.2. Wipe any excess homogenate off the pipette tip to avoid contaminating the edges of the mold.
--2.3. Insert the pipette tip into the mold and expel enough of the homogenate to coat the plunger (or previous layer).
----2.3.1. It is ideal to not expel all the homogenate as an air bubble is likely to form if you do.
--2.4. If necessary, tap the barrel flange side of the mold down onto a hard surface to remove air bubbles and/or help the homogenate to settle to an even layer.
--2.5. It is crucial to have the homogenate completely coat the previous layer.
- Freeze the newly added layer of homogenate.
--3.1. Dip the mold into the dry ice-chilled ethanol for at least 1 minute (or longer until the layer is frozen).
----3.1.1. Dip at least far enough such that the newly added homogenate is below the ethanol level but do not completely submerge (i.e. the open end of the mold should always be above the ethanol level).
- Repeat steps 2 and 3 for the additional spiked homogenate layers.
--4.1. To avoid the potential for contamination, layers should be added starting with the blank then the lowest concentration spiked homogenate and ending with the highest concentration spiked homogenate.
--4.2. Try to limit the amount of time between freezing sequential layers.
----4.2.1. Once the fresh homogenate layer is added, it should be frozen as quickly as possible to prevent the previous layer from melting and mixing with the new layer.
- Once the last layer has been added and frozen, remove the handle (allen key), wrap the mold in foil, and place in a -80°C freezer for at least 3 hours or overnight to ensure that it is thoroughly frozen.
See figure in Figures section.Sectioning the Mimetic Tissue Model (10 minutes)
Once the mimetic model is constructed for a given standard, applying the quantification of that standard to any tissue merely requires collecting a section of the mimetic model onto the same slide with the tissue to be quantified.
- Equilibrate the model to the cryostat temperature as you would any other tissue.
- Remove the model from the mold.
--2.1. If necessary, use the palm of your hand to slightly warm the edges of the mold.
----2.1.1. Be careful not to warm it too much or the tissue will smear as you are pushing the model out of the mold.
--2.2. Start to slowly push the model out of the mold.
----2.2.1. Use any tool to depress the plunger.
------2.2.1.1. (i.e. a pair of forceps or the remainder of the plunger that you cut when preparing the mold).
--2.3. As the model is pushed out of the mold it may stick to the plunger seal.
----2.3.1. Use the plunger to orient the tissue model for mounting to the chuck.
----2.3.2. Carefully separate the plunger from the tissue model.
See figure in Figures section.- Mount the mimetic model onto the cryostat chuck.
--3.1. Apply a strip of mounting medium to the cryostat chuck.
--3.2. Mount the mimetic model horizontally onto the chuck.
See figure in Figures section.- Sectioning the model.
--4.1. The model should be sectioned horizontally to avoid the potential for contamination between model layers.
--4.2. Trim into the mimetic model tissue until there is enough surface area to collect a section.
--4.3. Section the model at the same thickness as the tissue to be quantified.
--4.4. Thaw-mount onto the slide with the tissue to be quantified.
--4.5. Replace remainder of the mimetic model back into the -80°C freezer for storage.
See figure in Figures section.