3.1 Corrosion morphology and result analysis
By Fig. 1(a) and (b), the lower part of the two figures is the weld area, and the upper part is the heat affected zone and a small amount of base metal area. the weld area has the best corrosion resistance in the three areas and poor corrosion resistance in the other two areas, corrosion steps can be seen near the joint fusion line. The corrosion degree of the joint after cryogenic treatment is less than that of the untreated joint, Moreover, the corrosion in the three areas of the joint is relatively uniform on the whole, and the corrosion steps still exist, but not obvious. It can be seen that the welded joint after cryogenic treatment has good corrosion resistance, and the threat of local corrosion fracture of the joint is reduced due to the homogenization of corrosion.
3.3 Electrochemical test and result analysis
From Fig. 2(a), with the extension of the deep cold treatment time, the grain is refined, the self-corrosion potential of the joint sample is constantly negative shift, the corrosion current is gradually reduced, the width of the passivation steps is gradually increasing, the maximum value is reached when the time is 8h. Under this parameter, the protective film layer on the surface of the joint is the most stable and the corrosion resistance is the best. The difficulty of the cathodic reaction affects the corrosion rate of the joint sample. As can be seen from the figure that the cathodic reaction rate of different cryogenic parameters gradually decreases with the increase of cryogenic treatment time. Similarly, as shown in Fig. 2(b), with the decrease of cryogenic treatment temperature, the self-corrosion potential and self-corrosion current generally show a decreasing trend, the cathode branch curve shows a decreasing trend, the resistance of cathode hydrogen evolution reaction increases, and the hydrogen evolution corrosion rate of cathode gradually decreases. When the parameter is -180℃ for 8h, The corrosion resistance of the alloy is the best (the self corrosion potential is reduced by 0.017v compared with the untreated joint sample, and the corrosion current density is reduced by one order of magnitude), the passivation range is the most obvious, and the logarithm of corrosion current density is the lowest.
Since charge transfer and corrosion film are formed on the surface of the sample, the equivalent circuit is used to verify, and two sets of parallel resistors and capacitors are used to form a parallel circuit, as shown in Fig. 3, Rs is the solution resistance, Rct is the charge transfer resistance, Rfilm is the protective resistance formed by the corrosion film, Cdl is the electric double-layer capacitance, and Cfilm is the corrosion film capacitance.
CPE=Cfilm Cdl
Table 3
Fitting result of Rct and Rfilm of the joint by deep cryogenic of 8h for different holding time
|
Rct(ohm cm2)
|
Rfilm(ohm cm2)
|
untreated
|
920
|
2089
|
-100℃, 6h
|
1036
|
2103
|
-140℃, 6h
|
1251
|
2175
|
-180℃, 6h
|
1384
|
2631
|
Figure 4 shows the impedance spectrum of joint samples insulated for 6h at different cryogenic treatment temperatures, and the untreated samples are used as comparison. The diameter of the semicircular arc in the impedance spectrum is proportional to the electron transfer resistance, and the larger the diameter, the harder the sample is to be corroded[8, 9]. The diameter of the semi-circular arc was continuously increased during the cryogenic treatment, and the corrosion resistance of the sample gradually became better. When the cryogenic parameter is -180℃ for 6h, the semi-circular arc diameter is the largest and its corrosion resistance is the best. Table 3 shows the fitting results of the impedance values before and after cryogenic treatment by the corrosion current density calculation formula. Taking the untreated sample as a reference, when the cryogenic treatment parameter is -180℃ for 6h, the corresponding Rct and Rfilm are the maximum.
Table 4
Fitting result of Rct and Rfilm of the joint by deep cryogenic at 140℃ for Different Time
|
Rct(ohm cm2)
|
Rfilm(ohm cm2)
|
untreated
|
920
|
2089
|
-140℃, 4h
|
1093
|
2041
|
-140℃, 6h
|
1251
|
2175
|
-140℃, 8h
|
1591
|
3382
|
Figure 5 shows the impedance spectrum of joint samples insulated for different times at-140℃, with the extension of cryogenic treatment time, the diameter of semicircular arc increases and the corrosion resistance of the sample improves. The best corrosion resistance is -140℃ for 8h. According to the calculation formula of corrosion current density, the impedance values before and after cryogenic treatment were fitted, as shown in Table 4. With the extension of cryocooling treatment time, both Rct and Rfilm increased, and the corrosion resistance of the joint gradually became better. The distribution of β-phase and grain size affect the corrosion resistance of the sample. When the continuous distribution of β-phase is relatively rare and the grain size is large, it is difficult to block the corrosion of the α phase in the Cl− environment, Cryogenic treatment improves the microstructure of the joint and the distribution of β-phase, making the β-phase finely dispersed and strengthening the corrosion barrier effect of β-phase relative to α-phase. Cryogenic treatment also makes the structure uniform and refined, and improves corrosion resistance.