Comparison between FTIR and O-PTIR
In this section, we compare the characteristics of FTIR with O-PTIR spectra. As O-PTIR spectra, which are collected in reflection mode are comparable to FTIR spectra of thin films in transmission mode, an ATR correction was applied to the spectra of the standalone FTIR instrument collected in ATR mode. This accounted for the dependence of optical path length d on the wavelength λ. Details can be found in the 'Materials and Methods' section. IR spectra in Figure 1 are shown in the spectral range from 800 to 1800 cm-1. They were acquired with the standalone ATR-FTIR system (blue) and the O-PTIR instrument (red).
PE had its main IR bands near 1464 cm-1 (Figure 1a), resulting from the methylene deformation vibrations. The C-H groups were the only bonds with a permanent dipole initiating vibrations. Overall, the FTIR spectrum was in good agreement with the O-PTIR spectrum, though minor differences in the background and relative band intensities were observed.
Key IR bands of PP are related to methyl deformation vibrations near 1456 and 1375 cm-1 (Figure 1b). Weaker IR bands were detected in the lower wavenumber region between 800 and 1200 cm-1 that were assigned to C-C (808, 974 cm-1), and C-H (841, 999, 1167 cm-1). FTIR and O-PTIR spectra agreed very well.
Figure 1c shows the IR spectra of PVC. Methylene deformation vibrations of CH2 occurred at 1462 cm-1, similar to the IR spectrum of PE. The bands at 1074, 1276 and 1124 cm-1 correspond to C-O bands. The one near 1725 cm-1 related to C=O modes, which were unexpected in PVC as the main structure lacks such functional groups. However, they are common bands in PVC containing plasticizers [29] that can contribute up to 80% by weight in flexible tubes. The heavy chlorine atoms led to C-Cl vibrations usually observed in the wavenumber range below 800 cm-1. Typical HC-Cl bands were expected near 1250 cm-1 but overlapped with C-O bands of the plasticizers.
The IR spectra of PET in Figure 1d show variations in the relative intensities of C-O ester bands between 1050 and 1300 cm-1. The respective band for the C=O ester groups was found at 1726 cm-1 in the O-PTIR spectrum, while the band was shifted to 1716 cm-1 in the FTIR spectrum. This difference might be due to different crystallinities, as the band has been reported to be at 1725 cm-1 for amorphous PET and 1718 cm-1 for crystalline PET [30]. The C-C band of the phenyl ring at 1409 cm-1 occurred in both spectra at the same position. Although structurally different from PET, O-PTIR spectra of PVC and PET shared some similarities, e.g. bands near 1120 and 1280 cm-1, which resulted from the plasticization of PVC.
In Figure 1e, the IR bands of carbonyl bonds in PC near 1780 cm-1 were found at higher wavenumbers relative to carbonyl groups in PET and PMMA. The bands between 1161 and 1240 cm-1 originated from the ester groups of the molecules. Their shift towards wavenumbers resulted from the more ether-like character of the C-O bond. A unique spectral feature of PC was a band at 1504 cm-1, assigned to an aromatic ring vibration. The FTIR and O-PTIR spectra had a similar signature but varied band intensities and maxima.
Figure 1f shows the IR spectra of PS. Prominent IR bands at 1602 and 1492 cm-1 are related to aromatic ring vibrations and at 1452 cm-1 to methylene vibrations. The bands at 1492 and 1452 cm-1 were dipole-active only and thus, had no counterparts in the Raman spectrum (see Figure 2f). The band at 1028 cm-1 corresponded to the in-plane CH bending of the aromatic ring [31]. The spectra acquired with the FTIR and O-PTIR systems are in good agreement.
The IR bands of silicone were dominated by the Si-O stretching vibration between 1000 and 1200 cm-1 with a maximum near 1007 cm-1 and the symmetric Si-CH3 bending vibration at 1260 cm-1 (Figure 1g). Another band of Si-CH3 in the O-PTIR spectrum near 805 cm-1 was shifted towards 785 cm-1 for the FTIR instrument, and just the slope was evident. The band corresponds to the symmetric out-of-plane bending of Si-CH3, whereas the band of the antisymmetric bending vibration was found at 864 cm-1 [32]. The higher mass of the silicone atoms induced a shift of the main bands to lower wavenumbers compared to hydrocarbon polymers. The FTIR and the O-PTIR system spectra showed a mismatch in band intensities but only a slight spectral shift of the bands.
The C-O bond of the ester vibrations of PLA and their coupling effects in the polymer chain were associated with maxima at 1088, 1130 and 1184 cm-1 in Figure 1h. The bands located at 1375 and 1454 cm-1 originated from the methyl group. The carbonyl group also had IR bands near 1753 cm-1.
The IR spectra of PMMA are presented in Figure 1i. IR bands at 600, 812 and 1152 cm-1 were assigned to C-O vibrations in the ester group. Another band at 1000 cm-1 was assigned to a C-O-CH3 vibration of the ester group. Bands of methyl vibrations occur near 1450 cm-1. Similar to the carbonyl bands of PET and PC, an IR band of C=O stretch vibration of the carbonyl was found near 1727 cm-1. PMMA has similar functional groups to PLA.
Overall, both the FTIR and O-PTIR instruments conserved the position of the IR bands well, and the spectra are in good agreement. The quantitative analysis of their HQI confirmed the resemblance of the spectra (see section 'Hit quality index of the spectra'). Relative changes in intensity might occur due to local variations in the chemical composition. Due to the inter-system comparison, different sample positions have been probed.
Comparison between Raman and Raman option of O-PTIR
In comparison to IR spectroscopy, Raman scattering allowed detecting an extended wavenumber range below 800 cm-1 and down to 250 cm-1. In this section, we compare the spectral characteristics of Raman measurements (blue) with the Raman scattering collected simultaneously with O-PTIR (red) in Figure 2. Whereas the standalone Raman system performed an automatic intensity calibration, the spectra of the O-PTIR Raman option were manually calibrated to compensate for effects of different optical components such as the Cassegrain objective lens and to compare both Raman spectra properly. Between 1500 cm-1 and 1800 cm-1, the O-PTIR instrument had a lower collection efficiency than the standalone Raman system. The compensation through a larger calibration factor simultaneously enhanced the noise. Details are provided in the 'Materials and Methods' section.
Figure 2a shows the Raman spectra of PE. Bands due to CH2 deformations were found at 1295, 1418, 1441 cm-1, and a shoulder near 1460 cm-1. The latter band was evident at a similar position in the IR spectrum in Figure 1a. Additional bands in the Raman spectra originated from the C-C bonds' good polarizability, leading to vibrations near 1063 and 1129 cm-1. The relative band intensities between the O-PTIR system and the standalone Raman device differed slightly and showed a good agreement.
The main Raman bands of PP were located near 399, 810, 842, 1153, 1330, 1360, and 1460 cm-1 (Figure 2b). Most Raman bands were more intense than IR bands. Their intensities changed relative to the methyl-associated bands at 1371 and 1460 cm-1. Bands at 399, 810, 841 and 1330 cm-1 were assigned to various CH and CH2 deformation vibrations, and bands at 974 and 999 cm-1 to CH3 deformation vibrations. Bands at 1153 and 1168 cm-1 are related to CC stretch vibrations. The spectra of the standalone Raman and the O-PTIR instrument were in good agreement considering their relative intensities.
The prominent Raman bands of PVC are presented in Figure 2c. The heavy molecular mass of chlorine led to additional Raman bands of C-Cl vibrations in the lower wavenumber region at 636 and 695 cm-1. The bands around 1302 and 1437 cm-1 were assigned to the methylene groups. The carbonyl group was associated with the Raman bands near 1725 cm-1, which was consistent with the spectral contributions of plasticizers in IR spectra. In addition, Raman bands at 1040 and 1602 cm-1 were typical for the common plasticizer phthalate.
In Figure 2d, the spectra of PET show their main Raman bands at 1116 and 1287 cm-1 of C(O)-O vibrations in the ester group and C=O vibration of the carbonyl group at 1725 cm-1. The aromatic ring vibration at 1614 cm-1 was only present in the Raman spectrum (Figure 2d) and not in the IR spectrum (Figure 1d). The spectra of the Raman and the O-PTIR instrument coincided well.
The Raman spectra of PC are presented in Figure 2e. The C-O-C vibrations in ester groups observed in IR spectra have Raman counterparts at 1112, 1180 and 1236 cm-1. The most intense Raman band at 889 cm-1 was assigned to an O-C(O)-C stretching vibration, which gave only a weak IR signal. Conversely, the Raman band at 1780 cm-1 of the C=O stretching vibration was weak but very intense in the IR spectrum in Figure 1e. Vibrations of methyl groups showed relatively weak Raman bands near 1450 cm-1. The Raman band near 1604 cm-1 is due to the carbon-carbon vibration of the aromatic rings. Raman bands at 637 cm-1 (phenyl ring) and 706 cm-1 (CH out-of-plane bending vibration) were also typical of the aromatic moiety.
The Raman spectrum of PS (Figure 2f) was dominated by vibrational bands of the aromatic ring at 621 cm-1 (ring deformation mode), 795 cm-1 (CH out-of-plane deformation), 1002 cm-1 (symmetric ring breathing mode), 1032 cm-1 (CH in-plane deformation) and 1603 cm-1 (ring-skeletal stretch). Both Raman spectra were in good agreement. They show the distinctive narrow band characteristics of PS, which is recommended as a Raman standard for instrument calibration [33]. It was also the first polymer to produce a Raman spectrum [34]. Thus, it underlined the precision of both instruments and the performed intensity calibration, as the spectra overlay almost perfectly.
Main Raman bands of silicone (Figure 2g) were located at 490 cm-1 (symmetric Si-O stretch vibration) of the Si-O-Si chain and 710 cm-1 (symmetric C-Si-C stretch vibration) and 789 cm-1 (antisymmetric C-Si-C stretch vibration) of the CH3-Si-CH3 moieties. Further bands at 861, 1261 and 1413 cm-1 were assigned to CH3 vibrations. Due to the larger atomic mass of Si, the Si-related bands occurred in the lower wavenumber range. Most Raman bands had smaller bandwidths than IR bands.
In Figure 2h, the most intense band of PLA at 874 cm-1 was assigned to a C-COO vibration. A related C-CO vibration showed a band at 408 cm-1 [35]. In contrast to IR spectra, C-O bond ester vibrations in PLA were weak, and the Raman bands with maxima at 1043 and 1128 cm-1 were associated with C-CH3 vibrations. The Raman bands located at 1296 and 1454 cm-1 originated from CH and CH3 deformation vibrations, respectively. The carbonyl group had Raman bands at 1771 cm-1.
Raman spectra of PMMA are presented in Figure 2i. The carbonyl bonds occurred at 1727 cm-1. The ester vibrations of the C-O group were associated with bands between 1145 and 1240 cm-1. Except for two additional bands between 1550 and 1650 cm-1 with the O-PTIR device, both Raman spectra agreed well. The PMMA has similar functional groups to PLA, which is also evident in the spectral comparison with Figure 2h and Figure 2i.
After intensity calibration, most Raman spectra acquired from the standalone Raman instrument coincide very well with the simultaneous Raman option of the O-PTIR instrument. Only minor differences in relative band intensities were observed. A quantitative comparison of the hit quality index is described in the following section.
Hit quality index of the spectra
The hit quality index (HQI) is defined to compare the resemblance of two vectors and is established as a common measure to match an acquired spectrum X with a library spectrum Y through [36]
Here, the HQI was used to quantify the agreement between the IR spectra acquired with the FTIR system after ATR correction and the IR spectra measured with the O-PTIR system and between Raman spectra of the standalone Raman instrument and the simultaneous Raman option of the O-PTIR instrument. The matrix in Figure 3a summarizes the HQI values of the IR spectra in Figure 1. The matrix in Figure 3b shows the HQI values of the Raman spectra presented in Figure 2. As pre-processing, an intensity normalization based on feature scaling was applied to the spectra. In addition, the spectra were interpolated using a linear interpolation method to match the wavenumber values. All spectra were compared to assess the suitability of the O-PTIR method for polymer identification. The diagonal line indicates the resemblance between matching polymer types.
Figure 3a summarizes the HQIs of the IR spectra. The diagonal line consists of values from 0.7 for silicone to 0.96 for PS, indicating sufficient to almost perfect matching. The highest off-diagonal HQI is for PVC to PET at 0.83, showing the highest chance for misidentification using the O-PTIR system. The presence of phthalates as a plasticizer in PVC and as a structural element in PET showed a high spectral resemblance. However, even though the calculated HQI was high, the most significant hit for PVC was its FTIR counter-spectrum with an HQI of 0.92. While only five plastic IR spectra have an HQI above 0.5 with a false partner, the great majority of HQIs for IR spectra lie below 0.5 and even below 0.2. Therefore, it can be concluded that IR spectra of MP particles acquired with the O-PTIR instrument are suitable for material identification and can be incorporated in analysis procedures based on standard library matching against ATR-corrected FTIR spectra.
The HQIs of the Raman spectra are summarized in Figure 3b. The diagonal line ranges from 0.72 for PMMA to 0.97 for PET and silicone. Most of the remaining off-diagonal HQI values were lower than 0.2, and only one was close to 0.5, which ruled out the possibility of misidentification. Thus, the simultaneous Raman option of the O-PTIR instrument is also suitable for polymer material identification.
Robustness of the HQI
The HQI is mathematically based on the scalar or dot product of two vectors (Eq. 1). If the vectors are identical or parallel to each, their resemblance leads to an HQI = 1.0. The minimal possible resemblance with an HQI = 0.0 requires the vectors to be orthogonal. Here, each spectrum is interpreted as a vector. Following this train of thought, the HQI will therefore be influenced by the position of maxima rather than their intensity magnitude.
In Figure 4, the nine IR spectra acquired with the O-PTIR for the same material are compared against themselves. Without a spectral shift, they obviously result in an HQI = 1.0. If one of the spectra is slightly shifted in wavenumber compared to its original, their HQI reduces. Small spectral shifts below 4 cm-1 that were observed for most IR bands in figure 1 affect the HQI values only marginally. With a spectral shift larger than 4 cm-1, correct identification is significantly reduced for some materials whereas other polymers still show HQI values above 0.75 even for a 9 cm-1 shift.
Two-dimensional identification
In MP research, material identification is key to assessing the concentration and abundance of MP particles in a sample. Usually, MPs have been exposed to environmental influences that lead to degradation, building up a corona of contaminations and altering the chemical polymer signature. Furthermore, spectral contributions of pigments and additives might overlap with the vibrational fingerprint of plastic. As a consequence, the vibrational signature of environmental particles often gives an ambiguous identification result. We propose a two-dimensional (2D) identification approach based on acquiring both IR and Raman spectra to improve the correct identification rates. Thus, MP particles are identified based on their complementary chemical composition, bringing higher certainty for library matching, as both the Raman and IR spectra can be considered. In the 2D identification approach, the HQI of the IR spectrum is compared with the HQI of the Raman spectrum resulting in a 2D-HQI graph. As illustrated in Figure 5, this approach allows clearer differentiation between false materials (blue) and the correct matching (green). This is especially useful for materials with high HQI for multiple materials, e.g. as observed for PVC and PET based on IR spectra in Figure 3a. Even particles with a high HQI for one of the methods, here 0.83 for the IR spectra of PVC/PET, were ruled out by Raman spectra of the 2D identification process.
Application to microplastic particles
Different MP particles from commercial sources and MP particles prepared by wet grinding from bulk material were investigated as test samples for the 2D-HQI identification approach. Figure 6 shows the 2D-HQI graphs for different MP samples. The figure also shows that not all particles could be identified. This usually happens due to insufficient signal quality or spectra of contaminations. In the figure, these spectra are marked as black dots. Most particles have a good HQI above 0.6, allowing IR and Raman to match the reference. Although HQI values of each spectrum were calculated for all nine polymers, that were presented in Figures 3 and 5, the 2D-HQI values in Figure 6 were only plotted for the expected polymer for simplicity. Microscope images of the sample and the recorded spectra can be taken from the digital supplement (Figures S2 – S4).
The simultaneously acquired Raman and IR spectra from self-made PS MPs match overall well (Figure 6a) with the reference spectra acquired from bulk material (Figure 1f, Figure 2f). The Raman spectra of the commercial red and blue PS beads (10 µm diameter, respectively) are highly distorted through the interfering fluorescence and/or are dominated by resonance-enhanced bands of the pigments. However, combined with the O-PTIR signal, the materials can still be identified (Figure 6a). The HQIs of smaller PS beads (6 µm diameter) shows the strength of the combined acquisition of IR and Raman spectra. Raman spectra are in good agreement with the reference, while the signal quality of the IR spectra is considerably lower.
The situation is reversed for PC (Figure 6b). HQI values of better Raman spectra are spread between 0.6 and 0.9, and HQI values of IR spectra are between 0.3 and 0.9 due to poorer signal quality. The Raman spectra of PP tend to have a lower signal-to-noise ratio, which also results in a lower HQI (Figure 6c). Still, most spectra could be matched successfully with the PP reference using simultaneously acquired IR spectra of higher signal-to-noise ratio, overcoming the disadvantage of using just one detection mechanism.