3.1 Comparison of different collectors
The regent blanks of Bi2O3, NiO and PbO were respectively determined by Bi-FA-ICP-MS, NiS-FA ICP-MS and Pb-FA GF-AAS, and the values were shown in Table 4. It could be seen that the reagent blank of Pt and Pd in Bi2O3 was extremely low compared to commercial PbO and NiO. Therefore, Bi2O3 could be used as collector directly without purification. Besides, the toxicity of Bi2O3 was low and Bi-FA was a green analytical method.
Table 4
Total procedure blanks (mean data ± standard deviation, n = 5) for Pt and Pd using commercially available Bi2O3, NiO and PbO (ng g− 1)
Element | Bi2O3 | NiO | PbO |
Pt | 0.54 ± 0.12 | 56 ± 21 | 2.6 ± 0.9 |
Pd | 0.52 ± 0.09 | 75 ± 34 | 3.5 ± 0.6 |
3.2 Cupellation temperature
Metal Bi can be oxidized to Bi2O3 at ~ 300℃. However, the melting point of Bi2O3 is 820℃. Therefore, the cupellation temperature must be controlled to above 820℃ so that liquid Bi2O3 could be absorbed by the magnesia cupel. The effect of cupellation temperature was optimized with the data shown in Fig. 1. It can be seen that the cupellation speed was accelerated, and Bi content remaining in the Ag granule was also decreased with increasing of cupellation temperature. If the temperature was too high, the loss of Pt and Pd as well as muffle furnace would be increased. Therefore, 940℃ was selected as the cupellation temperature.
After Bi cupellation, the target Pt and Pd were trapped in the Ag granule. The Ag granule was annealed at 700 oC to further homogenize the alloy of Pt, Pd and Ag, and which was compressed into ~ 0.2 mm slices to facilitate the injection of the laser ablation system.
3.3 Preparation of Ag slices
The solid samples determined by LA-ICP-MS were required to be as uniform as possible, thus the effect of annealing on the signal stability of 105Pd and 195Pt were evaluated and shown in Fig. 2. It was observed that, the signal fluctuation before annealing was obviously larger than that after annealing. Therefore, the Ag granules were annealed at 700 oC to ensure the uniformity of the target Pt and Pd inside the Ag slices.
3.4 Mass spectral interferences
The isotopes of Pt, Pd and Ag were shown in Table S1, based on the principle of high abundance and free from isobaric interference, 195Pt and 105Pd were selected as measuring isotopes. Though 107Ag and 109Ag are close in abundance, the mass charge ratio difference of 109Ag/105Pd is larger than 107Ag/105Pd, then 109Ag was selected as internal standard isotope. The possible interferences from polyatomic molecular ions on 195Pt, 105Pd and 109 Ag were shown in Table S2. After Bi-FA and cupellation, there were only trace Bi, Au, Pt, Pd, Ru, Rh and Ir reserved in Ag granule. Compared to traditional solution injection system, laser ablation solid sample injection could avoid the introduction of large amount of Cl, N, H and O into the ICP. Based on the above means, the possible interferences of polyatomic molecular ions could be effectively eliminated.
3.5 Modified empirical coefficient method
The empirical coefficient method is based on the certified values and signal strength of a series of CRMs, using linear or nonlinear regression methods to obtain the coefficients for the quantification formula and allow quantitative sample analysis [30], which is usually used in X-ray fluorescence analysis of solid samples. However, the traditional empirical coefficient method has a high demand on sample matrix, which requires the composition and structure of the CRMs and the real sample to be tested should be highly similar. Up to now, only a few geochemical Certified Reference Materials included soil, stream sediments, peridotite, chromite and Pt-Pd ores were developed by China. Some special samples, such as polymetallic ore and black shale, the matrix was not identical to the existing CRMs, the accuracy of the method will be affected.
In this work, a fully matrix-matched Ag slices were obtained and modified empirical coefficient method LA-ICP-MS was established for the determination of ultra-trace Pt and Pd in multiple geochemical samples. Bi-FA was employed to enrich the target precious metal elements from the CRMs and real samples (such as soil, river sediment, chromite, olivinite and Pt-Pd ore) into the Bi granule. After Ag protection cupellation, Pt and Pd were enriched in the Ag slices, fully matrix-matched was achieved and the typical Ag slices of CRMs were shown in Fig. 3. Details of the matrix-matched Pt and Pd mixed external standard CRMs series in the Ag slices are shown in Table 5. The concentrations for Pt and Pd in the media of Ag slices were 0.4071-2958 µg g− 1 and 0.4071-2636 µg g− 1, respectively.
Table 5
The mass fractions of Pt and Pd in the Ag slices of CRMs
Standard serials | Certified values of Pt/ ng g− 1 | Certified values of Pd/ ng g− 1 | mAg/µg | Pt mass fractions in the Ag slices/µg g− 1 | Pd mass fractions in the Ag slices /µg g− 1 |
GBW07288 | 0.26 | 0.26 | 6387 | 0.4071 | 0.4071 |
GBW07340 | 0.66 | 0.66 | 6437 | 1.025 | 1.025 |
GBW07289 | 1.6 | 2.3 | 6292 | 2.543 | 3.655 |
GBW07290 | 6.4 | 4.6 | 6764 | 9.462 | 6.801 |
GBW07294 | 14.7 | 15.2 | 6208 | 23.68 | 24.48 |
GBW07291 | 58 | 60 | 6383 | 90.87 | 94.00 |
GBW07293 | 440 | 570 | 6626 | 664.1 | 860.2 |
GBW07341 | 1900 | 570 | 6423 | 2958 | 887.4 |
GBW07342 | - | 1670 | 6335 | - | 2636 |
3.6 Internal standard calibration strategy for LA-ICP-MS
In this work, internal standard calibration method was employed to reduce analysis error and correct biases resulting from fluctuations in laser output power as well as sample ablation amount and transport efficiency, to improve the method precision and accuracy. Due to Ag content in the Ag slices were almost identical between CRMs and real samples, then 109Ag in the Ag slices was selected as internal standard isotope for the determination of 195Pt and 105Pd. According to the basic principle and formula (Formula 1) of internal standard method [29], the concentration of the target element in real sample could be calculated.
\(\frac{{w_{t}^{{sam}}}}{{w_{t}^{{std}}}}=\frac{{I_{t}^{{sam}}w_{i}^{{sam}}/I_{i}^{{sam}}}}{{I_{t}^{{std}}w_{i}^{{std}}/I_{i}^{{std}}}}=\frac{{I_{t}^{{sam}}/I_{i}^{{sam}}}}{{I_{t}^{{std}}/I_{i}^{{std}}}}\) (1)
Where \(w_{t}^{{sam}}\) and \(w_{t}^{{std}}\) are the concentrations of target element (Pt and Pd) in Ag slices of real and standard samples (µg g− 1), respectively; \(w_{i}^{{sam}}\) and \(w_{i}^{{std}}\) are concentrations of internal standard element (Ag) in Ag slices of real and standard samples (µg g− 1), respectively. In our experiment, as the concentrations of internal standard element in real and standard samples were the same by adding the same amount of Ag standard solution during Ag protection cupellation procedure, then the formula is simplified; \(I_{t}^{{sam}}\) and \(I_{t}^{{std}}\) are the signal strength of target element in Ag slices of real and standard samples (cps), respectively; \(I_{i}^{{sam}}\) and \(I_{i}^{{std}}\) are the signal strength of internal standard element in Ag slices of real and standard samples (cps), respectively.
The internal standard and non-internal standard method RSDs (n = 10) were compared in Table 6. It is observed that the RSDs of 195Pt and 105Pd by non-internal standard method were between 5.71–6.68%. In comparison, the RSDs were reduced to 2.54%-4.05% when internal standard method was used.
Table 6
Determined RSDs of non-internal standard and internal standard LA-ICP-MS methods
| I195Pt/cps | I105Pd/cps | I109Ag/cps | I195Pt /I109 Ag | I105Pd /I109 Ag |
1 | 4432 | 3518 | 582357456 | 0.0000076097 | 0.0000060418 |
2 | 4768 | 3922 | 646279911 | 0.0000073784 | 0.0000060690 |
3 | 4646 | 3737 | 615638164 | 0.0000075466 | 0.0000060699 |
4 | 4256 | 3564 | 582410079 | 0.0000073067 | 0.0000061186 |
5 | 4292 | 3590 | 573730805 | 0.0000074804 | 0.0000062581 |
6 | 4593 | 3472 | 582566560 | 0.0000078833 | 0.0000059592 |
7 | 4225 | 3528 | 586953735 | 0.0000071989 | 0.0000060099 |
8 | 4289 | 3464 | 573790334 | 0.0000074740 | 0.0000060370 |
9 | 4900 | 3957 | 617387914 | 0.0000079373 | 0.0000064091 |
10 | 3911 | 3310 | 565210208 | 0.0000069203 | 0.0000058559 |
Average | 4431 | 3606 | 592632517 | 0.0000074736 | 0.0000060828 |
RSD / % | 6.68 | 5.71 | 4.29 | 4.05 | 2.54 |
3.7 Analytical performance
The Pt and Pd mixed external standard series were prepared with the concentrations of 0.4071-2958 µg g− 1 and 0.4071-2636 µg g− 1 in the media of Ag slices. The standard series of CRMs are shown in Table 5. At the optimum conditions, the intensities of 195Pt, 105Pd and 109Ag were detected by LA-ICP-MS, and the concentrations of target elements were calculated by formula 1.
The analytical performance of the proposed Bi-FA LA-ICP-MS method has been validated using the calibration curve equation, fit coefficient and LODs, shown in Table 7. Excellent curve fitting of Pt and Pd were obtained (0.4071 ~ 2958 µg g− 1 and 0.4071 ~ 2636 µg g− 1, respectively), with the correlation coefficients exceeding 0.9996. Based on 3δblank approach as recommended by IUPAC for spectrochemical measurements, the LODs of the proposed method for the target Pt and Pd were 0.074 and 0.037 ng g− 1, respectively. The LODs for Pt and Pd obtained by this method along with other methods were compared. The results in Table 8 revealed that, due to the high enrichment factor (about 1667 fold, 10 g sample weight pre-concentrated into ~ 6 mg Ag granules) the LODs obtained in this work and our previous Pb-FA LA-ICP-MS methods [29] were much lower than those low enrichment factor methods based LA-ICP-MS and NiS/Pb fire assay [24–27].
Table 7
Analytical performance data by the established Bi-FA LA-ICP-MS method for Pt and Pd.
Isotopes | Linear range/ µg g− 1 | Linear equation | R2 | LOD*/ ng g− 1 |
195Pt | 0.4071ཞ2958 | y = 0.0000009586x-0.0000009065 | 0.9998 | 0.074 |
105Pd | 0.4071ཞ2636 | y = 0.0000007354x + 0.000004545 | 0.9996 | 0.037 |
Table 8
Comparison of detection limits using this proposed technique and other conventional methods
Analytical technique | LODs | Ref. |
Pt | Pd |
Pb-FA Spark-OES | 10 ng g− 1 | 100 ng g− 1 | 25 |
Pb-FA LA-ICP-MS | 30 ng g− 1 | 25 ng g− 1 |
Pb-FA GD-MS | 7 ng g− 1 | 9 ng g− 1 |
NiS-FA LA-ICP-MS (dynamic reaction cell) | 20 ng g− 1 | 28 ng g− 1 | 26 |
NiS-FA LA-ICP-MS (focusing sector field MS) | 11 ng g− 1 | 17 ng g− 1 | 27 |
Pb-FA femtosecond LA-ICP-MS | 6 ng g− 1 | 9 ng g− 1 | 24 |
Pb-FA LA-ICP-MS | 0.06 ng g− 1 | 0.03 ng g− 1 | 29 |
Bi-FA LA-ICP-MS | 0.074 ng g− 1 | 0.037 ng g− 1 | This work |