Performance of the new SmartCardia wireless, wearable oximeter: a comparison with arterial SaO2 in healthy volunteers.

DOI: https://doi.org/10.21203/rs.3.rs-1202029/v1

Abstract

Backgroud: Our goal was to evaluate the performance of a new wearable arm located pulse oximeter.

Methods: Twelve volunteers were monitored with three pulse oximeters and underwent desaturation to 70% SaO2 .  

We compared the accuracy of SpO 2 reading from the SmartCardia system with SaO2 using two well established devices (Masimo and Nellcor ) as reference.

Oximetry was performed at different level of oxygen saturation varying from 70 to 100%. Bias, ARMS and precision were evaluated using Bland-Altman plots.

Results: The mean (SD) differences between SpO compared to SaO2 and the devices were as follows: SaO2 versus Masimo 2,12±1,01% (95% CI 1,45 to 2,79), SaO2 versus Nellcor 0,78±0,58 % (95 % CI -0,29 to 1,65) and SaO2 versus SmartCardia 0,42±0,24 % (95% CI -0,64 to 1,46). The bias between SmartCardia, Masimo, Nellcor devices and SaO2 was 0.16 (95% CI 0.05 to 0.33) and LoA (level of agreements) 2.96 (95% CI -2,68 to 2,89) for SmartCardia, 2,02 (95% CI 1,49 to 2,54) and LoA -6 to 11 for Masimo, and 0,76 (95% CI 0,5 to -1) and LoA -3,5 to 5,0 for Nellcor. ARMS for the 70-100% SaO 2 range was 1,4 for SmartCardia, 5,0 for Masimo and 2,31 for Nellcor.

Conclusions: The new wireless SmartCardia SpO2 measurement system demonstrated in-line results, bias, ARMS and precision in healthy volunteers, when compared with the gold standard SaO2 and with two well established systems, Masimo and Nellcor.

Trial registration: The present trial was prospectively registered at UCSF record (registration number:10-00437), on March 8, 2021.

Introduction

Oximetry is a procedure for measuring the level of oxygen in the blood expressed as oxygen saturation (13). This parameter is widely used in the evaluation of various medical conditions that affect the function of the heart and lungs and is a standard monitoring modality during anesthesia delivery (46).

Conventional oximeters use either the reflective or the transmitive method to measure SpO2. The reflective method is known to be less accurate. For this reason, most commercially available patches do not provide SpO2 measurements. The transmissive method is the more commonly used of the two. Transmitive technology transmits red and infrared light through the finger to a photo detector. The pulse oximeter sensor reads the transmitted light beams to estimate oxygen saturation of the blood and pulse rate. This method positions the transmitter and receiver in the same plane and the sensors can be placed on areas of the anatomy other than the finger. Conventional pulse oximeters have various limitations including false alarm and failed measurements (711) which may be related to positioning of the device located on the finger. A new generation of reflective pulse oximetry device (SmartCardia oximeter) has been designed and works on a wireless basis (12, 13). The present trial was designed to validate this technology and to demonstrate the potential of this device using the reflective method. We compared this new device to the Nellcor N-600 (Covidien, Boulder, CO, USA) and the Masimo Radical device (Masimo Corp., Irvine, CA, USA), both units considered to be the two most used and accurate FDA approved non-invasive devices for SpO2 measurements. SaO2 was used as the gold standard.

Methods

Ethical considerations

Even if the device was developed in Switzerland the trial was conducted in a reference laboratory accepted by the FDA, the Hypoxia Research Laboratory at UCSF. The study was reviewed and approved by the UCSF Ethical Committee on Human Research. The approval letter is on file at UCSF. The laboratory conforms to Good Clinical Practice Standards for the involvement of human subjects and handling of test data. Written informed consent was obtained from each participant.

System description:

The SmartCardia 7L device is a wireless patch with a low-cost disposable component and a re-usable electronic unit (Figure 1). The patch acquires ECG and measures heart rate (HR) and SpO2. The data are transmitted by Bluetooth to a mobile phone or router. The measured signals and parameters are also stored on the device. The device for SpO2 recording is placed on the left arm of the subject (Figure 1). The size of the patch is 55 x 130mm. The patch-based device offers up to 14-days monitoring and data storage and 7-days real-time connectivity with the cloud through a smartphone. The ability to receive, store, and interpret a broad range of signals offers the opportunity to go far beyond monitoring individual parameters. If the patient’s vital measurements reach a pathological value, the system gives an alert on the cloud and the physician can see the real-time parameters and ECG signals.

Protocol:

The protocol required brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained. The blood sample was analyzed for oxygen saturation with a gold standard bench CO-oximeter, currently a Radiometer ABL-90 multi-wavelength oximeter (Hemoximeter, Radiometer, Copenhagen (Denmark), serial 1393-090R0359N0002). This instrument contains factory certified calibration standards and quality control algorithms. Twenty to twenty-five arterial blood samples from each subject can be analyzed following a protocol aligned with current ISO and FDA guidance documents for pulse oximeter testing.

A radial arterial cannula was placed in either the left or right wrist of each subject for arterial blood sampling and blood pressure monitoring.

Our approach to obtaining stable, safe, and controlled hypoxia was breath-by-breath respiratory gas analysis and a computer program that permits the inspired gas mixture to be adjusted to achieve a level of lung alveolar gas that will achieve the desired degree of hypoxia. Typically, saturation is determined once with air breathing and then at each of the following levels, e.g., 93%, 90%, 87%, 85%, 82%, 80%, 77%, 75% and 70% saturation for about 30-60 seconds at each level. An arterial blood sample is obtained from the catheter at the end of each hypoxic plateau. The operator changes the inspired oxygen concentration at the end of each blood sampling to attain the next desired steady-state hypoxic condition. A run takes 10-15 minutes, and each run is terminated by a breath of 100% O2 followed by room air. Two runs together enable obtaining a total of 20-25 blood samples, two samples at each different plateau. Saturation of each arterial blood sample is determined by direct oximetry using the Radiometer ABL-90 multi- wavelength oximeter.

Sites for affixing pulse oximeter probes was the fingers for the Masimo device, the ears for the Nellcor, and the arm for the SmartCardia.

Statistics:

Pulse oximeter data was taken as 5 second averages corresponding to the point of SaO2 analysis. Individual data points may be missed or excluded for dropped signals or failure of the oximeter signal to achieve an appropriate plateau. Agreement in SpO2 and arterial SaO2 were analyzed using Bland Altman analysis (15). Bland Altman curve gives a graphical representation of the agreement between SpO2 and arterial SaO2 value. The average of the SpO2 and SaO2 is plotted on the x-axis while the difference between the two values is plotted on the y-axis. The more the graph points toward a zero difference with narrow dispersion of the ‘body’, the better the agreement.

Tables of mean, standard deviation, standard error, minimum, maximum, 95% CI, count and root mean square are provided for each oximeter’s bias, and all oximeters combined in the following ranges of SaO2: 70 - 100%, 60 - 70%, 70 - 80%, 80 -90%, and 90 - 100%. LoA was obtained between methods of measurement with multiple observations per individual (16). Root mean square error (ARMS) was calculated as it represents the best way to compare SaO2 measurements. It includes both the values of SaO2 and its stability. The following formula was used for calculation:


Data were managed on MS excel spreadsheet and analysed using stata 9.0 software (Stata Corp, College Station, TX). On Bland Altman curves linear regression is shown for all subjects combined and the equation with R2 is shown on the plot. Mean bias is displayed as a solid horizontal line, and the upper and lower limits of agreement (mean bias ± 1.96•SD*) are shown by dashed horizontal lines. For “pooled” plots, different markers are used for each pulse oximeter. Continuous variables were compared using ‘t’ test. A p value < 0.05 was considered as significant.

Results

The current study was performed on 4/13/2021 and 4/14/2021 at UCSF. Twelve volunteers with a mean age of 28 (range 21-29) (3 women and 9 men) were included in the trial (Table 1). All subjects enrolled in the study had normal Hb level (Hb ≥ 10 g/l) and all were healthy and non-smoking individuals.

Table 1

Mean±SD SpO2 values at different plateaus with the 3 oximeters for the 12 subjects. Mean differences between the 3 oximeters are shown as delta. H=hemoximeter, Ma=the Masimo system, Ne=the Nellcor system, SC=the SmartCardia system. Delta is the mean difference between SaO2 and the tested system.

Oxymeter

H

SaO2 in %

SC

SpO2 in %

Ma

SpO2 in %

Ne

SpO2 in %

Delta H/SC

Delta H/Ma

Delta H/Ne

Baseline

97,83±0,43

97,85±0,85

98±2,92

97,9±1,2

0,02

0,17

0,07

Plateau 1

94,81±1,33

94,58±1,39

96,75±2,38

94,83±1,45

-0,23

1,94

0,02

Plateau 2

91,25±2,48

90,78±3,45

93,75±3,11

91,54±3,02

-0,47

3,5

0,29

Plateau 3

87,14±3,04

87,09±3,32

89,29±3,99

87,25±3,84

-0,05

2,15

0,11

Plateau 4

82,96±3,25

83,16±3,43

86,04±4,62

84±4,75

0,20

3,08

1,04

Plateau 5

76,90±3,5

77,49±3,67

79,5±5,95

78,5±4,93

0,59

2,6

1,6

Plateau 6

72,22±3,42

73,03±3,56

74,64±5,86

72,79±4

0,81

2,42

0,63

Break

99,28±0,31

98,75±0,59

98,791,28

98,75±1,23

-0,53

-0,49

-0,53

Plateau 7

92,73±2,81

92,49±2,53

95,08±2,84

93,15±2,29

-0,24

2,35

0,42

Plateau 8

87,65±2,28

87,81±2,36

90,5±3,32

88,42±2,02

0,16

2,85

-0,23

Plateau 9

82,43±2,77

82,82±3,2

85,29±5,09

84,05±2,44

0,39

2,86

1,62

Plateau 10

76,26±2,36

76,72±2,21

78,71±6,1

77,95±2,33

0,46

2,45

1,69

Plateau 11

71,71±2,59

73,06±2,16

73,29±7,92

73,45±3,86

1,35

1,58

1,84

Mean

       

0,42±0,24

2,12±1,01

0,78±0,58

A total of 298 samples were obtained at the saturation plateaus during the study. Data were collected for each type of oximeter and probes. Mean (SD) differences measurements between SaO2 and SpO2 and between the devices tested were as follow: SaO2 versus Masimo 2,12±1,01% (95% confidence interval CI, 1,45 – 2,79), SaO2 versus Nellcor 0,78±0,58% (CI, -0,29 - 1,65) and SaO2 versus SmartCardia 0,42±0,24% (95% CI, -0,64 - 1,46). A statistically significant difference in precision was observed between the SmartCardia and the Masimo device (p<0.001), between the SmartCardia and the Nellcor device (p=0.03) and between the Nellcor and the Masimo (p<0.001) (Table 1 and figure 2).

Bias, precision and ARMS are presented in Table 2. Plots assessing the agreement between arterial SaO2 and SpO2 detected by the SmartCardia, the Masimo and the Nellcor pulse oximeters are shown in Figure 2. The bias between SmartCardia, Masimo, Nellcor and SaO2 was as follows: 0,16 (95% CI 0.05 to 0,33) and the LoA were 2,96 (95% CI -2,68 to 2,89) for SmartCardia, 2,02 (95% CI 1,49 to 2,54) and the LoA were -6 to 11 for Masimo, 0,76 (95% CI 0,5 to -1) and the LoA were -3,5 to 5,0 for Nellcor. This bias is significantly lower for SmartCardia compared to Masimo (p<0.001) and Nellcor (p<0.05), and for Nellcor compared with Masimo (p<0.05). ARMS for the 70-100% SaO2 range was 1,4 for the SmartCardia device, 5,0 for the Masimo device and 2,31 for the Nellcor device (Table 2 and figure 3).

Discussion

Currently, arterial oxygenation is primarily measured with pulse oximeters providing immediate continuous noninvasive monitoring of SpO2. Several devices are available in clinics with some limitations including false alarms and failed measurements when used in patients with low perfusion or during patient motion. Fingertip positioning makes measurements sometimes difficult and inconvenient for the patients. Some trials have shown variable sensitivity and precision in patients with hypotension and hypoxemia (15, 16), mainly due to the location of the device. We compared the performance of two well established pulse oximeters with a new generation pulse oximeter, the SmartCardia device. This wireless system is located on the patient's arm and the values are transmitted electronically and stored on the cloud, both representing major advantages compared to other devices. In addition, today, due to the Covid 19 pandemic, SpO2 measurement, even outside the hospital, has become more and more important. We have shown that despite using the typically less accurate reflective method, the SmartCardia device provides very accurate and precise results when compared with well established devices in clinical use. Limitations due to the presence of wired connections are avoided and patient hand movement is not limited by the presence of a finger mounted sensor. In addition, the oximeter can be worn during walking with permanent transmission of data to a control station or a phone providing reliable electronic wifi transmission. This allows access to permanent and uninterrupted information regarding patient oxygenation levels. To confirm the quality of the device, which is currently awaiting FDA approval, its performance was compared with the Masimo and Nellcor pulse oximeter devices which utilize the transductive method. In our study, the results are encouraging and were shown to be equivalent or even slightly better based on bias, ARMS and precision measurements. We also showed that this system provides good SpO2 assessments over the range of SaO2 tested between 70 to 100% oxygen saturation.

Our trial involved young healthy volunteer subjects in a reference laboratory setting. This allowed the use of a more rigorous protocol while maintaining control over variables.

As mentioned before, the fact that this system is placed on the subject’s arm is an advantage due to fewer disturbances related to the movement of the hand. Devices placed on the ear are also more prone to disturbances related to movements of the patient's head. We found significant performance differences among the 3 devices tested. Our results also showed that the Masimo device tended towards 2 to 3% higher SpO2 values than actual SaO2. This may be clinically relevant if confirmed in ICU patients. As our trial was not performed in a clinical setting, we cannot conclude that one oximeter would be better than the others in daily practice. Further trials in real life situations, such as ICU environments, hypotensive or hypothermic conditions, etc. are needed to further draw definitive conclusions, as it has been our experience that even well simulated clinical scenarios often lead to ambiguous results. We observed a significant difference in the measurement of SpO2 between the Nellcor and the Masimo favouring the Nellcor system. However, our data were collected in healthy volunteers, and previous trials comparing the Masimo, the Nellcor and the Philips Intellivue have not shown significant differences between these three pulse oximeters when tested in ICU patients (1617). Conversely, other trials conclude to a higher sensitivity and specificity of the Masimo device when compared to the Nellcor system (18). This confirms the need for additional studies utilizing real life environments and patient’s conditions to further corroborate our results.

Study Limitation

The major limitation is that we performed this trial in a laboratory. This may not fully replicate the characteristics of complex clinical settings exhibited by patient populations in the ICU or operating theater. In addition, the trial was conducted with a small number of patients. These results need confirmation in a larger cohort of subjects in a clinical environment.

Conclusion

The new wireless SmartCardia SpO2 measurement system demonstrated in-line results, bias, ARMS and precision in healthy volunteers, when compared with the gold standard SaO2 and with two well established systems, Masimo and Nellcor.

Abbreviations

Hb

hemoglobin blood level

UCSF

University of California, San Francisco

CI

confidence interval

ICU

intensive car unit

SpO2

saturation measured by the tested oximetry device

SaO2

saturation measured by the Hemoximeter

LoA

limits of agreements

ARMS

root mean square error

Declarations

Ethical approval and consent to participate: The study was reviewed and approved by the UCSF Ethical Committee on Human Research. The approval letter is on file at UCSF. The laboratory conforms to Good Clinical Practice Standards for the involvement of human subjects and handling of test data. Written informed consent was obtained from each participant.

Consent for publication: Publication of the data was accepted by each participant and was part of the informed written consent.

Availability of data and material: Data from the trail are stored on the UCSF files and can be consulted at any time.

Funding: The only funding body was SmartCardia. SmartCardia didn't play any role in the design of the study, as the study was designed taking into account only FDA requirements for SpO2 validation. SmartCardia collected data from its device only, and the lab where the study took place collected the gold standard data only. After the study, SmartCardia shared all data with the lab in a completely blind manner (SmartCardia never had access to the gold standard data), and the lab performed the analysis and interpretation of the data.

Competing interest: F. Rincon and S. Murali are developers of the SmartCardia oximeter and member of the board of SmartCardia. 

Authors’ contribution: All authors have read and approved the manuscript. F.Rincon and S.Murali collected the data. Redaction of the manuscript was done by JJ. Goy and statistics performed by J. Pidoux. 

Acknowledgements: not applicable.

References

  1. Jubran A. Pulse oximetry. Intensive Care Med 2004; 30:2017. 
  2. Neff TA. Routine oximetry. A fifth vital sign? Chest 1988; 94:227.
  3. Chan ED, Chan MM, Chan MM. Pulse oximetry: understanding its basic principles facilitates appreciation of its limitations. Respir Med 2013; 107:789.
  4. Eichhorn JH, Cooper JB, Cullen DJ, et al. Standards for patient monitoring during anesthesia at Harvard Medical School. JAMA 1986; 256:1017-1020.
  5. 5.International Task Force on Anaesthesia Safety. Eur J Anaesthesiol Suppl 1993; 10:12-15.
  6. 6.World Federation of Societies of Anaesthesiologists. 2008 International Standard for a Safe Practice of Anaesthesia. Available at: practice-of-anaesthesia.html. Accessed on September 8, 2008.
  7. Lawson D, Norley I, Korbon G, Loeb R, Ellis J. Blood flow limits and pulse oximeter signal detection. Anesthesiology 1987; 67:599-603.
  8. Severinghaus JW, Spellman MJ Jr. Pulse oximeter failure thresholds in hypotension and vasoconstriction. Anesthesiology 1990; 73:532-7.
  9. Wilson S. Conscious sedation and pulse oximetry: false alarms? Pediatr Dent 1990; 12:228-32.
  10. Plummer JL, Zakaria AZ, Ilsley AH, Fronsko RR, Owen H. Evaluation of the influence of movement on saturation readings from pulse oximeters. Anaesthesia 1995; 50:423-6.
  11. Wiklund L, Hök B, Ståhl K, Jordeby-Jönsson A. Post anesthesia monitoring revisited: incidence of true and false alarms from different monitoring devices. J Clin Anesth 1994; 6:182-8.
  12. S. Murali, F. Rincon, T. Cassina, S. Cook, JJ. Goy. Heart Rate and Oxygen Saturation Monitoring with a New Wearable Wireless Device in the Intensive Care Unit: Pilot Comparison Trial. JMIR Biomed Eng 2020;5(1): e18158.
  13. Murali S, Brugger N, Rincon F, Mashru M, Cook S, Goy JJ. Cardiac Ambulatory Monitoring: New Wireless Device Validated Against Conventional Holter Monitoring in a Case Series. Front. Cardiovasc. Med. 2020. 7:587945. doi: 10.3389/fcvm.
  14. Bland JM, Altman DG. Agreement Between Methods of Measurement with Multiple Observations Per Individual. J Biopharm Stat 2007; 17:571-582.
  15. Bland, J. M., & Altman, D. G. Measuring agreement in method comparison studies. Statistical methods in medical research. 1999; 8: 135-160.
  16. Ahrens T, Ott K. Comparison of three new generation pulse oximeters in a medical intensive care unit. Crit. Care Med: 2006;34: A 85.
  17. Shah N, Ragaswamy H, Govindugari K, Estanol L. Performance of the three new-generation pulse oximeters during motion and low perfusion in volunteers. J Clin Anesth. 2012; 24:385-391.
  18. Bipin J, Rakesh L, Kabra S. Comparison of two new generation pulse oximeters with arterial oxygen saturation in critically ill children. The Indian Journal of Pediatrics. 2014; 81:1297-1301.