Effect of Guided Ahmed Glaucoma Valve Implantation on Corneal Endothelial Cells: A 2-Year Comparative Study

DOI: https://doi.org/10.21203/rs.3.rs-1807414/v1

Abstract

Purpose: To compare the effects of guided and non-guided Ahmed glaucoma valve (AGV) implantation on the corneal endothelium.

Methods: Medical records of patients who underwent AGV implantation in the anterior chamber (AC) were reviewed retrospectively. The eyes were divided into two groups depending on the use of a guidance technique with spatula and a 4-0 nylon intraluminal stent. Specular microscopy was performed to measure corneal endothelial cell density (ECD) loss after surgery, and the rate of ECD change was calculated. Tube parameters were measured using anterior segment optical coherence tomography (AS-OCT).

Results: The ECD loss during 2 years of follow-up was significantly lower in the guided AGV implantation (gAGV) group than in the non-guided implantation (ngAGV) group, and the rate of ECD change was -0.62 ± 1.23 and -1.42 ± 1.57 %/month in the gAGV and ngAGV groups, respectively (p = 0.003). The mean tube–cornea distance (TCD) and mean tube–cornea angle (TCA) were significantly greater in the gAGV group than in the ngAGV group. The frequency of tube repositioning within 2 years after surgery was 0 % in the gAGV group and 12.66 % in the ngAGV group (p = 0.005).

Conclusions: The use of a guidance technique can reduce corneal endothelial loss during the first 2 years after AGV implantation in the AC. The tube was positioned at a more distant and wider angle from the cornea in the eyes of the gAGV group, which may have contributed to the reduced need for tube repositioning to prevent corneal decompensation.

Introduction

The frequency of glaucoma drainage device (GDD) surgery in cases of medically uncontrolled glaucoma is increasing worldwide. According to the 2012 Tube Versus trabeculectomy (TVT) study, in comparison with trabeculectomy with mitomycin C, tube shunt surgery showed comparable intraocular pressure (IOP) reduction and requirement for additional glaucoma medications, as well as a higher success rate over 5 years of follow-up in eyes with previous ocular surgery.1

Although the effectiveness and safety of GDD surgery have been proven for decades, corneal decompensation is one of the most common long-term complications after GDD surgery, and it has devastating consequences. Variable rates of corneal decompensation (8%-19%) have been reported in previous studies.24 The TVT study showed that 16% of tube patients developed persistent corneal edema with 8% undergoing subsequent keratoplasty over 5 years of follow-up.5 These rates were twice as high as those in the trabeculectomy group, which implies that GDD surgery possesses a higher risk of corneal decompensation than trabeculectomy. The Ahmed Baerveldt Comparison study also found the corneal decompensation rate to be 20% during 5 years of follow-up.6 The cumulative probability of corneal decompensation after Ahmed glaucoma valve (AGV) implantation was 3.3% over 5 years in the study by Kim et al.,7 but the difference was not significant in comparison with control eyes after 2 years. However, a recent study by Beatson et al. reported that the cumulative probability of developing decompensation after GDD surgery at 3, 6, 9 years was 4.7%, 9.2%, and 14.8%, respectively, with the risk persisting over time.8

Recent studies on tube parameters found that a longer tube–cornea distance (TCD) and deeper tube–cornea angle (TCA) induced less endothelial cell density (ECD) loss after AGV implantation in the anterior chamber (AC), emphasizing the importance of accurately positioning the tube.9,10 However, to the best of our knowledge, no comparative studies have been conducted on the surgical methods that may be useful for tube positioning in the AC.

In this study, we introduced a guidance technique using a spatula and a 4 − 0 nylon intraluminal stent during tube insertion in the AC. This method enables direct identification of the lowest entry point in the angle and firmly supports the flexible tube from kinking and bending when the tube enters the AC. We compared the corneal ECD change, surgical outcomes, and frequency of tube repositioning between the guided AGV implantation (gAGV) and non-guided AGV implantation (ngAGV) groups during 2 years of follow-up to evaluate the effect of guided AGV implantation on the corneal endothelium.

Methods

This retrospective study was approved by the Institutional Review Board of our university hospital (IRB number: 2022-05-030-001) and was conducted in accordance with the tenets of the Declaration of Helsinki. The requirement for written informed consent was waived owing to the retrospective nature of the study.

We reviewed the medical records of patients with glaucoma who were unresponsive to maximal medical treatment or had intolerable complications of anti-glaucoma drugs who underwent AGV implantation between 2016 and 2021. Patients with at least six months of postoperative follow-up were included in this study. Exclusion criteria included previous corneal transplant, preexisting corneal diseases that could affect the corneal endothelium, preoperative ECD less than 1200 cells/mm2, previous AGV implantation in the same eye, and any previous intraocular surgery within 6 months. Patients who were unable to complete the examinations required for this study were also excluded. If patients underwent other intraocular surgeries during follow-up, data before surgery were collected.

Patient demographics and ocular characteristics were collected by reviewing medical records, including age, sex, glaucoma diagnosis, laterality, number of glaucoma medications, best-corrected visual acuity (BCVA), IOP, systemic diseases, previous intraocular surgery, central corneal thickness, axial length, and AC depth. Postoperative data included BCVA, IOP, anterior segment inflammation, the number of glaucoma medications, and surgical complications. In both groups, surgical times (in seconds) were measured from the start of using a spatula in the guided group and the start of using a 23-gauge needle in the non-guided group; surgical time measurements in both groups were stopped at the time of successful AC tube insertion.

Corneal endothelial evaluation was performed using a non-contact specular microscope (Topcon SP-3000P; Topcon Corp., Tokyo, Japan) in patients within 1 month before AGV implantation and repeated postoperatively until the last day of follow-up. The central corneal area was examined three times consecutively to obtain the average measurements, while the patients fixed their gaze at the target in the instrument. The manual center-dot method was used to measure the central corneal ECD, coefficient of variation (CV), and hexagonality, marking at least 50 contiguous endothelial cells.11

Anterior segment optical coherence tomography (AS-OCT) (Triton SS-OCT; Topcon Corp., Tokyo, Japan) was performed 3–6 months after surgery to determine tube location in the AC. Patients were asked to fixate on an internal fixation target. Images that lacked artifacts of eye motion and blinking were included in the analyses, and the best-quality images were obtained using methods similar to those described by Hau et al..12 The scanning axis was positioned along the tube such that the spatial relationship between the tube and other structures of the AC could be clearly visualized. The perpendicular TCD between the anterior tip of the tube and the cornea, the length of the tube (TL) from the point of entrance into the AC to the anterior tip of the tube, and the TCA between the posterior corneal surface and anterior surface of the tube were measured using a built-in ruler and angle indicator.

Repositioning of the AGV tube was performed if the postoperative ECD decreased by more than 50% of the preoperative ECD and reached less than 1000 cells/mm2. The previous tube was withdrawn from the AC and inserted into the ciliary sulcus (CS) along the newly created sclerotomy path by using a 23-gauge needle.

All surgeries were performed by a single surgeon (CKL), and all AGVs were FP-7. The AGV tube was placed in the AC in all cases by using a previously reported standard method.13 The method was gradually changed to include the guided approach since 2020.

Under sub-Tenon anesthesia, a fornix-based flap of the conjunctiva and Tenon’s capsule was created in the superior temporal or nasal quadrants. Two partial-thickness scleral flaps were made with approximately one-third to half of the total thickness of the sclera to stabilize the tube (Fig. 1A). The tube was flushed with a balanced salt solution to check patency. Partial ligation of the tube and 4 − 0 nylon intraluminal stent was performed at the proximal site using 8 − 0 vicyl, and the stent was withdrawn. The AGV was anchored between the superior rectus and lateral or medial rectus muscles using 7 − 0 nylon, with the anterior edge of the plate at least 8 mm posterior to the limbus. The tube was trimmed to an adequate length by using a bevel-up edge. In the ngAGV group, a tube was inserted into the AC along the scleral tunnel created 2–3 mm posterior to the limbus under the scleral flap by using a bent 23-gauge needle, followed by an ophthalmic viscoelastic device (OVD) injection. In the gAGV group, a spatula with a round tip was inserted through the nasal paracentesis site after the OVD injection, and the spatula tip was gently rubbed to mark the insertion site for the tube under the scleral flap to be visible from the outside of the eyeball (Fig. 1B). Notably, placing the spatula too low would capture the peripheral iris and show how it was dragged. A bent 23-gauge needle was inserted into the AC from the marking location and pre-cut 4 − 0 nylon was docked into the needle inside the AC (Fig. 1C). After 4 − 0 nylon was pulled outside the insertion site, it was placed into the trimmed tube, and the tube was gently inserted into the AC by using nylon as a guide (Fig. 1D). In both groups, the tube was covered with a scleral flap sutured using 10 − 0 nylon, and the conjunctiva and Tenon’s capsule were closed using 8 − 0 vicryl. Topical steroids and antibiotic eye drops were administered for 8 weeks after surgery.

All data were analyzed using SPSS version 24.0 (SPSS, Inc., Chicago, IL, USA). Demographic data and surgical results were compared between the two groups by using Student’s t-test for continuous variables and Pearson’s chi-square test or Fisher’s exact test for categorical variables. Tube parameters and postoperative corneal ECD measurements were compared between the two groups by using Student’s t-test. Paired t-tests were used to compare pre- and postoperative ECD. The rate of ECD change after AGV implantation was determined using linear regression, and the slope of each regression equation represented the rate of ECD change in %/month. The rates of ECD change between the two groups were compared using Student’s t-test. Statistical significance was set at P < 0.05.

Results

A total of 135 eyes from 135 patients were included in this study, with 56 and 79 eyes in the gAGV and ngAGV groups, respectively. Preoperative characteristics of the patients showed no significant difference between the two groups, including age, sex, laterality, BCVA, IOP, axial length, AC depth, central corneal thickness, number of glaucoma medications, glaucoma diagnosis, systemic diseases, previous intraocular surgery and total follow-up period (p > 0.05, Table 1). This study included combined cataract surgery and AGV implantation, i.e., triple surgery, and the percentage of triple surgery in the gAGV and ngAGV groups was 41.07% and 26.58%, respectively, with no significant intergroup difference (p = 0.077, Table 1).

Table 1

Baseline characteristics of the patients.

 

Total

n = 135

gAGV

n = 56

ngAGV

n = 79

p value*

Age (years)

69.30 ± 10.90

70.07 ± 10.34

68.76 ± 11.31

0.493

Male / Female (n)

90 / 45

37 / 19

53 / 26

0.888†

Right / Left (n)

70 / 65

30 / 26

40 / 39

0.740†

Triplea [n (%)]

44 (32.59)

23 (41.07)

21 (26.58)

0.077†

Follow-up period (months)

17.33 ± 5.99

15.84 ± 6.49

18.39 ± 5.41

0.100

Preoperative BCVA (LogMAR)

0.96 ± 0.90

0.84 ± 0.88

1.05 ± 0.90

0.181

Preoperative IOP (mmHg)

27.32 ± 8.77

26.43 ± 7.95

27.95 ± 9.30

0.322

Preoperative number of glaucoma medication (n)

3.57 ± 0.83

3.64 ± 0.82

3.52 ± 0.85

0.397

AL (mm)

23.83 ± 1.21

23.85 ± 1.01

23.81 ± 1.34

0.836

ACD (mm)

3.52 ± 0.78

3.49 ± 0.91

3.55 ± 0.68

0.663

CCT (µm)

534.03 ± 38.17

537.98 ± 34.76

531.23 ± 40.39

0.313

Glaucoma diagnosis [n (%)]

     

0.346‡

POAG

39 (28.89)

19 (33.93)

20 (25.32)

 

NVG

36 (26.67)

12 (21.43)

24 (30.38)

 

Pseudoexfoliation

26 (19.26)

13 (23.21)

13 (16.46)

 

Uveitis

8 (5.93)

1 (1.79)

7 (8.86)

 

PACG

8 (5.93)

4 (7.14)

4 (5.06)

 

Othersb

18 (13.33)

7 (12.50)

11 (13.92)

Systemic diseases [n (%)]

     

0.080‡

HTN

62 (45.93)

37 (66.07)

25 (31.65)

 

DM

44 (32.59)

17 (30.36)

27 (34.18)

 

Autoimmune

6 (4.44)

2 (3.57)

4 (5.06)

 

Previous intraocular surgery

[n (%)]

     

0.842†

Phaco /c PCL

83 (65.91)

29 (65.91)

54 (64.29)

 

TLE

13 (10.16)

4 (9.09)

9 (10.71)

 

PPV

17 (13.28)

7 (15.91)

10 (11.90)

 

Othersc

15 (11.72)

4 (9.09)

11 (13.10)

 
* = Results of Student’s t-test. † = Results of Pearson’s chi-square test. ‡ = Results of Fisher’s exact test. Continuous variables are presented as means with standardized deviations. Categorical variables are presented as frequencies and percentages.
a = Combined phacoemulsification with IOL implantation and Ahmed glaucoma valve implantation.
b = Other glaucoma diagnoses include angle recession glaucoma, steroid-induced glaucoma, and mixed-mechanism glaucoma.
c = Other intraocular surgeries including bleb needling, intravitreal injection, and laser iridotomy.
gAGV, guided Ahmed glaucoma valve implantation; ngAGV, non-guided Ahmed glaucoma valve implantation;
BCVA, best-corrected visual acuity; IOP, intraocular pressure; POAG, primary open angle glaucoma; NVG, neovascular glaucoma; PACG, primary angle closure glaucoma; HTN. hypertension; DM, diabetes mellitus; AL, axial length; ACD, anterior chamber depth; CCT, central corneal thickness; Phaco /c PCL, phacoemulsification with posterior chamber IOL implantation; TLE, trabeculectomy; PPV, pars plana vitrectomy.

The surgical results in both groups showed comparable postoperative BCVA, IOP at different time intervals, number of glaucoma medications, and number of topical carbonic anhydrase inhibitors (CAI) (p > 0.05, Table 2). Postoperative AC cells showed no significant difference between the two groups, and iridocorneal touch did not occur in either group. Hypotony and hyphema were the two most common complications, accounting for one-quarter of the postoperative complications. The percentages of these complications were not significantly different between the two groups. Other postoperative complications also showed comparable results between the two groups, except for flat AC, which required viscoelastic injection. The mean surgical time was higher in the gAGV group, but the difference was not statistically significant (Table 2). Repositioning of the AGV tube in the CS was required in 0 eyes in the gAGV group and in 10 eyes (12.66%) in the ngAGV group, showing a significant difference between the groups (p = 0.005, Table 2).

Table 2

Comparison of surgical results after guided and non-guided Ahmed glaucoma valve implantation.

 

Total

n = 135

gAGV

n = 56

ngAGV

n = 79

p value*

Postoperative BCVA (LogMAR)

0.97 ± 0.92

0.83 ± 0.88

1.08 ± 0.95

0.118

Postoperative IOP (mmHg)

       

12 months

12.41 ± 3.74

12.90 ± 4.36

12.21 ± 3.45

0.386

24 months

13.65 ± 4.47

14.75 ± 4.97

13.43 ± 4.37

0.355

Final visit

13.50 ± 4.45

14.05 ± 4.21

13.11 ± 4.61

0.228

Postoperative number of glaucoma medication (n)

1.56 ± 1.06

1.54 ± 1.04

1.58 ± 1.07

0.802

Postoperative number of topical CAI (n)

0.45 ± 0.50

0.52 ± 0.50

0.41 ± 0.49

0.197

Postoperative AC cella

       

At 1 day

1.01 ± 0.79

1.05 ± 0.70

0.97 ± 0.85

0.568

At 1 week

0.56 ± 0.87

0.45 ± 0.74

0.65 ± 0.95

0.191

At 1 month

0.04 ± 0.21

0.05 ± 0.23

0.04 ± 0.19

0.668

Postoperative complications ≤ 1 month [n (%)]

     

0.642‡

Hypotonyb

16 (11.85)

6 (4.44)

10 (12.66)

0.729†

Hyphema

19 (14.07)

6 (4.44)

13 (16.46)

0.354†

Flat ACc

7 (5.19)

0

7 (8.86)

0.021

CD

5 (3.70)

2 (1.48)

3 (3.80)

0.659‡

IOL related

1 (0.74)

0

1 (1.27)

0.585‡

Vitreous hemorrhage

4 (2.96)

1 (1.79)

3 (3.80)

0.448‡

Vitreous prolapse

4 (2.96)

1 (1.79)

3 (3.80)

0.448‡

Iris related

2 (1.48)

1 (1.79)

1 (1.27)

0.659‡

Iridocorneal touch (n)

0

0

0

 

Surgical time (s)

251.74 ± 192.10

267.68 ± 168.37

240.00 ± 208.22

0.428

Repositioning of tube [n (%)]

10 (7.41)

0 (0)

10 (12.66)

0.005

* = Results of Student’s t-test. † = Results of Pearson’s chi-square test. ‡ = Results of Fisher’s exact test. Continuous variables are presented as means with standardized deviations. Categorical variables are presented as frequencies and percentages.
a = measurement by the standardization of uveitis nomenclature (SUN) grading system.
b = intraocular pressure ≤ 5 mmHg by Goldmann applanation tonometry.
c = viscoelastic injection into the AC required.
gAGV, guided Ahmed glaucoma valve implantation; ngAGV, non-guided Ahmed glaucoma valve implantation;
BCVA, best-corrected visual acuity; IOP, intraocular pressure; CAI, carbonic anhydrase inhibitor; AC, anterior chamber; CD, choroidal detachment; IOL, intraocular lens.

Comparisons of postoperative corneal ECD showed statistically significant differences between the gAGV and ngAGV groups (Table 3). While the two groups showed no significant differences in preoperative corneal endothelial cell measurements, including ECD, CV, and hexagonality, the postoperative ECD at the final visit was 1949.39 ± 488.95 cells/mm2 in the gAGV group which was significantly higher than the corresponding value in the ngAGV group (1571.95 ± 708.82 cells/mm2; p = 0.001, Table 3). The percentage of postoperative ECD loss was 7.91 ± 11.72% in the gAGV group and 29.34 ± 22.20% in the ngAGV group (p < 0.0001, Table 3). The rate of ECD change per month was significantly lower in the gAGV group than in the ngAGV group (-0.62 ± 1.23% and − 1.42 ± 1.57%, respectively, p = 0.003, Table 3).

Table 3

Comparisons of preoperative and postoperative corneal endothelial cell measurements between the guided and non-guided Ahmed glaucoma valve implantation groups.

 

Total

n = 135

gAGV

n = 56

ngAGV

n = 79

p value*

Preoperative ECD (cells/mm2)

2140.61 ± 509.75

2110.98 ± 473.77

2161.61 ± 535.78

0.572

Preoperative CV

29.19 ± 10.26

27.54 ± 10.17

30.37 ± 10.23

0.115

Preoperative hexagonality (%)

31.42 ± 41.65

25.91 ± 39.86

35.19 ± 42.67

0.208

Postoperative ECD (cells/mm2)

1728.52 ± 652.25

< 0.0001§

1949.39 ± 488.95

< 0.0001§

1571.95 ± 708.82

< 0.0001§

0.001

Postoperative ECD lossa (%)

20.45 ± 21.35

7.91 ± 11.72

29.34 ± 22.20

0.000

Rate of monthly ECD changeb (%)

-1.08 ± 1.49

-0.62 ± 1.23

-1.42 ± 1.57

0.003

* = Results of Student’s t-test. § = Results of paired t-test. Values are presented as mean ± standard deviation.
a ECD loss (%) = (preoperative ECD – postoperative ECD)/preoperative ECD × 100.
b The rate of monthly ECD change (%) was calculated and averaged from the slope of each linear regression equation.
gAGV, guided Ahmed glaucoma valve implantation; ngAGV, non-guided Ahmed glaucoma valve implantation;
ECD, endothelial cell density; CV, coefficient of variation.

Figure 2 shows the mean percentage changes of residual corneal ECD over time after AGV implantation in the two groups. The loss of ECD showed linear patterns overall in both groups, with a regression coefficient (B) of -0.399% in the gAGV group and − 1.149% in the ngAGV group (p = 0.009 and < 0.0001, respectively).

The tube parameters measured by AS-OCT are listed in Table 4. Both groups had comparable TL, while the gAGV group had a longer TCD (1101.32 ± 376.99 mm) and greater TCA (32.33 ± 6.65 °), than the ngAGV group (863.35 ± 357.31 mm and 28.51 ± 5.91 °, respectively, p = 0.004 and p = 0.007, respectively).

Table 4

Comparison of tube parameters between the guided and non-guided Ahmed glaucoma valve implantation groups.

 

Total

n = 135

gAGV

n = 56

ngAGV

n = 79

p value*

TL (mm)

1656.53 ± 533.80

1702.76 ± 494.25

1603.38 ± 577.70

0.392

TCD (mm)

988.00 ± 384.59

1101.32 ± 376.99

863.35 ± 357.31

0.004

TCA (°)

30.53 ± 6.57

32.33 ± 6.65

28.51 ± 5.91

0.007

* = Results of Student’s t-test. Values are presented as mean ± standard deviation.
gAGV, guided Ahmed glaucoma valve implantation; ngAGV, non-guided Ahmed glaucoma valve implantation;
TL, tube length; TCD, tube–cornea distance; TCA, tube–cornea angle.

Discussion

In this retrospective study, we compared and analyzed the effect of AGV implantation with and without guided tube insertion into the AC on corneal ECD and surgical outcomes. Previous studies have reported corneal ECD loss after AGV implantation in the AC.14,15 Kim et al.14 reported a 10.5% reduction in the central ECD after 12 months, and Lee et al.15 reported a 15.3% and 18.6% reduction in average ECD at 12 and 24 months after surgery, respectively.

Various hypotheses have been proposed to explain the mechanism leading to corneal ECD reduction after tube shunt surgery. Mechanical factors include the foreign body effect of the silicone tube, progressive tube migration, peripheral anterior synechiae (PAS), and transient tube-corneal/uveal contact with blinking and rubbing.12,14,16−18 Jets of aqueous humor fluid through the tube occurring in sync with heartbeats could result in ECD loss near the tube.19 Chronic inflammation caused by the silicone tube and chronic trauma results in increased endothelial cell permeability and depletion of nutrients and oxygen, leading to corneal edema.19,20 Moreover, the aqueous humor protein concentration was increased 10-fold in a previous study, suggesting an impairment in the blood-aqueous barrier to allow oxidative, apoptotic, and inflammatory proteins to enter the AC and alter the aqueous humor environment.21,22

In addition to tube-related factors, other factors can influence corneal ECD, including age, toxicity from glaucoma medication, high preoperative IOP, longer duration of high IOP before surgery, higher number of previous intraocular surgeries, and history of uveitis.23 Differences in ECD reduction have been reported according to the type of glaucoma, wherein eyes with angle closure glaucoma and pseudoexfoliation glaucoma showed significantly decreased ECD.24,25

Efforts to minimize the reduction in corneal ECD after tube shunt surgery are ongoing. AGV tube insertion into the CS using the iris as a mechanical barrier has significantly lower rates of ECD change than tube insertion in the AC in previous studies.2628 However, tube implantation in the CS may be related to a higher incidence of postoperative hemorrhage, since the CS shows greater vascularization than the AC,29 as well as intraocular lens (IOL)-related complications and capsular bag stability and zonular weakness should also be considered, and the structural difficulty of tube insertion in the CS may be present in Asian eyes with relatively small axial lengths. Placing the tube on the par plana can minimize the effect of the tube on the corneal endothelium but is limited by the prerequisite that previous vitrectomy must be performed.

In our study, no significant differences were observed in the demographic data between the gAGV and ngAGV groups, indicating that the influence of nonsurgical factors such as age, preoperative IOP, or glaucoma types that affect corneal ECD loss can be excluded. Comparisons of surgical results also showed no significant differences in postoperative IOP and the number of glaucoma medications, including topical CAIs between the two groups, thereby excluding the influence of these factors on the corneal endothelium. We also assessed postoperative inflammation by grading AC cells up to 1 month and found that postoperative inflammation in the AC did not increase in the gAGV group in comparison with that in the ngAGV group.

Postoperative complications and surgical times were compared to ensure the safety and efficiency, respectively, of the guidance technique. Surgical time was measured in seconds only for the part where the surgical technique of the two groups was different; a spatula and a guiding stent were used in the gAGV group while the tube was directly inserted into the ngAGV group. Guided AGV implantation took an average of 30 s longer than simple tube insertion because the former required more steps; however, based on the standard deviation, the guided procedure could be completed without much deviation from the mean time. The surgical time was relatively short in many eyes in the ngAGV group, but the procedure took much longer in some eyes in which the tube was not inserted into the desired position, or the tube was bent and did not enter the AC well. Postoperative complications were not significantly different between the two groups except for flat AC, which might be caused by leakage of the sclerotomy site after multiple punctures to obtain the desired tube position in the ngAGV group. Therefore, guided AGV implantation can be considered a relatively safe surgical technique that requires little additional time and sometimes saves time by facilitating tube insertion.

A guidance technique was proposed with the goal of ideal positioning of the AGV tube in the AC as far as possible from the cornea and parallel to the iris. To assess the tube position, the tube parameters were measured and compared between the two groups. The TCD and TCA are both considered important tube parameters, but the TCD is not a fully independent variable since it is affected by the TL and TCA.9 In this study, the mean TLs in both groups were not significantly different; therefore, the influence of TL on TCD could be excluded.

Postoperative ECDs at the final visit were significantly lower than the corresponding preoperative value in eyes of the gAGV and ngAGV groups (p < 0.0001). However, the percentage of postoperative ECD loss was lower in the gAGV group, and the rate of ECD change was also lower in the gAGV group, which implied that less corneal endothelial damage occurred in the gAGV group within 2 years after surgery. The rate of ECD change with guided AGV implantation in our study appeared to be comparable to the rate of ECD change with tube insertion in the CS in previous studies (-0.36%/month in Kim et al.27 and − 0.72%/month in Zhang et al.28). However, the overall degrees and rates of ECD change in this study were relatively higher than those reported previously. One possible explanation is that approximately 30% of the eyes included in our study had concurrent phacoemulsification, although lens status was not associated with treatment failure in the TVT study.1

Changes in residual corneal ECD over time after AGV implantation showed a clear difference in the slope between the two groups, and the gap in the remaining ECD widened as the follow-up period increased. The transient rapid decline in ECD at 1 month and recovery at 3 months suggest that lost endothelium due to corneal injuries during surgery might be recovered by stem cells from a niche at the posterior limbus.30 The mean percentage loss in corneal ECD was more prominent during the first 12 months than after 12 months. In fact, the mean follow-up period until tube repositioning was 13.40 ± 6.15 months in the ngAGV group, indicating the importance of careful monitoring of the remaining ECD in the first 12 months after surgery. Eyes that underwent tube repositioning within 2 years in this study had a mean ECD loss of 54.42% and a mean rate of ECD change of 4.80%/month. These figures were similar to those in the eyes included in this study that developed corneal decompensation after 2 years, with a mean ECD loss of 50.30% and a mean rate of ECD change of 4.09%/month. This result suggests that our criteria for tube repositioning are effective in selecting eyes at risk of corneal decompensation for early management.

This study had some limitations due to its retrospective nature. First, the follow-up interval and specular microscopic examination interval were not the same among the patients. Linear regression analysis was used to calculate the rate of ECD change in each eye to compensate for this limitation, and at least three ECD measurements were performed in most eyes to obtain a reliable slope. Second, the mean follow-up period in the gAGV group was shorter than that in the ngAGV group, although the difference was not significant. As the surgical method was gradually changed from simple tube insertion to a guidance-based technique, the ngAGV group had a greater chance of undergoing long-term follow-up examinations. However, we tried to compensate for this limitation by strictly obeying the inclusion criteria and limiting the data to within two years. In addition, the mean follow-up period in the gAGV group was longer than that until tube repositioning. An advantage of the gradual change in the surgical method is that there was no selection bias between the two groups because patients who visited in the specific period underwent surgery using the same surgical method regardless of their age, sex, or glaucoma diagnosis. Therefore, we excluded cases one month after the first guided AGV implantation in consideration of the learning curve and selection bias. Third, the postoperative PAS score, which may affect the corneal endothelium was not assessed in this study. We examined the iridocorneal contact by slit-lamp examination but did not use gonioscopy. Fourth, the AGV tube may change its position and length over time in the AC; therefore, tube positioning assessment using AS-OCT may vary depending on the time of evaluation.31,32 In this study, AS-OCT was performed 3–6 months after surgery to access a relatively early tube position. Finally, as the mean time from GDD surgery to corneal decompensation was over 24 months in a previous study,8 longer follow-up periods would be necessary to further investigate the association with corneal decompensation.

In conclusion, in comparison with the conventional method, guided AGV implantation resulted in a lower corneal ECD loss and frequency of tip repositioning within 24 months. Guided AGV implantation was associated with less corneal ECD loss and a lower rate of postoperative ECD change in comparison with non-guided AGV implantation. Tube parameter analysis showed that the guidance technique could be used to position the tube further from the corneal endothelium at both distances and angles. The two groups showed no difference in the frequency of postoperative complications, and tube insertion was consistently completed within the mean surgical time. Thus, guided AGV implantation is a safe and time-efficient surgical technique that may contribute to the prevention of postoperative corneal decompensation when an AGV tube is inserted into the AC.

Declarations

Conflict of Interest: No conflicting relationship exists for any author.

Funding

This research was supported by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science and ICT (NRF-2017R1C1B5018279)  

Commercial relationships disclosures

C.K Lee, Norvartis(S), Samil (S), Hanlim(S), Lilly(S), ChonkunDang(S), J.H Chey, none 

References

  1. Gedde SJ, Schiffman JC, Feuer WJ, Herndon LW, Brandt JD, Budenz DL; Tube versus Trabeculectomy Study Group. Treatment outcomes in the Tube Versus Trabeculectomy (TVT) study after five years of follow-up. Am J Ophthalmol. 2012;153:789-803.
  2. Purtskhvanidze K, Saeger M, Treumer F, Roider J, Nölle B. Long-term results of glaucoma drainage device surgery. BMC Ophthalmol. 2019;19:14
  3. Garmany A, Knier C, Kung F, Mejia C, Sargent J, Jamali Dogahe S, et al. Rate of Corneal Grafting Post-Glaucoma Drainage Device Use in Pediatric and Adult Patients. J Glaucoma. 2021;30:347-51.
  4. Mandalos A, Sung V. Glaucoma drainage device surgery in children and adults: a comparative study of outcomes and complications. Graefes Arch Clin Exp Ophthalmol. 2017;255:1003-11.
  5. Gedde SJ, Herndon LW, Brandt JD, Budenz DL, Feuer WJ, Schiffman JC; Tube Versus Trabeculectomy Study Group. Postoperative complications in the Tube Versus Trabeculectomy (TVT) study during five years of follow-up. Am J Ophthalmol. 2012;153:804-14.
  6. Budenz DL, Feuer WJ, Barton K, Schiffman J, Costa VP, Godfrey DG, et al. Postoperative Complications in the Ahmed Baerveldt Comparison Study During Five Years of Follow-up. Am J Ophthalmol. 2016;163:75-82.
  7. Kim KN, Lee SB, Lee YH, Lee JJ, Lim HB, Kim CS. Changes in corneal endothelial cell density and the cumulative risk of corneal decompensation after Ahmed glaucoma valve implantation. Br J Ophthalmol. 2016;100:933-8.
  8. Beatson B, Wang J, Boland MV, Ramulu P, Schein O, Fliotsos MJ, et al. Corneal Edema and Keratoplasty: Risk Factors in Eyes With Previous Glaucoma Drainage Devices. Am J Ophthalmol. 2022;238:27-35.
  9. Lee HM, Kim KN, Park KS, Lee NH, Lee SB, Kim CS. Relationship between Tube Parameters and Corneal Endothelial Cell Damage after Ahmed Glaucoma Valve Implantation: A Comparative Study. J Clin Med. 2020;9:2546.
  10. Koo EB, Hou J, Han Y, Keenan JD, Stamper RL, Jeng BH. Effect of glaucoma tube shunt parameters on cornea endothelial cells in patients with Ahmed valve implants. Cornea. 2015;34:37-41.
  11. Hirst LW, Ferris FL 3rd, Stark WJ, Fleishman JA. Clinical specular microscopy. Invest Ophthalmol Vis Sci. 1980;19:2-4.
  12. Hau S, Scott A, Bunce C, Barton K. Corneal endothelial morphology in eyes implanted with anterior chamber aqueous shunts. Cornea. 2011;30:50-5.
  13. Lee CK, Ma KT, Hong YJ, Kim CY. Long-term clinical outcomes of Ahmed valve implantation in patients with refractory glaucoma. PLoS One. 2017;12:e0187533.
  14. Kim CS, Yim JH, Lee EK, Lee NH. Changes in corneal endothelial cell density and morphology after Ahmed glaucoma valve implantation during the first year of follow up. Clin Exp Ophthalmol. 2008;36:142-7.
  15. Lee EK, Yun YJ, Lee JE, Yim JH, Kim CS. Changes in corneal endothelial cells after Ahmed glaucoma valve implantation: 2-year follow-up. Am J Ophthalmol. 2009;148:361-7.
  16. Topouzis F, Coleman AL, Choplin N, Bethlem MM, Hill R, Yu F, et al. Follow-up of the original cohort with the Ahmed glaucoma valve implant. Am J Ophthalmol. 1999;128:198-204.
  17. Heuer DK, Budenz D, Coleman A. Aqueous shunt tube erosion. J Glaucoma. 2001;10:493-6.
  18. Kalinina Ayuso V, Scheerlinck LM, de Boer JH. The effect of an Ahmed glaucoma valve implant on corneal endothelial cell density in children with glaucoma secondary to uveitis. Am J Ophthalmol. 2013;155:530-5.
  19. McDermott ML, Swendris RP, Shin DH, Juzych MS, Cowden JW. Corneal endothelial cell counts after Molteno implantation. Am J Ophthalmol. 1993;115:93-6.
  20. Macdonald JM, Geroski DH, Edelhauser HF. Effect of inflammation on the corneal endothelial pump and barrier. Curr Eye Res. 1987;6:1125-32.
  21. Rosenfeld C, Price MO, Lai X, Witzmann FA, Price FW Jr. Distinctive and pervasive alterations in aqueous humor protein composition following different types of glaucoma surgery. Mol Vis. 2015;21:911-8.
  22. Anshu A, Price MO, Richardson MR, Segu ZM, Lai X, Yoder MC, et al. Alterations in the aqueous humor proteome in patients with a glaucoma shunt device. Mol Vis. 2011;17:1891-1900.
  23. Sinha S, Ganjei AY, Ustaoglu M, Syed ZA, Lee D, Myers JS, et al. Effect of shunt type on rates of tube-cornea touch and corneal decompensation after tube shunt surgery in uveitic glaucoma. Graefes Arch Clin Exp Ophthalmol. 2021;259:1587-95.
  24. Zimmermann N, Wünscher M, Schlötzer-Schrehardt U, Erb C. Corneal endothelial cell density and its correlation with the severity of pseudoexfoliation. Klin Monbl Augenheilkd. 2014;231:158-63.
  25. Sihota R, Lakshmaiah NC, Titiyal JS, Dada T, Agarwal HC. Corneal endothelial status in the subtypes of primary angle closure glaucoma. Clin Exp Ophthalmol. 2003;31:492-5.
  26. Bayer A, Önol M. Clinical outcomes of Ahmed glaucoma valve in anterior chamber versus ciliary sulcus. Eye (Lond). 2017;31:608-14.
  27. Kim JY, Lee JS, Lee T, Seo D, Choi W, Bae HW, et al. Corneal endothelial cell changes and surgical results after Ahmed glaucoma valve implantation: ciliary sulcus versus anterior chamber tube placement. Sci Rep. 2021;11:12986.
  28. Zhang Q, Liu Y, Thanapaisal S, Oatts J, Luo Y, Ying GS, et al. The Effect of Tube Location on Corneal Endothelial Cells in Patients with Ahmed Glaucoma Valve. Ophthalmology. 2021;128:218-26.
  29. Weiner A, Cohn AD, Balasubramaniam M, Weiner AJ. Glaucoma tube shunt implantation through the ciliary sulcus in pseudophakic eyes with high risk of corneal decompensation. J Glaucoma. 2010;19:405-11.
  30. Tan AN, Webers CA, Berendschot TT, de Brabander J, de Witte PM, Nuijts RM, et al. Corneal endothelial cell loss after Baerveldt glaucoma drainage device implantation in the anterior chamber. Acta Ophthalmol. 2017;95:91-6.
  31. Lopilly Park HY, Jung KI, Park CK. Serial intracameral visualization of the Ahmed glaucoma valve tube by anterior segment optical coherence tomography. Eye (Lond). 2012;26:1256-62.
  32. Mathew DJ, Anuradha A, Low SAW, Belkin A, Buys YM, Trope GE. Long-term Follow-up of Ahmed Glaucoma Valve Tube Position Changes. J Glaucoma. 2019;28:276-80.

Summary

What was known before

What this study adds