DOI: https://doi.org/10.21203/rs.3.rs-149964/v1
PURPOSE: Choroidal thickness is associated with many ocular conditions, interchangeability among different generations of optical coherence tomography is therefore important for both research purpose and clinical application. Hence, we compared choroidal thickness measurements between spectral-domain optical coherence tomography (SD-OCT) and swept-source optical coherence tomography (SS-OCT) in healthy pediatric eyes.
METHODS: Children from the population–based Hong Kong Children Eye Study were recruited. Choroidal thickness was measured by both devices. Intra-class correlation coefficient (ICC) was used to compare the measurements.
RESULTS: A total of 114 children with mean age of 7.38±0.82 years were included. The central foveal choroidal thickness (CFCT) measured by SD-OCT and SS-OCT was 273.24±54.29μm and 251.84±47.12μm respectively. Inter-device correlation coefficient was 0.840 (95%CI: 0.616-0.918). However, choroidal thickness obtained by SD-OCT was significantly thicker than that measured by SS-OCT with a mean difference of 21.40±33.13μm (P<0.001). Bland-Altman limit of agreement on the relative difference scale for SD-OCT/SS-OCT was 86.33μm. Validated conversion equation for translating SD-OCT CFCT measurement into SS-OCT was SS-OCT = 35.261 + 0.810 x SD-OCT.
CONCLUSIONS: ICC shows an acceptable agreement between SD-OCT and SS-OCT, however, there was a significant inter-device difference of choroidal thickness measurements in normal children eyes. Therefore, the measurements are not interchangeable.
The highly-vascularised choroid is the middle tunic of the eye, supplying nearly 90% of ocular artery blood flow. It is crucial in thermoregulation, regulation of intraocular pressure (IOP) and drainage of aqueous humor. Longitudinal studies of choroid in children suggest its role in eye development, including emmotropization, growth factor secretion and scleral growth modulation. 1, 2
Associations of choroidal thickness as measured by spectral-domain optical coherence tomography (OCT) with enhanced depth imaging (EDI) protocol have been reported for retinal pathologies, including myopia. Consistent evidence has shown the close link between myopia and choroidal thinning in both children and adults. 1–4 Therefore, choroidal changes since childhood may be an early indictor in myopic development. Accurate and reliable monitoring of temporal changes in choroidal thickness may help in studying the role of choroid in myopia.
Using in the EDI mode with spectral-domain OCT, peak sensitivity of measurements is positioned at the inner sclera, enabling visualisation of deeper tissue layers including choroidal structure. 5–8 Whereas SS-OCT, the newer generation of OCT, uses a tunable laser beam to sweep across various layers with a single photodiode detector. By employing light beam with a longer wavelength, SS-OCT could provide a higher quality of choroid imaging by having less signal noise and better penetration in deeper structures. 5–7
Owing to the advantage of greater scanning depth and availability of commercialised SS-OCT, many researchers have switched from SD-OCT to SS-OCT in assessing deep structures like choroid. Nevertheless, it is still questionable whether measurements made by SD-OCT and SS-OCT images are interchangeable. Many studies comparing the two commonly used generations of OCT (SD-OCT and SS-OCT) have been focusing on either normal 9–15 or diseased 12, 15, 16–18 adult eyes. However, little is known about the inter-device variability in choroidal thickness measurements in children eyes.
However, such information may have been obtained from different generations of OCT across the years, determining the interchangeability among two generations of OCT are therefore of significant use. Furthermore, valid conversion equations would enable studies to pool data from two generations of OCT and
In view of the potential utilisation of choroidal thickness data from different generations of OCT to monitor choroidal thickness changes for research purpose or clinical application, we would like to compare the choroidal thickness measurements between SD-OCT and SS-OCT in school children in the Hong Kong Children Eye Study (HKCES). Conversion equations enable longitudinal trials to pool data from two generations of OCTs to monitor changes across the years.
Data for this analysis were derived from the Hong Kong Children Eye Study (HKCES), a population-based cohort study of eye conditions in children aged 6 to 8 years. 19–23 Study subjects were recruited consecutively from November to December 2015. All underwent the comprehensive ophthalmic examination, physical examination and standardised questionnaires at the Chinese University of Hong Kong Eye Centre.
Informed consent was obtained from all subjects and their legal guardians before participation. The study procedure was performed in accordance with the Declaration of Helsinki and the protocol was approved by the Ethic Committee Board of the Chinese University of Hong Kong. Subjects with ocular diseases (except myopia and hypermetropia), congenital eye malformations, prior eye trauma, previous eye surgery or incapable to complete optical coherence tomography, were excluded. In this study, only right eyes were analyzed in view of the high correlation between both eyes.
Each child’s cornea, anterior chamber, iris, pupil reflex, crystalline lens, anterior vitreous and eyelids were examined by an ophthalmologist using Haag-Streit slit-lamp (Koeniz, Switzerland) to exclude underlying ocular diseases. Pre- and post-cycloplegic refractive status were measured by autorefractor (Nidek ARK-510A, Gamagori, Japan). Two rounds of cycloplegic agents (1% cyclopentolate (Alcon, Belgium) and 1% tropicamide (Santen, Japan)) were applied at 10 minutes apart. Post-cycloplegic spherocylindrical autorefraction was measured at least 30 minutes after the last drop of cycloplegic agent.
Body height and weight were measured by a professional integrated set (seca, Hamburg, Germany). Waist and head circumferences were measured using a tape measurer.
For Spectral-domain Optical Coherence Tomography (SD-OCT), Spectralis OCT (Heidelberg Engineering, Heidelberg, Germany) was employed for choroidal imaging. In the enhanced depth imaging (EDI) protocol, an 870-nm-wavelength light source was used at a scanning speed of 40,000 A-scans per second for image acquisition. A 45-degree cross-line scan pattern was adopted. 24 All images were inspected, and choroidal layer was manually segmented using a MATLAB software by a well-trained examiner (C.O.L.). In the segmentation for each radial scan, the fovea was denoted by a dot and 31 points were plotted for both the upper and lower boundaries of choroidal layer.
For Swept-source Optical Coherence Tomography (SS-OCT), Triton DRI-OCT (Topcon, Inc, Tokyo, Japan) was used. A 1050-nm-wavelength swept-laser was used with a scanning speed of 100,000 A-scans per second. Image was obtained in 12-line radial scan pattern with a resolution of 1059 × 400. An average of 32 overlapped consecutive scans, covering an area of 12 mm × 9 mm, was acquired for each image. Choroidal layer was segmented using a built-in software. All segmented images were further inspected and manually modified to ensure accuracy.
We defined choroidal thickness as the perpendicular distance between the outer border of the retinal pigment epithelium and the inner border of the sclera. 25 Measurement of choroidal thickness of a study subject through SD-OCT and SS-OCT imaging is shown in Figs. 1 and 2 respectively. For SS-OCT, choroidal thickness map was presented using the Early Treatment Diabetic Retinopathy Study (ETDRS) grid. 26 Mean regional thickness was calculated for each of the 9 sectors in the ETDRS grid. For SD-OCT, we divided the whole region into 9 sectors as in the ETDRS grid for easier comparison with SS-OCT. Central foveal choroidal thickness (CFCT) was derived from the average thickness from 4 radial scans inside central foveal circle. Choroidal thicknesses in other regional sectors (i.e. S1, S2, I1, I2, T1, T2, N1 and N2) were derived from the average thicknesses from 2 radial scans forming two linear edges of the annulus and 1 radial scan passing through the mid-point of outer and inner circular (Fig. 3). 27
For quality control, all participants completed both optical coherence tomography on the same day. To avoid the discrepancy between different examiners, only one person (C.O.L.) was trained to measure the choroidal thickness for both SD-OCT and SS-OCT. The examiner was masked to all participant characteristics. We exclude participants if any of the optical coherence tomography images in either device was in poor quality. Images were graded as poor quality if either the fovea or choroid-scleral junction was poorly visualised. Intra-grader reliability was high for both SD OCT and SS-OCT with intra-class correlation coefficient (ICC) of 0.969 or greater (Supplementary Table 1).
Data were analyzed with SPSS (SPSS, Inc., Chicago, IL, version 24.0). Continuous parameters were expressed as mean ± standard deviation (SD). Bland-Altman plot was used to assess the agreement and limit of agreement. ICCs were analyzed for the agreement between the two devices. A P-value of less than 0.05 was considered statistically significant.
Conversion equations were derived to translate measurement of choroidal thickness from SD-OCT into a predicted SS-OCT value. Outliers (N = 14) (defined as CFCT measurement difference by two devices is > 2 SDs away from the mean difference) were excluded from the derivation of conversion equations. Random half sample of the remaining participants (N = 48) were used to determine the conversion equations using linear regression and second half of the participants (N = 47) were used to validate the derived equations. Validity is evaluated by comparing observed and predicted SS-OCT measurements computed by the conversion equations.
After excluding 31 children with poor quality of optical coherence tomography images, a total of 114 Chinese children, 57 (50.0%) boys and 57 (50.0%) girls, who successfully completed ophthalmic examinations were included for analysis. The mean age was 7.38 ± 0.82 years. Demographic characteristics, systemic and ocular parameters of the included participants are summarized in Table 1.
n = 114 |
Total (mean ± SD) |
---|---|
Gender, female |
n = 57 (50.0%) |
Age, years |
7.38 ± 0.82 |
Height, cm |
123.51 ± 6.23 |
Weight, kg |
24.00 ± 4.20 |
Spherical equivalent refraction (SER), diopters (D) |
0.15 ± 1.40 |
Post-cycloplegic spherical power, diopters (D) |
0.50 ± 1.43 |
Post-cycloplegic cylindrical power, diopters (D) |
-0.67 ± 0.81 |
Axial length, mm |
23.05 ± 0.92 |
SD: standard deviation; SD-OCT: spectral-domain optical coherence tomography; SS-OCT: swept-source optical coherence tomography. |
By SD-OCT measurement, the CFCT was 273.24 ± 54.29µm. Other regional sectors were inner superior (S1): 270.39 ± 50.91µm; outer superior (S2): 263.25 ± 45.84µm; inner inferior (I1): 266.94 ± 53.94µm; outer inferior (I2): 256.57 ± 50.65µm; inner temporal (T1): 283.41 ± 50.81µm; outer temporal (T2): 286.88 ± 43.48µm; inner nasal (N1): 252.39 ± 51.43µm; and outer nasal (N2): 224.48 ± 47.20µm. By SS-OCT measurement, the CFCT was 251.84 ± 47.12µm. Other regional sectors were S1: 261.32 ± 48.93µm; S2: 247.65 ± 44.54µm; I1: 255.07 ± 48.17µm; I2: 240.40 ± 46.40µm; T1: 274.56 ± 45.21µm; T2: 271.32 ± 40.11µm; N1: 230.68 ± 56.99µm; and N2: 183.90 ± 61.63µm (Table 2).
Total Participants (n = 114) |
||||
---|---|---|---|---|
Mean ± SD |
SD-OCT |
SS-OCT |
P-value |
Mean Difference |
Center,µm |
273.24 ± 54.29 |
251.84 ± 47.12 |
< 0.001 |
21.40 ± 33.13 |
S1,µm |
270.39 ± 50.91 |
261.32 ± 48.93 |
0.008 |
9.07 ± 36.07 |
S2,µm |
263.25 ± 45.84 |
247.65 ± 44.54 |
< 0.001 |
15.60 ± 27.42 |
I1,µm |
266.94 ± 53.94 |
255.07 ± 48.17 |
0.001 |
11.87 ± 37.47 |
I2,µm |
256.57 ± 50.65 |
240.40 ± 46.40 |
< 0.001 |
16.17 ± 32.43 |
T1,µm |
283.41 ± 50.81 |
274.56 ± 45.21 |
0.003 |
8.85 ± 31.38 |
T2,µm |
286.88 ± 43.48 |
271.32 ± 40.11 |
< 0.001 |
15.56 ± 28.50 |
N1,µm |
252.39 ± 51.43 |
230.68 ± 56.99 |
< 0.001 |
21.71 ± 43.41 |
N2,µm |
224.48 ± 47.20 |
183.90 ± 61.63 |
< 0.001 |
40.57 ± 49.61 |
Spherical equivalent refraction (SER) |
||||
Myopia (SER ≤ -0.50 D) (n = 28) |
||||
Center,µm |
244.83 ± 46.86 |
227.36 ± 43.05 |
0.022 |
17.48 ± 38.02 |
S1,µm |
245.72 ± 42.69 |
236.68 ± 47.00 |
0.336 |
9.04 ± 48.88 |
S2,µm |
241.48 ± 32.50 |
226.04 ± 42.87 |
0.038 |
15.44 ± 37.35 |
I1,µm |
240.49 ± 47.23 |
236.00 ± 52.04 |
0.622 |
4.49 ± 47.56 |
I2,µm |
233.34 ± 47.98 |
224.46 ± 50.33 |
0.341 |
8.87 ± 48.43 |
T1,µm |
257.71 ± 46.93 |
251.96 ± 47.24 |
0.434 |
5.74 ± 38.27 |
T2,µm |
263.49 ± 38.48 |
250.79 ± 43.27 |
0.049 |
12.71 ± 32.61 |
N1,µm |
225.08 ± 39.75 |
209.57 ± 58.15 |
0.161 |
15.51 ± 56.97 |
N2,µm |
201.28 ± 36.97 |
165.25 ± 62.43 |
0.006 |
36.03 ± 64.05 |
Emmetropia (-0.50 D < SER < + 0.50 D) (n = 33) |
||||
Center,µm |
275.68 ± 60.87 |
251.06 ± 50.27 |
0.001 |
24.62 ± 38.20 |
S1,µm |
267.84 ± 51.11 |
257.12 ± 48.92 |
0.031 |
10.72 ± 27.31 |
S2,µm |
260.02 ± 42.83 |
244.79 ± 43.29 |
< 0.001 |
15.23 ± 20.17 |
I1,µm |
271.01 ± 60.78 |
251.7 ± 46.85 |
0.005 |
19.31 ± 36.93 |
I2,µm |
258.48 ± 55.95 |
235.94 ± 45.24 |
< 0.001 |
22.55 ± 28.01 |
T1,µm |
288.11 ± 57.04 |
275.09 ± 46.69 |
0.026 |
13.02 ± 32.09 |
T2,µm |
286.54 ± 44.4 |
268.94 ± 33.78 |
0.002 |
17.60 ± 30.63 |
N1,µm |
253.82 ± 56.47 |
223.55 ± 53.53 |
< 0.001 |
30.28 ± 31.35 |
N2,µm |
222.88 ± 50.25 |
177.42 ± 58.65 |
< 0.001 |
45.46 ± 36.14 |
Hypertrophia (SER ≥ + 0.50 D) (n = 53) |
||||
Center,µm |
286.73 ± 48.60 |
265.26 ± 42.43 |
< 0.001 |
21.47 ± 26.86 |
S1,µm |
285.01 ± 50.30 |
276.96 ± 44.67 |
0.087 |
8.05 ± 33.55 |
S2,µm |
276.77 ± 49.40 |
260.85 ± 42.07 |
< 0.001 |
15.92 ± 25.67 |
I1,µm |
278.39 ± 48.68 |
267.25 ± 43.95 |
0.012 |
11.14 ± 31.17 |
I2,µm |
267.66 ± 45.10 |
251.60 ± 42.75 |
< 0.001 |
16.06 ± 23.02 |
T1,µm |
294.07 ± 44.48 |
286.17 ± 39.09 |
0.038 |
7.90 ± 27.00 |
T2,µm |
299.44 ± 40.88 |
283.64 ± 37.90 |
< 0.001 |
15.80 ± 25.05 |
N1,µm |
265.93 ± 48.73 |
246.28 ± 54.93 |
0.001 |
19.65 ± 41.63 |
N2,µm |
237.73 ± 45.92 |
197.79 ± 60.84 |
< 0.001 |
39.93 ± 48.87 |
SD: standard deviation; SD-OCT: spectral-domain optical coherence tomography; SS-OCT: swept-source optical coherence tomography; Center: Central foveal; S1: Inner superior: S2: Outer superior; I1: inner inferior; I2: outer inferior; T1: Inner temporal; T2: Outer temporal; N1: Inner nasal; N2: Outer nasal. |
For both devices, it was observed that choroidal thicknesses in parafoveal regional sectors were thicker than those in perifoveal regional sectors. Also, temporal sectors had the thickest choroid while nasal sectors had the thinnest choroid among all regional sectors.
There was good intra-class correlation for CFCT with an ICC of 0.840 (95% confidence interval (CI): 0.616–0.918) for absolute measurements and 0.881 (95%CI: 0.828–0.918) for relative agreement (P < 0.001) (Table 3). Other regional sectors showed similar results with good ICCs, except for nasal sectors with slightly poor ICCs (N1: 0.774, 95%CI: 0.595–0.864; N2: 0.636, 95%CI: 0.150–0.815 for absolute measurements). Nevertheless, the mean difference for CFCT was 21.40 ± 33.13µm and the thickness measurements from SD-OCT in all regional sectors were statistically significant thicker than that from SS-OCT (P < 0.001) (Table 2). Bland-Altman limit of agreement on the relative difference scale for SD-OCT/SS-OCT was 86.33µm (Fig. 4). The derived conversion equation for translating SD-OCT CFCT measurement into SS-OCT was SS-OCT = 35.261 + 0.810 x SD-OCT. Validation showed satisfactory ICC (0.942, 95%CI: 0.898–0.967). The predicted value fell within 10% of the actual SS-OCT CFCT measurement 94.7% of the time (Table 4).
Total Participants (n = 114) |
||||
---|---|---|---|---|
Regional sectors |
Absolute ICC (95%CI) |
Relative ICC (95%CI) |
P-value |
|
Center |
0.840 (0.616–0.918) |
0.881 (0.828–0.918) |
< 0.001 |
|
S1 |
0.843 (0.869–0.893) |
0.850 (0.783–0.896) |
< 0.001 |
|
S2 |
0.871 (0.722–0.930) |
0.899 (0.853–0.930) |
< 0.001 |
|
I1 |
0.833 (0.746–0.889) |
0.845 (0.775–0.893) |
< 0.001 |
|
I2 |
0.849 (0.716–0.912) |
0.875 (0.818–0.913) |
< 0.001 |
|
T1 |
0.873 (0.810–0.915) |
0.881 (0.827–0.918) |
< 0.001 |
|
T2 |
0.837 (0.672–0.908) |
0.869 (0.810–0.909) |
< 0.001 |
|
N1 |
0.774 (0.595–0.864) |
0.810 (0.724–0.869) |
< 0.001 |
|
N2 |
0.636 (0.150–0.815) |
0.743 (0.628–0.823) |
< 0.001 |
|
Spherical equivalent refraction (SER) |
||||
Myopia (SER ≤ -0.50 D) (n = 28) |
||||
Center |
0.754 (0.454–0.888) |
0.783 (0.530–0.899) |
< 0.001 |
|
S1 |
0.579 (0.097–0.805) |
0.579 (0.090–0.805) |
0.014 |
|
S2 |
0.654 (0.271–0.838) |
0.682 (0.314–0.853) |
0.002 |
|
I1 |
0.709 (0.366–0.866) |
0.703 (0.358–0.863) |
0.001 |
|
I2 |
0.680 (0.314–0.852) |
0.680 (0.308–0.852) |
0.002 |
|
T1 |
0.804 (0.579–0.909) |
0.802 (0.572–0.908) |
< 0.001 |
|
T2 |
0.794 (0.552–0.905) |
0.811 (0.593–0.913) |
< 0.001 |
|
N1 |
0.505 (-0.039-0.767) |
0.514 (-0.050-0.775) |
0.033 |
|
N2 |
0.307 (-0.285-0.652) |
0.362 (-0.380-0.705) |
0.125 |
|
Emmetropia (-0.50 D < SER < + 0.50 D) (n = 33) |
||||
Center |
0.825 (0.525–0.925) |
0.867 (0.732–0.935) |
< 0.001 |
|
S1 |
0.910 (0.809–0.957) |
0.919 (0.837–0.960) |
< 0.001 |
|
S2 |
0.913 (0.673–0.967) |
0.942 (0.882–0.971) |
< 0.001 |
|
I1 |
0.842 (0.629–0.927) |
0.869 (0.735–0.935) |
< 0.001 |
|
I2 |
0.874 (0.515–0.952) |
0.918 (0.834–0.960) |
< 0.001 |
|
T1 |
0.883 (0.750–0.944) |
0.895 (0.788–0.948) |
< 0.001 |
|
T2 |
0.781 (0.479-0.900) |
0.822 (0.641–0.912) |
< 0.001 |
|
N1 |
0.844 (0.275–0.946) |
0.912 (0.821–0.956) |
< 0.001 |
|
N2 |
0.737 (-0.133-0.913) |
0.877 (0.751–0.939) |
< 0.001 |
|
Hypertrophia (SER ≥ + 0.50 D) (n = 53) |
||||
Center |
0.855 (0.495–0.940) |
0.905 (0.836–0.945) |
< 0.001 |
|
S1 |
0.853 (0.746–0.915) |
0.858 (0.754–0.918) |
< 0.001 |
|
S2 |
0.888 (0.707–0.947) |
0.915 (0.853–0.951) |
< 0.001 |
|
I1 |
0.861 (0.747–0.922) |
0.873 (0.779–0.926) |
< 0.001 |
|
I2 |
0.895 (0.680–0.954) |
0.926 (0.872–0.957) |
< 0.001 |
|
T1 |
0.877 (0.784–0.930) |
0.884 (0.799–0.933) |
< 0.001 |
|
T2 |
0.851 (0.622–0.930) |
0.888 (0.805–0.935) |
< 0.001 |
|
N1 |
0.778 (0.570–0.880) |
0.808 (0.668–0.889) |
< 0.001 |
|
N2 |
0.635 (0.111–0.827) |
0.741 (0.552–0.851) |
< 0.001 |
|
SD: standard deviation; SD-OCT: spectral-domain optical coherence tomography; SS-OCT: swept-source optical coherence tomography; ICC: Intra-class correlation coefficient of two devices; Center: Central foveal; S1: Inner superior: S2: Outer superior; I1: inner inferior; I2: outer inferior; T1: Inner temporal; T2: Outer temporal; N1: Inner nasal; N2: Outer nasal. |
Conversion Equation (N = 48) |
Validation (N = 47) |
||||
---|---|---|---|---|---|
% of values within 10% of each other |
Absolute ICC (95%CI) |
P-value |
Difference in means |
||
Center |
SS-OCT = 35.261 + 0.810 x SD-OCT |
94.7% |
0.942 (0.898–0.967) |
< 0.001 |
3.57µm |
S1 |
SS-OCT = 60.149 + 0.739 x SD-OCT |
85.3% |
0.827 (0.707–0.899) |
< 0.001 |
2.78µm |
S2 |
SS-OCT = -7.057 + 0.968 x SD-OCT |
88.4% |
0.887 (0.806–0.935) |
< 0.001 |
0.44µm |
I1 |
SS-OCT = 46.750 + 0.783 x SD-OCT |
88.4% |
0.832 (0.718–0.903) |
< 0.001 |
1.57µm |
I2 |
SS-OCT = 15.353 + 0.881 x SD-OCT |
86.3% |
0.851 (0.747–0.914) |
< 0.001 |
2.06µm |
T1 |
SS-OCT = 57.869 + 0.768 x SD-OCT |
91.6% |
0.883 (0.800-0.933) |
< 0.001 |
1.87µm |
T2 |
SS-OCT = 41.405 + 0.818 x SD-OCT |
94.7% |
0.892 (0.814–0.938) |
< 0.001 |
2.33µm |
N1 |
SS-OCT = 20.396 + 0.827 x SD-OCT |
81.1% |
0.729 (0.563–0.839) |
< 0.001 |
5.53µm |
N2 |
SS-OCT = -11.260 + 0.837 x SD-OCT |
75.8% |
0.606 (0.387–0.760) |
< 0.001 |
0.68µm |
SS-OCT: swept-source optical coherence tomography; SD-OCT: spectral-domain optical coherence tomography; ICC: Intra-class correlation coefficient of two devices; Center: Central foveal; S1: Inner superior: S2: Outer superior; I1: inner inferior; I2: outer inferior; T1: Inner temporal; T2: Outer temporal; N1: Inner nasal; N2: Outer nasal. | |||||
95 participants were included in the conversion equation analysis after exclusion of outliers (N = 19). Conversion equations were derived based on random half sample of participants (N = 48) and validation were based the second half sample (N = 47). |
The study children were further categorized according to their post-cycloplegic spherical equivalent refraction (SER), which was defined as the spherical diopters (D) plus one-half cylindrical diopters. Subjects are classified into (1) myopia (SER≤-0.50 D) (n = 28); (2) emmetropia (-0.50 D < SER < + 0.50 D) (n = 33); (3) hypertrophia (SER ≥ + 0.50 D) (n = 53). As shown in both OCT devices, children with myopia (CFCT from SD-OCT: 244.83 ± 46.86µm; CFCT from SS-OCT: 227.36 ± 43.05µm) tended to have a thinner choroid than emmetropic children (CFCT from SD-OCT: 275.68 ± 60.87µm; CFCT from SS-OCT: 251.06 ± 50.27µm) while children with hypertrophia tended to have a thicker choroid (CFCT from SD-OCT: 286.73 ± 48.60µm; CFCT from SS-OCT: 265.26 ± 42.43µm). All subgroups showed statistically significant difference in measured CFCT between the two OCTs (Table 2). While the inter-device correlation coefficients were excellent in both hypertrophic (CFCT: 0.855, 95% CI: 0.495–0.940) (P < 0.001) and emmetropic eyes (CFCT: 0.825, 95% CI: 0.525–0.925) (P < 0.001), but moderate in myopic eyes (CFCT: 0.754, 95% CI: 0.454–0.888) (P < 0.001) (Table 3).
Our study shows generally satisfactory intra-device correlation coefficients of the widely commercially available SD-OCT and SS-OCT in normal Chinese children population, regardless of the refractive status. However, a statistically significant intra-device difference exists in the measurements. We also noted a marked superiority in the quality to visualise the choroid-sclera interface in SS-OCT (Fig. 2) compared to SD-OCT (Fig. 1), same as the finding previously reported in other studies. 9, 28
In a study of 35 healthy adult eyes, Matsuo et al 11 found a high intra-device correlation coefficient between SD-OCT and SS-OCT, which is consistent with our current study. Excellent correlation was also noted in other similar studies in normal adults. 10, 12, 15, 16 In contrast to our findings, Matsuo et al 11 and Ikuno et al 29 revealed a thicker measured choroidal thickness from SS-OCT than from SD-OCT, which were both manually determined. The differences may be attributed to the automated segmentation of SS-OCT in our study. Zafar et al 15 observed that the mean subfoveal choroidal thickness measured manually by SD-OCT and SS-OCT was greater than that automatically determined SS-OCT. They reported that a considerably higher choroid-sclera interface was identified by automatic software. Nevertheless, in our current study, automatically segmented images were further inspected and the border was manually adjusted if required, so the difference may not be fully explainable.
Another postulation from an earlier study by Michalewski et al 30 was the difference in the measurement method. Manually determined SD-OCT choroidal thickness measurements were based on focal measurements whereas that of automatically determined SS-OCT were calculated from the average thickness in a circular area with a diameter of 1000µm. As choroidal layer becomes thinner with the distance away from the fovea, focal measurements near fovea tend to overestimate the thickness and this may not be fully compensated by averaging focal measurements from different spots. 30
In a pediatric study on choroidal thickness measurements by SS-OCT, Xiong et al 31 reported subfoveal choroidal thickness of 272 ± 61µm, 283 ± 63µm and 269 ± 61µm for healthy children aged 6, 7 and 8 respectively. Similar studies done on Chinese school-age children also reported consistently thicker choroid than our subjects (251.84 ± 47.12µm). 26, 32–34 Of note, our previous study showed that Hong Kong has a higher prevalence of myopia than other Chinese cities. The notable difference could be attributed to the more intensive pre-school education, along with frequent near-work activities and lack of outdoor time due to crowded living places in Hong Kong compared to other regions of China. 22 Given the widely accepted postulation of the negative relationship between choroidal thickness and myopia, 26, 35 it is conceivable that the measured choroid is thinner in our study population.
In addition, we observed a statistically significant difference between the measurements obtained from two devices, which differs from previous study on healthy adult eyes. 10–12, 15, 16 Hanumunthadu et al 36 reported thicker large choroidal vessel wall and thinner medium vessel wall at subfoveal area in children eyes compared to adult eyes. A denser and high-flow vascular network was also found in children eyes. We believe that the age-related variation in choroidal vasculature, in particular thicker and more crowded vessels, may result in a higher tissue density. It may affect the accuracy in automatic detection of borders of choroidal layer in SS-OCT. Furthermore, lowered penetrance may affect the visualisation of the lower borders, predominantly in SD-OCT with shorter wavelength.
As for topographical variation, our results are consistent with previous paediatric studies showing thickest and thinnest choroid layers in temporal and nasal regions respectively. Similarly, parafoveal sectors are thicker than perifoveal sectors as observed in these studies. 26, 31–34 Our study further demonstrates poorer ICCs in nasal sectors compared to other sectors. We believe that the thin nasal choroidal layer results in a larger percentage error. Also, its proximity to the optic nerve may account for its higher variability compared to other regions. This also explains the less favourable validation data of conversion equations in nasal sectors.
In subgroup analysis, we observed similar ICCs in hypertrophic and emmetropic eyes, but poorer ICCs in myopic eyes in all regions. We believe that the inter-individual topographic variation of choroidal thickness during myopic growth may account for the difference. It is generally believed that choroid is stretched towards temporal direction when eyeball grows axially during myopic change. 37–40 Chui et al 41 proposed the ‘slippage’ hypothesis that difference in the rate and extent of stretching of tissue layers (retina, choroid and sclera) may result in slippage between ocular tissues during axial elongation. The ongoing process of scleral stretching gives rise to dynamic topographic asymmetry in choroidal thickness. In this regard, there is an inter-individual topographic variation of choroidal thickness depending on individual’s myopic progression. Unlike in the circular measurements by SS-OCT, focal measurements by SD-OCT may not truly reflect the variation and therefore results in a greater discrepancy.
Our study, to the best of our knowledge, is the first to compare the choroidal thickness measurements by two generations of OCT instruments in children. We directly compared measurements by different devices on the same subject.
Nevertheless, limitations of this study should be acknowledged. As mentioned above, SD-OCT choroidal thickness measurements were derived from focal measurements on different linear radial scans while that of SS-OCT were derived from regional measurements in a circular area. The discrepancy may be significant in children with greater topographical variation in choroidal thickness. Of note, most of these children are myopic, who are not the main focus of our current study.
In conclusion, we have shown a high consistency in choroidal thickness measurements between the two generations of OCT device in healthy paediatric eyes. With a satisfactory ICC, it is justifiable to directly compare two sets of measurements for monitoring choroidal thickness changes in population-based epidemiological research study. Nevertheless, it should be noted that choroidal thickness measured by SD-OCT was statistically significantly greater than that measured by SS-OCT, especially in myopic children. Therefore, it is not recommended to interchange the two OCT-results when assessing disease progression in individual patient in clinical practice.
This study was supported in part by CUHK Jockey Club Children Eye Care Programme; the General Research Fund (GRF), Research Grants Council, Hong Kong (14111515 and 14103419 [JCY]); the Health and Medical Research Fund (HMRF), Hong Kong (5160836 [LJC] and 07180826 [XJZ]); the Direct Grants of the Chinese University of Hong Kong (4054193 [LJC] and 4054121 & 4054199 [JCY]); the UBS Optimus Foundation Grant (8984 [JCY]); and the Centaline Myopia Fund [JCY]. The funding organization had no role in the design or conduct of this research.
No conflicting relationship exists for any author.
Concept and design: All authors.
Acquisition, analysis, or interpretation of data: All authors.
Statistical analysis: COL, XJZ, JCY.
Drafting of the manuscript: COL, XJZ, CYC, JCY.
Critical revision of the manuscript for important intellectual content: All authors.
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.