Anatomical Considerations for Appropriate Mini-plate Positioning in Open-Door Laminoplasty To Avoid Plate Impingement and Screw Facet Violation

DOI: https://doi.org/10.21203/rs.3.rs-1026135/v1

Abstract

Objective

This study aimed to describe a safe zone for mini-plate positioning that can avoid instrument-related complications in laminoplasty.

Methods

The posterior surface length and inferior pole angle of the lateral mass were measured at each level using computed tomography. The safe zone was defined based on these measurements. Incidences of screw facet violation and plate impingement were recorded.

Results

Among 40 patients included, 15 (37.5%) had inappropriate plate positioning, causing screw facet violation or plate impingement, which more commonly occurred at distal (C5, C6) and proximal (C3, C4) levels, respectively. Lateral mass posterior surface length was shorter at the proximal levels, and the inferior pole angle of the lateral mass was smaller at the distal levels, signifying that the lateral mass became thin and long at the distal levels. Inserting the mini-plate with plate-to-lateral mass inferior pole distances of 4–5 mm and 5–6 mm at the C3–C5 and C6–C7 levels, respectively, would avoid instrument-related complications.

Conclusion

The risk of plate impingement was higher at the proximal level, whereas the risk of screw facet violation was higher at the distal level in open-door cervical laminoplasty. These risks coincide with anatomical differences at each level.

Introduction

Open-door laminoplasty is a safe and widely applied surgical technique for the treatment of cervical myelopathy.1,2 Various instruments have been used to maintain the opening of the lamina, including bone block, stay suture, wire, and mini-plate.3 Mini-plate fixation has been reported to reduce the risk of lamina reclosure while enhancing bone healing at the trough side.4,5 Fixation of the mini-plate is generally a safe procedure, and critical complications related to its instrumentation are rare since it is fixed with short screws.5,6

Although mini-plate fixation with screws is free from adverse complications such as vertebral artery injury or nerve root irritation, which could occur with lateral mass screw or pedicle screw fixation, instrument-related complications can still occur.3,7−9 Considering the anatomy of the lateral mass, mini-plates located too caudally would result in screw facet violation. However, when the plate is located excessively cranially, the inserted plate would impinge on the lateral mass at the proximal level. Previous studies have demonstrated that screw facet violation is not rare in mini-plate fixation for laminoplasty, where incidence ranges from 34.1–37.4%.3,8,9 Screw facet joint violation or plate impingement caused by the mini-plate raises concern for possible facet degeneration and postoperative neck pain.9 Aggravation of neck pain and kyphosis after laminoplasty are common, possibly caused by instrument positioning.10–12 It has also been reported that facet joint violation by mini-screws can decrease cervical range of motion.9

While previous studies have described optimal insertion points and trajectories for lateral mass screws or pedicle screws, not many have demonstrated an optimal method for mini-plate fixation for laminoplasty.3,8,13,14 Therefore, this study was conducted to (1) describe the incidence of screw facet joint violation and plate impingement in open-door laminoplasty using mini-plate fixation, (2) define a safe zone to avoid instrument-related complications, and (3) identify whether inappropriately positioned mini-plates would adversely affect clinical outcomes.

Materials And Methods

Study design and participants

This was a retrospective cohort study approved by the institutional review board of our institute (Dongguk University Ilsan Hospital Institutional Review Board 2021-09-020). All methods were carried out in accordance with relevant guidelines and regulations. Informed consent was waived owing to the study’s retrospective nature. The study was conducted in accordance with the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement for cohort studies.

The medical records of 51 patients who underwent laminoplasty for cervical myelopathy caused by spondylosis or ossification of the posterior longitudinal ligament between September 2012 and March 2019 were retrospectively reviewed. Patients (1) who underwent surgery due to trauma, infection, or tumor; (2) who lacked radiographic or clinical data; and (3) who had a follow-up period of less than 1 year were excluded.

Patients with screw facet joint violation or possible plate impingement with cranial lateral mass observed on postoperative computed tomography (CT) were classified into the inappropriate position patient group (IP group). Patients with no identifiable mini-plate-related complications were defined as the appropriate position patient group (AP group).

Surgical technique

Patients were placed in a prone position with their heads located on the Mayfield headrest. A midline posterior approach was used to expose the spinous process and lamina of the indicated levels. Dissection was performed until the lamina-lateral mass junction was exposed. The spinous processes were resected at the base. Open-side and hinge-side troughs were made at the lamina-lateral mass junction. After the lamina was carefully opened to avoid complete fracture of the hinge side, a mini-plate (Centerpiece, Medtronic, Minneapolis, MN, USA) was used to maintain the lamina opening. We attempted to position the mini-plate at the center of the posterior surface of the lateral mass. Two 5-mm screws were used to fix the plate at the lamina, and two 5-mm screws were inserted at the lateral mass to anchor the plate. Screws into the lateral mass were inserted perpendicular to the posterior surface of the lateral mass. For C7, partial laminectomy rather than open-door laminoplasty was performed to preserve muscle insertion in the spinous process.15

Variables and radiographic measurements

The neck pain visual analog scale (VAS), arm pain VAS, and neck disability index (NDI) were recorded preoperatively and at each postoperative follow-up. Three-dimensional CT scans were taken preoperatively for surgical planning and at 2 days postoperatively to evaluate adequate decompression and instrument position.16

Possible plate impingement was diagnosed when the cranial edge of the mini-plate reached the caudal edge of the lateral mass of adjacent proximal level (Figure 1A). Screw facet violation was defined as the screw penetrating the ventral surface of the lateral mass detected on axial or sagittal reconstructed CT images (Figure 1B).

The posterior surface length of the lateral mass was measured as the distance between the cranial and caudal edges of the lateral mass on sagittal CT images. A sagittal image showing a pedicle-lateral mass junction was selected because the mini-plate is usually fixed to the lateral mass at this location. Within the same sagittal image, the inferior pole angle of the lateral mass was measured as the angle between the line drawn through the posterior surface of the lateral mass and the line drawn through the ventral-inferior border of the lateral mass that composed the facet joint. Measurements were performed bilaterally, and the mean value was used for evaluation (Figure 2).

Definition of safe zone

A safe zone that can avoid both screw facet violation and plate impingement at the proximal adjacent level was defined. To define the safe zone, we first measured the size of the mini-plate (Centerpiece, Medtronic, Minneapolis, MN, USA) used for all patients included in the study. The distances between the edge of the plate and the center of the screw fixation hole, between the center of each screw fixation hole, and between the cranial and caudal edges of the plate were 2 mm, 4 mm, and 8 mm, respectively (Figure 3A).

Second, the minimum distance from the inferior pole of the lateral mass needed to avoid screw facet violation was calculated. The minimum distance for screw placement (x) was calculated using the inferior pole angle (a) and the length of the screw using the following equation: [x = length of screw ÷ tangent (a)] where tangent (a) would be the same as [length of screw ÷ x] (Figure 2).

Finally, a safe zone that could avoid both screw facet violation and plate impingement was defined as the distance between the caudal edge of the mini-plate and the inferior pole of the lateral mass. We assumed that the screw was inserted perpendicular to the posterior surface of the lateral mass. The minimum distance of the safe zone would be [distance needed to avoid screw facet violation (x) – 2] where the plate caudal edge-to-plate caudal screw hole center distance is 2 mm. The maximum distance of the safe zone would be [lateral mass posterior surface length – 8] since the plate cranial edge-to-plate caudal edge distance is 8 mm (Figure 3B).

Statistical analysis

Student’s t-test was performed to compare the patient-reported outcome measure results between the IP and AP groups. Paired t-test was performed to compare preoperative and postoperative values. All data management and statistical analyses were performed using SPSS version 21.0 software (SPSS Inc., Chicago, IL, USA). Statistical significance was set at P < 0.05.

Results

Patient characteristics and inappropriate plate positioning

Approximately 40 patients with 120 levels met the inclusion criteria and were included in the study. Screw facet violation was observed in 8 patients (20.0%) and in 8 levels (6.7%). Furthermore, plate impingement was detected in 9 patients (22.5%) and in 11 levels (9.2%). In total, 15 patients (37.5%) had inappropriate plate positioning and were classified as the IP group (age, 67.7 ± 9.8 years; male, 76.2%; follow-up, 75.0% [15/20]). The remaining 25 patients (62.5%) were classified into the AP group (age, 62.1 ± 11.6 years; male, 76.2%; follow-up, 80.6% [25/31]). Baseline patient characteristics were not significantly different between the AP and IP groups (Table 1).

Table 1

Patient characteristics

 

AP group

IP group

P value

N

25

15

 

Sex

   

1.00

Male

16 (76.2%)

8 (72.7%)

Female

5 (23.8%)

3 (27.3%)

Age

62.1±11.6

67.7±9.8

0.25

Follow-up (m)

19.0±20.9

20.2±16.7

0.63

Number of levels operated

4.1±1.1

4.7±1.1

0.22

AP, appropriately positioned; IP, inappropriately positioned; m, months
Sex was analyzed using a chi-square test
Age, follow-up period, number of levels operated were analyzed using student’s t-test

Screw facet joint violation was more frequently observed at the distal levels, including C5 and C6, whereas no screw violation was observed in C3. In contrast, plate impingement was more frequently detected at the proximal levels including C3 and C4, whereas no plate impingement was detected at the most caudal level, C6 (Table 2) (Figure 4). No plate/screw pullout or breakage occurred.

Table 2

Inappropriate plate positioning

 

Facet joint violation

Plate impingement

C3

0% (0/22)

18.2% (4/22)

C4

2.6% (1/39)

15.4% (6/39)

C5

13.3% (4/30)

3.3% (1/30)

C6

10.3% (3/29)

0% (0/29)

Radiographic results

The lateral mass posterior surface length was longer at the distal levels. Furthermore, the inferior pole angle of the lateral mass tended to decrease at these levels. Because of the smaller inferior pole angle at the distal levels, the calculated minimum distance from the inferior pole to insert screws by [x = length of screw ÷ tangent (a)] was higher at the distal levels (Table 3) (Figure 4).

Table 3

Radiographic measurements

Levels

Posterior surface length (mm)

Inferior pole angle (°)

Distance from inferior pole needed to avoid screw facet violation (5mm)

Distance from inferior pole needed to avoid screw facet violation (7mm)

C3

14.3±1.9

50.5±4.8

4.2±0.7

5.9±1.0

C4

13.3±2.1

48.0±5.4

4.6±0.9

6.5±1.2

C5

14.3±1.5

45.6±4.9

5.0±0.9

7.0±1.2

C6

15.2±1.6

38.7±4.9

6.4±1.2

9.0±1.6

C7

16.1±1.6

33.5±4.3

7.7±1.3

10.8±1.8

Distance from the in inferior pole needed to avoid screw facet violation was calculated by [length of screw ÷ tangent (inferior pole angle)]

Patient-reported outcome measures

Neck pain VAS, arm pain VAS, and NDI significantly improved after the operation in both groups (P < 0.01). No significant difference in patient-reported outcome measures between the AP and IP groups was observed at each follow-up period (Table 4).

Table 4

Patient reported outcome measures

   

AP group

IP group

P value

Neck pain

VAS

Preoperative

6.1±2.6

6.2±2.1

0.91

Post op 6w

2.2±2.1

1.7±2.0

0.46

Post op 1y

1.2±1.3

0.6±1.3

0.26

Final follow-up

1.6±1.3

2.0±3.1

0.46

Arm pain

VAS

Preoperative

7.6±1.1

8.4±0.8

0.06

Post op 6w

2.6±2.2

3.6±2.1

0.13

Post op 1y

1.5±1.6

1.6±1.8

0.94

Final follow-up

1.6±1.1

2.0±3.1

0.88

NDI

Preoperative

25.3±5.9

27.7±7.3

0.57

Post op 6w

13.4±7.0

14.8±9.6

0.92

Post op 1y

11.2±9.9

7.8±5.0

0.77

Final follow-up

11.9±10.1

14.8±7.1

0.95

AP, appropriately positioned; IP, inappropriately positioned; VAS, visual analogue scale; NDI, neck disability index
Comparisons between two groups were performed using student’s t-test

Suggested safe zone

The suggested safe zone for appropriate plate positioning in laminoplasty is summarized in Table 5 and Figure 5. When inserting a 5-mm screw for the caudal screw for mini-plate lateral mass fixation, 2- to 3-mm distancing of the plate from the inferior pole of the lateral mass was required for the C3, C4, and C5 levels. However, at the C6 or C7 levels, a distance of approximately 5 mm from the inferior pole of the lateral mass was needed to avoid screw facet violation. Distancing the mini-plate by more than 5–6 mm at C3–C5 and 7–8 mm at C6–C7 from the inferior pole of the lateral mass was not safe because it would cause plate impingement at the cranial level. The safe zone was narrower when inserting a 7-mm screw for the caudal screw as more distance is needed to avoid screw facet violation. Inserting a 7-mm screw for the caudal screw at the C7 level leaves no safe zone due to the thin lateral mass at this level, as demonstrated by the small inferior pole angle at C7.

Table 5

Suggested safe zone defined by the distance between inferior pole of lateral mass and caudal edge of mini-plate

Levels

Distance between inferior pole of lateral mass and caudal edge of plate (mm)

5mm screw for caudal screw

7mm screw for caudal screw

Minimum

Maximum

Median

Minimum

Maximum

Median

C3

2.2

6.3

4.3

3.9

6.3

5.1

C4

2.6

5.3

4.0

4.5

5.3

4.9

C5

3.0

6.3

4.7

5.0

6.3

5.7

C6

4.4

7.2

5.8

7.0

7.2

7.1

C7

5.7

8.1

6.9

     
Minimum distance was calculated by [Distance from in inferior pole needed to avoid screw facet violation - 2]
Maximum distance was calculated by [Posterior surface length -8]

Considering the median value of the safe zone, for C3–C5, leaving a 4-to 5-mm distance when inserting a 5-mm screw and distancing the mini-plate 5–6 mm when inserting a 7-mm screw from the inferior pole would avoid both screw facet violation and plate impingement. For C6–C7, leaving 6–7 mm when inserting a 5-mm screw would avoid instrument-related complications. However, for C6–C7, 7-mm screw insertion for caudal screw in mini-plate fixation would be considered unsafe because of the narrow or no safe zone.

Discussion

Optimal mini-plate insertion for cervical laminoplasty would adequately prevent the reclosure of an open hinge without instrument-related complications such as screw pullout, plate breakage, screw facet violation, and plate impingement with approximate level.3,5,8,17 Several studies have demonstrated that aggravation of kyphosis, decreased range of motion, and postoperative neck pain are common after laminoplasty.10–12 Although injuries to the posterior neck musculature have been commonly discussed as a factor causing these adverse outcomes, inappropriate instrument positioning such as plate impingement or screw facet violation would also have a negative effect on axial symptoms and may have been underestimated.3,8,9 While transfacet fixation has been reported as a viable technique for fusion, screw facet violation by mini-screws does not limit facet joint motion and would accelerate the degenerative process at the involved level.18 Chen et al. demonstrated that screw facet joint violation during mini-plate fixation results in decreased range of motion and aggravated neck pain, although neurological recovery was not affected.9

Two previous studies have suggested safe mini-screw insertion points for laminoplasty.3,8 Chen et al. demonstrated the safe zone using 3D image rendering.8 However, the safe zone definition in this study is complex considering that intraoperatively, surgeons can only adjust the plate position in the cranial or caudal direction. Min et al. also demonstrated the minimal safe distance of mini-screw insertion.3 The limitation of this study is that the measuring method has not been objectively described. Furthermore, although these two studies suggest that a certain distance is needed from the inferior pole of the lateral mass to avoid screw facet violation, they did not consider the possibility of plate impingement when the plate is located too cranially.3,8 Therefore, the present study attempted to define the safe zone of mini-plate placement by considering both minimum (to avoid screw facet violation) and maximum distances (to avoid plate impingement) from the inferior pole of the lateral mass.

In this study, screw facet violation was more common at the distal levels, including C5 and C6, whereas plate impingement was more common at the proximal levels, such as C3 and C4. The results of radiological measurements demonstrate that this trend is consistent with the anatomical differences between each level. The possibility of plate impingement would be higher at the proximal level because the posterior surface length is shorter at these levels. However, a smaller inferior pole angle at the distal level signifies a thin lateral mass at these levels, which increases the possibility of screw facet joint violation. Considering such anatomical differences at each level, locating the mini-plate more caudally at the proximal levels and more cranially at the distal levels would help avoid instrument-related complications.

The safe zone was described based on the distance between the inferior pole of the lateral mass and the caudal edge of the mini-plate. Although previous reports have used the screw insertion area as the reference point, we used the caudal edge of the mini-plate because it is easier to identify intraoperatively. The suggested safe zone demonstrates that inserting the mini-plate with a plate-to-lateral mass inferior pole distance of 4–5 mm for the C3–C5 levels and 5–6 mm for the C6–C7 levels would avoid instrument-related complications. Min et al. also demonstrated that more distance from the inferior border of the lateral mass is needed at distal levels to avoid screw facet joint violation.3 It is known that the lateral mass is generally thin at C7, which makes pedicle screw a more preferred choice than lateral mass screw.19,20 The results of the present study also demonstrated that inserting a 7-mm screw for mini-plate fixation in C6–C7 would be unsafe owing to the thin lateral mass at these levels as demonstrated by the small inferior pole angle. This finding supports performing partial laminectomy rather than laminoplasty at C7 due to a higher chance of instrument-related complications at this level.20,21

Laminoplasty is often performed in patients with cervical spondylosis, which distorts the anatomical landmarks due to bony spurs and spondylolisthesis. Lee et al. demonstrated that screw facet joint violation is more common in severely degenerative cervical spine than in mildly degenerative spine.7 Although the suggested safe zone in the present study could be used as a reference while placing the mini-plate, such distortion of anatomic landmarks would make it difficult to identify appropriate insertion points. Therefore, individual assessment and preoperative planning with radiographic measurements used in this study would further enhance the safety of laminoplasty.

The results of the present study demonstrated that despite the occurrence of screw facet joint violation or plate impingement, clinical results such as neck pain VAS or NDI were not adversely affected, which is contradictory to the findings of Chen et al., which suggested that screw facet joint violation is related to increased neck pain and decreased range of motion.9 However, both studies included a small number of patients with instrument-related complications, which warrants further evaluation.

This study had several limitations. First, the lateral mass posterior surface is not a plane surface, but rather has a round curvature, and the measuring method of the present study would have limitations reflecting such curved surfaces. However, within the confinement of using two-dimensional images, the round curvature of the lateral mass cannot be completely measured. Furthermore, the minimal safety distance to avoid screw facet violation demonstrated in this study corresponds to that reported in other previous reports.3,8 Second, as previously discussed, the study has limited capacity to demonstrate the clinical impact of inappropriate instrument positioning due to the small sample size. Finally, the study is not free from the possibility of selection bias because it was a retrospective, single-center study.

In conclusion, the risk of plate impingement was higher at the proximal level, whereas the risk of screw facet violation was higher at the distal level in open-door cervical laminoplasty. These risks coincide with the anatomical differences at each level. The demonstrated safe zone can be used as a reference for plate positioning. Despite inappropriate positioning of the mini-plate, the clinical outcomes were not adversely affected.

Declarations

Data availability

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Acknowledgements

This work was supported by the research program of Dongguk University.

Author contributions

Conceptualization and design of the work was done by JJY. Analysis, data collection, writing, and editing was done by SHP.

Competing interests

The authors declare no competing interests.

References

  1. Choi, S. H. & Kang, C. N. Degenerative Cervical Myelopathy: Pathophysiology and Current Treatment Strategies. Asian Spine J 14, 710-720, doi:10.31616/asj.2020.0490 (2020).
  2. Hirabayashi, K., Miyakawa, J., Satomi, K., Maruyama, T. & Wakano, K. Operative results and postoperative progression of ossification among patients with ossification of cervical posterior longitudinal ligament. Spine (Phila Pa 1976) 6, 354-364, doi:10.1097/00007632-198107000-00005 (1981).
  3. Min, W. K., Seo, I., Na, S. B., Choi, Y. S. & Choi, J. Y. Radiological analysis of minimal safe distance and optimal screw angle to avoid facet violation in open-door laminoplasty using precontoured plate. J Orthop Surg (Hong Kong) 25, 2309499017736562, doi:10.1177/2309499017736562 (2017).
  4. Matsumoto, M. et al. Impact of lamina closure on long-term outcomes of open-door laminoplasty in patients with cervical myelopathy: minimum 5-year follow-up study. Spine (Phila Pa 1976) 37, 1288-1291, doi:10.1097/BRS.0b013e3182498434 (2012).
  5. Rhee, J. M., Register, B., Hamasaki, T. & Franklin, B. Plate-only open door laminoplasty maintains stable spinal canal expansion with high rates of hinge union and no plate failures. Spine (Phila Pa 1976) 36, 9-14, doi:10.1097/BRS.0b013e3181fea49c (2011).
  6. Chen, H. et al. Effect of Mini-plate Fixation on Hinge Fracture and Bony Fusion in Unilateral Open-door Cervical Expansive Laminoplasty. Clin Spine Surg 29, E288-295, doi:10.1097/BSD.0000000000000131 (2016).
  7. Lee, D. H. et al. A CT-Based Simulation Study to Compare the Risk of Facet Joint Violation by the Cervical Pedicle Screw Between Degenerative and Nondegenerative Cervical Spines. Spine (Phila Pa 1976) 42, E136-E141, doi:10.1097/BRS.0000000000001730 (2017).
  8. Chen, H. et al. Optimal area of lateral mass mini-screws implanted in plated cervical laminoplasty: a radiography anatomy study. Eur Spine J 26, 1140-1148, doi:10.1007/s00586-016-4785-9 (2017).
  9. Chen, H. et al. Facet joint disturbance induced by miniscrews in plated cervical laminoplasty: Dose it influence the clinical and radiologic outcomes? Medicine (Baltimore) 95, e4666, doi:10.1097/MD.0000000000004666 (2016).
  10. Kim, B. J., Cho, S. M., Hur, J. W., Cha, J. & Kim, S. H. Kinematics after cervical laminoplasty: risk factors for cervical kyphotic deformity after laminoplasty. Spine J, doi:10.1016/j.spinee.2021.06.010 (2021).
  11. Niu, S., Anastasio, A. T. & Rhee, J. M. Laminoplasty Achieves Improved Outcomes Despite Leading to a More Positive Sagittal Balance: Neither Preoperative Nor Postoperative Sagittal Balance Correlated With Spine-specific Outcome Data. Clin Spine Surg, doi:10.1097/BSD.0000000000001165 (2021).
  12. Wang, S. J., Jiang, S. D., Jiang, L. S. & Dai, L. Y. Axial pain after posterior cervical spine surgery: a systematic review. Eur Spine J 20, 185-194, doi:10.1007/s00586-010-1600-x (2011).
  13. Lee, C. S., Chung, S. S., Kang, K. C., Park, S. J. & Shin, S. K. Normal patterns of sagittal alignment of the spine in young adults radiological analysis in a Korean population. Spine (Phila Pa 1976) 36, E1648-1654, doi:10.1097/BRS.0b013e318216b0fd (2011).
  14. Lee, S. H. et al. Cervical pedicle screw placement using the "key slot technique": the feasibility and learning curve. J Spinal Disord Tech 25, 415-421, doi:10.1097/BSD.0b013e3182309657 (2012).
  15. Umeda, M. et al. A less-invasive cervical laminoplasty for spondylotic myelopathy that preserves the semispinalis cervicis muscles and nuchal ligament. J Neurosurg Spine 18, 545-552, doi:10.3171/2013.2.SPINE12468 (2013).
  16. Kim, G. U., Chang, M. C., Kim, T. U. & Lee, G. W. Diagnostic Modality in Spine Disease: A Review. Asian Spine J 14, 910-920, doi:10.31616/asj.2020.0593 (2020).
  17. Park, A. E. & Heller, J. G. Cervical laminoplasty: use of a novel titanium plate to maintain canal expansion--surgical technique. J Spinal Disord Tech 17, 265-271, doi:10.1097/01.bsd.0000095401.27687.c0 (2004).
  18. Husain, A. et al. A comparison of open versus percutaneous cervical transfacet fixation. J Neurosurg Spine 25, 430-435, doi:10.3171/2016.2.SPINE151334 (2016).
  19. Lee, G. W. et al. Feasibility Study of Free-Hand Technique for Pedicle Screw Insertion at C7 without Fluoroscopy-Guidance. Asian Spine J 10, 38-45, doi:10.4184/asj.2016.10.1.38 (2016).
  20. Yoshihara, H., Passias, P. G. & Errico, T. J. Screw-related complications in the subaxial cervical spine with the use of lateral mass versus cervical pedicle screws: a systematic review. J Neurosurg Spine 19, 614-623, doi:10.3171/2013.8.SPINE13136 (2013).
  21. Zhang, P. et al. Preserving the C7 spinous process in laminectomy combined with lateral mass screw to prevent axial symptom. J Orthop Sci 16, 492-497, doi:10.1007/s00776-011-0115-6 (2011).