Since in vivo dosimetry verification should an essential part of the standard QA procedure recommended by the International Atomic Energy Agency (IAEA) [23], numerous reports of EPID-based in vivo dose verification have appeared, either reporting experiences of in vivo use [8, 9, 14] or discussing the sensitivity of error detection [18, 19, 24, 25]. However, in those studies, the simulation errors were constant values that did not account for gantry inertial and gravitational forces during gantry rotation in VMAT. To the best of our knowledge, this is the first study to investigate the sensitivity of the iViewDose dosimetry system in lung SBRT for dynamic errors caused by gravity. This study also discussed other errors that may occur during delivery that are not affected by gantry rotation. The results of the phantom measurements provided us with sensitivity to possible errors for various iViewDose system scenarios.
In the pre-treatment QA procedure, gantry angles and dose outputs are usually checked daily at static angles such as 0° or 180° rather than dynamic angles. Possible errors in the output, gantry angle and MLC positions during VMAT delivery must be represented by periodic functions. Although there are several types of periodic functions that can be used, the result should not change depending on the choice of function. A sinusoidal function was used in this study to simulate the output, gantry angle and MLC position errors.
The results presented here are comparable with those of previous publications. EPID-based in vivo dosimetry did not detect 1-5% dynamic MU errors in this study; only the DRP showed an increasing trend as the error amplitude increased, with a maximum of 2.6% for the 5% dynamic MU error. Mijnheer et al. investigated the detectability of 5% MU errors in four planning models, including lung cases [18]. 5% MU more could be detected by \({\gamma }_{1\%}\), MU errors of less than 5% could not be detected. In the sinusoidal form, the MU errors decreased from 0° to 180° and increased from 180° and 360°, which may counteract the error results of the full arc. Additionally, the error amplitude may be too small to detect because the sinusoidal function reduces the magnitude of errors.
As shown in Fig. 1, the gantry angle dynamic errors were not detected for any of the \({\gamma }_{mean}\), \({\gamma }_{1\%}\), \({\gamma }_{pass rate}\) or DRP alerts, and the values were close to the original plan. A similar observation was made by Mijnheer et al. for lung cases [18]. Theoretically, gantry angle errors are serious errors; however, sometimes they are not severe enough to have a serious impact on treatment quality. For instance, the modulation is not overly strong for gantry angle errors, or the anatomy does not change considerably over angle errors. Both scenarios would yield the same primary dose transmission and could escape detection in EPID-based transit dosimetry. The model used in this study was a lung case, the body contour was regular and the target was a small sphere, indicating that even 5° gantry angle errors did not have a strong influence. Furthermore, except for 90° and 270°, the sinusoidal function used to simulate gantry errors would reduce the error amplitudes when compared with the constant errors. This would further reduce the errors in the results.
The dynamic MLC errors were quite successfully captured by the iViewDose dosimetry system for lung SBRT plans. \({\gamma }_{mean}\) and \({\gamma }_{pass rate}\) could distinguish all the dynamic MLC error scenarios, except for the 1 mm error case. \({\gamma }_{1\%}\) exceeded the alert threshold only for the 5 mm error scenario, and none of the DRP results exceeded the alert criterion. Oliver et al. showed that shifting MLC banks in opposite directions resulted in smaller or larger field shapes that tended to be more impactful than shifting systematic leaf banks in the same directions or random MLC errors [26]. In this study, dynamic MLC errors were caused by shifting all leaf banks toward the gravitational direction without changing the gap between them. The MLC apertures were not fundamentally alerted, which could make sense for none of the DRP results exceeded the alert threshold.
In contrast to previous studies [18, 19], one leaf pair situated at the position of the isocenter position errors was also designed in our study. The error amplitudes were constant and shifted, resulting in larger field sizes. \({\gamma }_{mean}\) and \({\gamma }_{1\text{\%}}\) could distinguish all MLC error scenarios with the exception of the 2 mm error case, and for the indicator \({\gamma }_{pass rate}\) and the DRP, all the errors could be captured by EPID-based transit dosimetry. This was consistent with the results of Mijnheer et al. [18]. In that study, single and multiple leaf position errors in lung cases were detected and were much more sensitive than other cases. Small changes in the MLC could be captured due to the smaller average field size.
Incorrect collimator angles for all scenarios escaped from the test. Mijnheer et al. [18] reported that collimator angle errors in lung plans could be detected by the indicator \({\gamma }_{1\text{\%}}\) when the angle was changed from 20° to 340° or vice versa; however, the DRP did not have a considerable change. In contrast, in the Yedekci et al. [19] study, all incorrect collimator angles from 0.5 to 5° escaped detection in the prostate SBRT plan, which was consistent with our findings. Although the tumor site differed from the lung plan, the target shapes were similar. The targets were regular, and rotation symmetry might be the main reason for the EPID-based transit dosimetry detection failure.
The Y1 jaw moving 5 mm inwards could be detected by the indicators \({\gamma }_{mean}\) and \({\gamma }_{pass rate}\), but the outwards scenario could not be detected for all kinds of indicators. This makes sense because, even when the jaw moves outwards, the leaves still there to keep the field shape and consequently have no influence on the transmit dose. The jaw moving inwards has the opposite effect; moving inwards would have an impact on the field edge fluence, further influencing the \({\gamma }_{mean}\) and \({\gamma }_{pass rate}\) values. The DRP was defined in the mass center of the PTV, which would be affected if the jaw blocked the define point.
A 2 cm setup error for the CIRS phantom in the anterior, left and superior directions was successfully detected. This makes sense when the phantom is moved 2 cm to the left, which would increase the thickness of the body during the irradiation fields and result in a decrease in the dosimetry received by the EPID. Moving two centimeters in the superior direction moved the superior spherical target, resulting in an increased isocenter dose and a decreased anterior shift SSD, resulting in an increased DRP.
Thickness errors were detected by all four indicators in the EPID-based dosimetry system for lung SBRT plans. After taping a 1 cm bolus on the phantom to simulate the weight changes that frequently occur during a treatment course, the expected dose decreased in the isocenter due to an increase in built-up dose. The DRP decreased from 0.6% to -6.3%, which was consistent with the results of the Mijnheer et al. [18] study, which showed a 6.9% difference for lung cases. Furthermore, \({\gamma }_{pass rate}\) obtains the worst value during all error scenarios simulated in this study.
Gamma analysis is widely used to compare the planned and measured dose distributions in the clinic. Christos Moustakis et al. recommended thin CT slices and tighter criteria (2%/2 mm), as well as local gamma evaluation in hypofractionated VMAT irradiation for 3D in vivo dosimetry verification[27]. Gamma analysis combines dose-difference and distance-to-agreement in a mathematical model, and contains no information on the dose distribution. Van der et al. compared DVH-based and multiparametric \(\gamma\)-based methodologies for the iViewDose system for pelvic VMAT plans and concluded that \({\gamma }_{mean}\) and \({\gamma }_{pass rate}\) were strongly correlated with DVH indicators and equivalent for 3D in vivo dosimetric verification of VMAT pelvic treatment [28]. The iViewDose system has four indicators: \({\gamma }_{mean}\), \({\gamma }_{1\%}\), \({\gamma }_{pass rate}\) and DRP. \({\gamma }_{mean}\) can detect 13 serious errors out of a total of 30 errors (43.3%), \({\gamma }_{1\%}\) can detect 8 out of 30 errors (26.7%), \({\gamma }_{pass rate}\) can detect 14 out of 30 errors (46.7%), and the DRP can detect 8 out of 30 errors (26.7%). Because the site used in the study was lung SBRT, with a regular and rotation-symmetric target, detection failure was expected for gantry angle and collimator angle error scenarios by EPID-based transit dosimetry. Excluding those errors, the detection rates for \({\gamma }_{mean}\), \({\gamma }_{1\%}\), \({\gamma }_{pass rate}\) and DRP were 13 out of 21 (61.9%), 8 out of 21 (38.1%), 14 out of 21 (66.7%), and 8 out of 21 (38.1%), respectively. The main error scenarios not detected by the \({\gamma }_{mean}\), and \({\gamma }_{pass rate}\) indicators were output errors. As previously discussed, one possible reason was that the amplitude of the simulated MU errors was too small to detect in sinusoidal form. Aside from the output errors, these indicators could detect almost all the error scenarios, with detection rates of 13 out of 16 (81.3%) and 14 out of 16 (87.5%). The ability of the \({\gamma }_{1\%}\) indicator and the DRP to serve as additional reference indicators appears to be limited, as no error was captured solely by either of them.
Alert criteria are generally a compromise between the need to detect all deviations and the workload required to analyze alerts. Some deviations were technically classified as serious errors, but they were not severe enough to have a serious impact on treatment quality. Treatment site specificity should be considered an important factor when alerts occur, and the alert threshold should also be related to a specific site [14].