DOI: https://doi.org/10.21203/rs.2.11154/v1
Most trauma systems and traumatic spinal injury guidelines mandate spinal stabilization from the site of injury to a radiological confirmation or refutal of spinal injury [1, 2]. Vacuum mattresses have been advocated for patients in need of prehospital spinal stabilization [3, 4]. A vacuum mattress (VM) is a full-body sized device that is wrapped around the patient (Fig. 1). It contains small plastic beads that render the VM rigid upon activation (evacuation of air). A VM is thought to provide better stabilization and comfort, compared to the backboard [4, 5].
Norwegian national guidelines recommend the use of a VM for patients in need of prehospital spinal stabilization [6]. Therefore, Norwegian Emergency Department trauma teams will probably receive more patients where a VM is applied and activated.
Our previous study indicated that it is feasible to maintain spinal stabilization with certain VMs throughout a trauma CT examination [7]. Rather than removing the VM upon arrival at the hospital it may be desirable to maintain the VM for spinal stabilization throughout the resuscitation bay initial assessment. However, this necessitates that it is feasible to perform the initial assessment including chest and pelvic X-ray examinations with a VM in place. In order to perform a complete clinical assessment the VM must be temporarily deactivated.
We have not been able to identify prior studies of artefacts from modern VMs on resuscitation bay radiographs. The aim of our study was to investigate the effect of different VMs on standard resuscitation bay conventional radiography (CR) of chest and pelvis, especially regarding artefacts.
To cover all technical and practical issues, the study group included members covering the treatment chain from the trauma site (paramedic), emergency department (anaesthesiologist and Head of Trauma; ER nurse), trauma CR-acquisition (radiographer), interpretation (radiologists), and quality control (physicist).
We used an anthropomorphic whole body phantom (PBU-60, Kyoto Kagaku Co. LTD, Kyoto, Japan). The phantom resembles a small Caucasian adult patient with a weight of 50 kg and a height of 165 cm (Fig. 2). It contains a synthetic skeleton, soft tissue, liver, kidneys, and vessels (Fig. 2). We performed chest and pelvic radiography using three different vacuum mattresses (Table 1). We made sets of CR with the vacuum mattresses both moulded around the phantom and activated with the straps attached, then de-activated and lastly de-activated plus stretching the mattress in order to minimize folds. We repeated each set three times. Being the present protocol in our institution, we also did CR with a trauma transfer board. A reference image without any of the devices applied was acquired and made available during the evaluation for comparison. We used the trauma table (JI-2100, Jostedal Industrier, Jostedal, Norway) that is standard in our institution.
Table 1. List of immobilisation devices investigated
We performed the CR on a mobile X-ray machine (MobileDaRt Evolution, Shimadzu, Kyoto, Japan) with a flat panel detector (CXDI-55-C High Sensitivity Detector, Canon, Tokyo, Japan) with fixed tube potential and tube current (Chest 125 kV/4mAs, Pelvis 78kV/25mAs).
Two specialists in radiology independently reviewed the trauma radiographs regarding artefacts and were blinded to the specific type of device. The order of the CRs was randomized by a pseudo random number generator (Mersenne-Twister shuffle generator from the NumPy computer package [8]) assigning each examination with a random patient name and number.
The artefacts were categorised as no artefacts (category 1), artefacts not impeding clinical judgement (category 2), artefacts impeding clinical judgement (category 3), and artefacts rendering the investigation unsuitable for clinical judgement (category 4) (examples of artefacts in Fig. 3). Artefacts not visualised inside the phantom were classified as category 2. Artefacts lateral of the glenoid cavity, not affecting the lungs, where categorized as 2 in the chest radiographs.
Agreement between the two radiologists was assessed using Cohen’s kappa test statistic. The degree of observer agreement indicated by kappa values was interpreted as: [0–0.2] slight agreement; (0.2–0.4] fair agreement; (0.4–0.6] moderate agreement; (0.6– 0.8] substantial agreement; and (0.8–1.0] almost perfect agreement [9].
All radiographs were of good technical quality and fully diagnostic.
The two radiologists agreed on 46 of 56 evaluations regarding artefacts (Table 2). The inter-observer agreement was 0,82 with a kappa coefficient of 0,71 (95% confidence interval 0.55, 0.87). Apart from one single disagreed evaluation in VM 3, category 3 artefacts were seen exclusively with activated VMs. Category 3- and 4-artefacts were not seen with VM1 altogether.
Table 2
We found substantial agreement between the two radiologists regarding the presence and category of artefacts. Overall the artefacts were subtle, especially with the VM deactivated.
We performed CR with activated VMs although this is not presently applicable to our knowledge – CR will probably be executed during the initial assessment, which includes clinical examination necessitating exposure of the patient. Therefore, the VM must be deactivated during this phase.
Category 3 artefacts were with one exception seen exclusively in activated VMs. The single exception applies to VM3; however it was interpreted as category 2 after stretching the VM. This indicates that all three VMs could be used with resuscitation bay x-ray using present protocol at our institution. If the VM must be used activated, VM1 produced only category 2 artefacts when activated.
No added value was seen from stretching of the VM in the remaining results.
The CR with VMs was done without trauma transfer board, in accordance with the Norwegian guidelines. Our evaluation showed no artefacts from the trauma transfer board and artefacts with VMs de-activated are minor. Thus we have no indication that a combined use should result in major artefacts. This is, however, an assumption that should be further investigated.
We hypothesized that artefacts might be caused by mattress polymer surface folds, plastic valves and securing straps with plastic buckles as well as uneven distribution of the polystyrene beads. We chose to repeat each set of CR three times with mattress re-orientation, deflation and inflation between the sets to allow for potential redistribution of artefacts.
There are several limitations to our study. First, all radiographs were performed solely with the same mobile x-ray machine and trauma table. We cannot be certain that alternative setups would give similar results. Second, the phantom represents a small adult patient. As the mattress is uni-size we assume this will result in relatively more accentuated artefacts compared to radiographing with a larger patient. However, the opposite would be the case for children. Third, it was not possible to blind the interpreters completely to the generic type of device. However, they were blinded to manufacture type. Our radiologists are not familiar to immobilisation devices in daily practise and we have assumed they therefore do not carry any bias on device type. Fourth, our radiologists were not accustomed to interpreting radiographs of a phantom for artefacts so this may have affected the results.
Together with our previous findings [7] our present results indicate that some VMs may be used throughout the initial resuscitation bay assessment and CT examination. However, if instituting such a protocol the quality assurance should be rigid. In addition, such a protocol should perhaps be subject to a research project before general application.
CONCLUSION
Our results indicate that it is feasible to maintain some vacuum mattresses on trauma patients during standard resuscitation bay conventional radiography of the chest and pelvis. Two of the mattresses did not result in artefacts that impeded clinical judgement when de-activated.
VM: Vacuum mattress
CT: Computed tomography
CR: Conventional radiography
ER: Emergency room
Not applicable.
Not applicable.
DICOM images and acquired datasets will be made available from the corresponding author on reasonable request.
SA has served as an advisor for AB Germa, Sweden and Kohlbrat & Bunz GmbH, Austria but has gained no economic compensation or any other benefits. All other authors declare no competing interests, economic or otherwise.
This project was funded by Sørlandet Hospital.
PKH conceived the study. PJS, EA, MMB, AIM, CPGA, SA, and PKH designed the study. Data acquisition was performed by PJS, EA, AIM and CPGA. PJS and EA analysed and interpreted the data. PJS drafted the manuscript. All authors contributed to revision of the manuscript and approved the final version.
We wish to thank the University of Tromsø for lending us the phantom, radiologist Jon Bache Marthinsen, SSHF, for artefact assessments, and dr. Lars Webb for language assistance.
Table 1. List of immobilisation devices investigated
Device |
Name and vendor |
|
Vacuum mattress 1 |
RedVac VM0991X01, Kohlbrat & Bunz, Radstadt, Austria |
|
Vacuum mattress 2 |
Germa EasyFix, Ferno International, Wilmington, Ohio, USA |
|
Vacuum mattress 3 |
Germa All in One Continental 90 cm, Ferno International, Wilmington, Ohio, USA |
|
Trauma transfer board |
TraumaTransfer, Eson Comfort, Landeryd, Sweden |
Table 2. Assessment of artefacts category by two radiologists.
Device |
Status |
Score Thorax |
Score Pelvis |
|
|
Reviewer 1 / Reviewer 2 |
Reviewer 1 / Reviewer 2 |
No mattress or transfer board |
Reference |
Reference |
|
Transfer board |
1 / 1 |
1 / 1 |
|
RedVac |
Activated |
2 / 2 |
1 / 1 |
RedVac |
Activated |
2 / 2 |
1 / 2 |
RedVac |
Activated |
2 / 2 |
2 / 1 |
RedVac |
Deactivated |
2 / 2 |
1 / 1 |
RedVac |
Deactivated |
2 / 2 |
1 / 1 |
RedVac |
Deactivated |
2 / 2 |
1 / 1 |
RedVac |
Deactivated and stretched |
2 / 2 |
1 / 1 |
RedVac |
Deactivated and stretched |
2 / 2 |
1 / 1 |
RedVac |
Deactivated and stretched |
2 / 2 |
1 / 1 |
Germa All in One Continental 90 cm, Ferno |
Activated |
3 / 2 |
2 / 2 |
Germa All in One Continental 90 cm, Ferno |
Activated |
2 / 2 |
2 / 3 |
Germa All in One Continental 90 cm, Ferno |
Activated |
3 / 3 |
3 / 2 |
Germa All in One Continental 90 cm, Ferno |
Deactivated |
1 / 1 |
1 / 1 |
Germa All in One Continental 90 cm, Ferno |
Deactivated |
1 / 1 |
2 / 2 |
Germa All in One Continental 90 cm, Ferno |
Deactivated |
1 / 1 |
1 / 1 |
Germa All in One Continental 90 cm, Ferno |
Deactivated and stretched |
2 / 1 |
1 / 1 |
Germa All in One Continental 90 cm, Ferno |
Deactivated and stretched |
1 / 1 |
1 / 1 |
Germa All in One Continental 90 cm, Ferno |
Deactivated and stretched |
1 / 1 |
1 / 1 |
Germa EasyFix, Ferno |
Activated |
3 / 3 |
3 / 3 |
Germa EasyFix, Ferno |
Activated |
3 / 3 |
3 / 2 |
Germa EasyFix, Ferno |
Activated |
3 / 3 |
3 / 3 |
Germa EasyFix, Ferno |
Deactivated |
2 / 2 |
3 / 2 |
Germa EasyFix, Ferno |
Deactivated |
2 / 2 |
2 / 2 |
Germa EasyFix, Ferno |
Deactivated |
2 / 2 |
2 / 1 |
Germa EasyFix, Ferno |
Deactivated and stretched |
2 / 2 |
2 / 1 |
Germa EasyFix, Ferno |
Deactivated and stretched |
2 / 2 |
1 / 1 |
Germa EasyFix, Ferno |
Deactivated and stretched |
2 / 2 |
2 / 2 |
Legend table 2. The artefacts were categorized independently by the two radiologists as no artefacts (category 1), artefacts not hampering clinical judgment (category 2), artefacts hampering clinical judgment (category 3) and artefacts rendering the investigation unsuitable for clinical judgment (category 4).