Optimal insertion depth of central venous catheter through the right internal jugular vein, verified by transesophageal echocardiography

DOI: https://doi.org/10.21203/rs.3.rs-2079296/v1

Abstract

This prospective observational study investigated the optimal insertion depth of the central venous catheter through the right internal jugular vein using transesophageal echocardiography. After tracheal intubation, the anesthesiologist inserted a probe for esophageal echocardiography into the patient’s esophagus. The investigators placed the catheter tip 2 cm above the crista terminalis with echocardiography, which was defined as the optimal point. We measured the inserted length of the catheter. Pearson correlation tests were performed with the measured optimal depth and some patient parameters. We made a new formula for placing the catheter at the optimal position. A total of 89 subjects were enrolled in this trial. The correlation coefficient between the measured optimal depth and the patient’s parameters was highest for patient height (0.703, p < 0.001). We made a new formula of ‘height (cm)/10–1.5 cm’. The accuracy rate of this formula for the optimal zone was 71.9% (95% confidence interval; 62.4–81.4%), which was highest among the previous formulas or guidelines, when we compared. In conclusion, the central venous catheter tip was evaluated with transesophageal echocardiography, and we could make a new formula of ‘height (cm)/10–1.5’, which seemed to be better than other previous guidelines.

Introduction

For central venous catheterization, the right internal jugular vein is most commonly used because of the advantages/disadvantages of each route 1,2. On the other hand, it is well known that there are fatal complications due to the inadequate depth of central catheters 36. Therefore, numerous studies have been conducted to find the optimal depth for central venous catheterization through the right internal jugular vein. For the use of the right jugular vein as the route of central venous catheterization, some previous studies have presented some guidelines regarding the ideal depth on the basis of patient height 710. Other guidelines recommend that clinicians check the depth of the central catheter inserted, with the use of the carina on X-ray as a landmark 11. However, there has been no consensus regarding the ideal depth 12. We think that the disagreement surrounding the ideal depth is due to the indirect measurement of the ideal depth and the different methods used for measurements in each study. For example, authors in previous studies have investigated the ideal depth by various methods, such as calculating lengths using anatomical landmarks 1315, radiologic findings 11,16−19, cardiogenic electrical findings 20,21, or patient height 79. We believe that the results of these previous studies may have been inconsistent because the previous studies used indirect methods such as simple chest radiographs.

On the other hand, until now, the most relevant factor for the ideal insertion depth in central venous catheterization in many studies and clinical practice has seemed to be considered as patient height 79. If there was a highly accurate formula to estimate the optimal depth on the basis of patient height, it would be a great help in actual clinical practice without image analysis.

It may be considered an ideal method to measure the optimal insertion depth of the central venous catheter while observing the catheter tip with transesophageal echocardiography (TEE) in real-time practice. However, there are absolute limitations for the routine use of TEE in central venous catheterization 22. The purpose of this study was to measure the optimal insertion depth of the central venous catheter through the right internal jugular vein using TEE in cardiac surgeries and to derive a simple meaningful formula for prediction on the basis of the measured optimal depth and patient parameters mostly correlated with the depth. The secondary aim was to compare a new simple formula for the optimal depth from our data with some guidelines introduced in previous studies.

Methods

This prospective, observational, clinical trial was approved by the Institutional Review Board of Dongsan Medical Center (no. 2016-09-009). The study protocol was registered at clinicaltrials.gov (NCT03116724; https://register.clinicaltrials.gov/prs/app/action/SelectProtocol?sid=S0006WY1&selectaction=Edit&uid=U00032RV&ts=2&cx=-8q8uk6, 17/04/2017) before enrolment of subjects. This study was performed in accordance with the Declaration of Helsinki 2013 and written informed consent was obtained from all patients. Patients were enrolled between April 2017 and August 2017. Adult patients (age > 18) scheduled for elective open-heart surgery, requiring central venous catheterization through the right internal jugular vein and TEE, were screened for eligibility in the study. The exclusion criteria were patients who had to receive central venous catheterization at another site, such as the left internal jugular vein, subclavian vein, or femoral vein, for any reason. After we obtained informed consent from the patients, we recorded the patients’ characteristics, including age, sex, height, weight, and body mass index (BMI).

First, we defined the optimal insertion depth of the central venous catheter (CVC) as 2 cm above the crista terminalis from a review of previous studies and clinical meaning. Additionally, we defined the optimal zone for the position of the tip of the CVC as 1 cm above the crista terminalis to 3 cm above it in the study. Anesthesia and surgical techniques were performed in a standardized manner regarding the routine practice in our institute during the trial. All patients arrived in the operating room without any premedication. Patient monitoring, including electrocardiography, peripheral oxygen saturation, noninvasive blood pressure, and bispectral index began before the induction of anesthesia. After indwelling the radial artery catheter under local anesthesia with 1% lidocaine injection, general anesthesia was induced with 0.15 mg/kg midazolam, and continuous intravenous infusion of remifentanil was performed using a target concentration infusion (TCI) system. After loss of patient consciousness, 0.8 mg/kg rocuronium was intravenously administered for muscle relaxation while the patients’ lungs were manually ventilated with 100% oxygen and sevoflurane. After tracheal intubation, mechanical ventilation was applied. Maintenance of anesthesia was provided with sevoflurane and continuous infusion of remifentanil. Next, a multi-plane probe (6VT-D/8.0–3.0 MHz, GE Healthcare Technologies, Wauwatosa, WI) for TEE was inserted into the patient’s esophagus at the mid-esophageal level for intraoperative cardiac monitoring, and a bicaval view was obtained with control of the probe.

For central venous catheterization through the right jugular vein, patients were placed in the 8° Trendelenburg position by tilting the operating table and the patient’s head was turned approximately 30–40° after a standard pillow was placed under the patient’s right shoulder. After sterile preparation and draping of the right jugular site, an investigator inserted an introducer needle into the sterilized area to puncture the right internal jugular vein under real-time ultrasound guidance with an ultrasound machine (Vivid™ S70, GE Healthcare Technologies, Wauwatosa, WI) equipped with a linear array probe (9L-D/10.0-2.4 MHz, GE Healthcare Technologies, Wauwatosa, WI). The puncture site was located at the level of the crease of the cricoid cartilage in all patients. Catheterization was performed with a 7 Fr. triple lumen catheter (Presep®, Edwards Lifescience, Irvine, CA) by the modified Seldinger maneuver after fresh dark blood was smoothly aspirated into the syringe, which was attached to the introducer needle. In the process of placing the central venous catheter through the right internal jugular vein while removing the guidewire, the catheter tip was first placed at the superior edge of the crista terminalis under a bicaval view of real-time transesophageal echocardiography 19,23. The TEE examination was performed by one board-certified cardiothoracic anesthesiologist. If the catheter tip was not clearly visible, the position of the tip was confirmed by injecting fluid into the distal lumen and observing the microbubble (Fig. 1) 19,24,25. The length of the inserted part of the catheter from the skin was checked when it was finally confirmed that the catheter tip was placed at the superior edge of the crista terminalis by TEE. Next, the catheter was drawn by 2 cm to move the catheter tip to the optimal position, which was defined as 2 cm above the crista terminals in our study. The practitioner fixed the catheter by anchoring it at the skin, and a sterile dressing was placed. The final insertion length was recorded as the optimal depth for each patient. Then, anesthesia and surgery proceeded in line with routine practices in our institute.

The primary end point was the optimal insertion depth, and we placed the catheter tip 2 cm above the superior edge of the crista terminalis using real-time TEE. The secondary endpoint was a new simple formula with the most correlated patient parameters to calculate the accuracy of the new formula and some previous guidelines for placing the tip of the inserted catheter within the newly proposed optimal zone (from 1 cm above to 3 cm above the edge of the crista terminalis) in the present study and to compare the accuracy rate between our new formula and the previous guidelines in terms of the placement of the central venous catheter within the optimal zone. Additionally, we evaluated the vertical distance from the catheter tip to the carina in the postoperative chest X-ray (Fig. 2).

When we reviewed the results of the previous study by Ahn et al. 19, the central venous catheter according to Peres’ formula was located in the optimal zone, which was newly proposed by us in the present study (from 3 cm above the upper margin of the crista terminalis to 1 cm above it), in approximately 57% of patients in their study. Considering that our new formula based on the data of the present study could guide the tip of the central catheter within the newly proposed optimal zone in a higher percentage by 20% than Peres’ formula in our study population, a total of 85 patients were required. Considering an ~ 5% drop-out rate, a total of 89 subjects were needed.

Statistical analyses

Categorical data are presented as numbers. Continuous data are presented as the mean and standard deviation (SD). Pearson’s correlation test was used to assess correlations between the real-measured optimal depth for the central venous catheter through the right internal jugular vein and the patient characteristics, including age, sex, height, weight, and BMI. In addition, we made a new simple formula for the optimal depth using the most correlated variable with the real-measured optimal depth. We calculated the predictability of a new formula for the placement of the central venous catheter within the optimal zone, which was newly proposed by us in our study population. Additionally, we calculated the predictability of some formulas or guidelines, including Peres’ formula. The previous formula and guidelines included ‘height(cm)/10’ by Peres 7, ‘height(cm)/10–1’ by Czepizak et al. 8, ‘height(cm)/10–1.3’ by Lum et al. 9, a fixed depth of 15 cm by Kim et al. 10, or ‘to the carina 11. We compared the predictability of the optimal depth between our new formula and the previous formula or guidelines including Peres’ formula, by using generalized estimating equation (GEE) analysis and a post- hoc test with Bonferroni correction. All data were analyzed using SPSS 21.0 (SPSS Inc. Chicago, IL, USA) and SAS 9.4 (SAS Institute Inc. Cary, NC, USA). A P-value less than 0.05 was considered statistically significant. The 95% confidence interval was calculated for adequate values.

Results

A total of 89 patients were enrolled in this trial. Central venous catheterization was successful with the study protocol in all patients. There were no complications in the study. Table 1 shows the patient characteristics. The optimal depth of the catheter to the optimal point, which was 2 cm above the crista terminalis, was 14.5 ± 1.6 cm (95% confidence interval; 14.2 cm to 14.8 cm) under real-time ultrasound transesophageal echocardiography. The distance from the ideal depth point to the carina on postoperative chest X-ray was − 1.3 cm ± 1.6 cm (95% CI; -1.0 cm to -1.7 cm); a negative value means that the catheter tip was lower than the carina. The correlation coefficients between the optimal depth and patient parameters, including sex, age, height, weight, and BMI are presented in Table 1. Among the variables, patient height was most correlated with the optimal depth.

Table 1

Demographic characteristics for the study patients and correlation coefficient between the optimal depth of central venous catheter and the patient characteristics

Characteristics

(n = 89)

Coefficient

P-value

Sex (M/F, n)

51/38

-0.591

< 0.001

Age (y)

66.1 ± 14.5

-0.180

0.092

Height (cm)

160.4 ± 9.3

0.703

< 0.001

Weight (kg)

61.7 ± 11.3

0.550

< 0.001

BMI (kg/m2)

23.8 ± 3.0

0.166

0.121

Data except sex were presented as mean ± standard deviation. Sex was presented as numbers of patients. BMI: body mass index

We made a simple formula, ‘height (cm)/10–1.5 cm’, to best fit the optimal zone, which was defined in our study (1–3 cm above the upper margin of the crista terminalis), on the basis of the patient’s height, which was mostly correlated with the patient’s parameters in our results and the previous formulas using the patient’s height.

The accuracy rate of our formula for the optimal zone of our study was calculated as 71.9% (95% confidence interval; 62.4–81.4%).

The accuracy rate of our formula and the previous guidelines for the optimal zone in our study population are presented in Table 2. When we compared the accuracy rate with the generalized estimating equation test, there was a significant difference in guidelines, including our new formula (p < 0.001). When we compared the new formula and other previous guidelines with the Bonferroni correction, the accuracy rate of our new formula was significantly higher than that of the previous guidelines except the formula suggested by Phillip Lum et al. (Table 2).

Table 2

Accuracy rate of new formula and other previous guidelines for placing the central venous catheter in the optimal zone, which was defined between 3 cm above the crista terminalis and 1 cm above it.

Guidelines

Equation (cm)

Accuracy rate

(95% CI)

OR (95% CI)

P-value

New formula*

H/10–1.5

71.9% (62.4–81.4)

   

Peres’

H/10

34.8% (24.7–44.9)

4.79 (2.53, 9.09)

< 0.001

Phillip Lum’

H/10–1.3

67.4% (57.5–77.3)

1.24 (0.92, 1.66)

0.153

Czepizak’

H/10–1

58.4% (48.0-68.9)

1.82 (1.25, 2.66)

0.002

Kim’

15

50.6% (40.0-61.2)

2.50 (1.45, 4.32)

0.001

The carina

to the carina on the CXR

36.0% (25.8–46.1)

4.56 (2.38, 8.73)

< 0.001

CI; confidence interval, OR; odds ratio when the new formula*, which was suggested from our results, was compared with the others as reference, H; patient’s height (cm), CXR; chest X-ray after the surgeries, P-value was presented as the Bonferroni corrected P-value after generalized estimating equations (GEE) analysis.

Discussion

In our study, we verified that the patient’s height was the parameter most correlated with the optimal depth. In addition, we showed that the new formula of ‘height (cm)/10–1.5 cm’ was the best fit for locating the catheter tip within the optimal zone among various formulas or guidelines introduced in previous studies.

Many previous studies have investigated the optimal depth of central venous catheters through the right internal jugular vein. However, there remains no consensus regarding the depth 12. The reason may be the methodological differences between previous studies, the differences in the definition of the optimal position or optimal zone of the catheter tip, or racial differences in regards to human anatomy.

There were some different methods used to determine the optimal insertion depth for central venous catheter tip placement in previous studies. These methods included measuring the imaginary optimal depth on chest X-ray along the predicted route, measuring the distance between some points on a patient’s skin surface, and measuring the inserted guidewire length on monitored electric signals, which were biphasic rhythms. However, a limitation of these studies may be the possibility that the catheter tip was not exactly located at the defined optimal point in their studies because they indirectly measured the length. We placed the central venous catheter via a withdrawal of 2 cm after directly observing that the catheter tip was at the crista terminalis with TEE.

We did not present a new formula for the optimal depth on the basis of real-time TEE measurements but on the patient’s height. After a review of previous studies on this issue, we predicted that the patient’s height would be a main correlating factor for the optimal depth 79. We hoped that the new formula from our data would be simple for use in every clinical situation. Some previous studies used some measures from chest X-ray, which seems to be very simple to use. However, central venous catheterization should be performed abruptly without chest X-ray in some emergent situations. Additionally, chest X-ray during pre-anesthesia evaluation is not mandatory in all healthy and young patients in some countries.

Another possible reason for the controversy in determining the optimal depth may be the difference in the defined optimal position or optimal zone for the placement of the central catheter tip in each study. The US Food and Drug Administration recommends that the catheter should not be located in or allowed to migrate into the heart 12. Some previous studies showed that the mean SVC length was approximately 7 cm 10,26. Some previous studies recommended that the catheter tip should be above the pericardial reflection 11,27. However, the upper margin of the pericardial reflection could be up to 5 cm above the SVC-RA junction in some people 11, and catheter tip placement more than 4 cm above the SVC-RA junction can lead to catheter malfunction 28. In a recent study, moreover, the authors argued that free-floating catheter tips should be more important than placement above the pericardial reflection 19. Finally, some previous studies suggested that the optimal position of the catheter tip should be 2 cm above the atrio-caval junction 5,29,30. From this review, we defined the optimal point of the catheter tip as 2 cm above the superior edge of the crista terminalis, and we defined the optimal zone as within 3 cm above to 1 cm above the edge of the crista terminalis with a safety margin of 1 cm from the optimal point.

It may be true that there are racial differences in human anatomy. We investigated the optimal depth of the central venous catheter through the right internal jugular vein in Korean only. To verify our results in many races or to determine a new simple formula or guideline to fit all humans regardless of race, multiple studies or international multi-center studies are needed.

There were also some studies comparing the accuracy of different guidelines from previous studies for the ideal depth of central venous catheterization due to this discrepancy 19,31,32. We also compared the accuracy of our new formula and that of the previous guidelines. For the optimal zone, which was defined in our study, the accuracy was higher with our new formula than with some previous guidelines except one study by Phillip Lum et al., which was similar to ours. We performed this trial with a sample size of 89. Considering that there are numerous performances of central venous catheterization, further studies with large sample sizes should be performed to verify whether our new formula is superior.

There are some limitations of our study. First, we defined the optimal zone as to 1–3 cm above the crista terminalis. This could not be absolute or based on any strong evidence. Even though the optimal point of our study could be below the pericardial reflection, we agreed with the concept of the study of Ahn et al. 19, that is, the catheter could float parallel to the vascular wall of the SVC if the catheter tip was placed at its lower part. However, we believe that the catheter tip should not be placed in the cardiac chamber. Therefore, we defined the optimal zone for the catheter tip similarly in our study. Additionally, we speculate that the reason why perforation of the SVC by the central catheter tip has rarely been reported or has not been reported in these days is due to the advancement of the material and the design of the catheter tip 33. Second, we calculated the sample size from the result of the study of Ahn et al. 19. Our sample size seemed to be small for applying our new formula in the general population, considering that there could be anatomical differences between persons and races.

In conclusion, we evaluated the optimal depth of the central venous catheter through the right jugular vein with real-time TEE. The optimal depth was most correlated with patient height, and we found a new formula of ‘height (cm)/10–1.5’ for the optimal depth, which was better than other guidelines in our study population.

Declarations

Author contributions statement

JML and JL designed the study. JHP and JL performed the study. YK, JML, and SO performed statistical analysis. The main manuscript was written and revised by JML and JL. All authors have read and approved the final manuscript.

Data availability statement

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Conflicts of Interest

The authors declare no conflicts of interest.

Disclosure of Funding Source

None 

References

  1. McGee, D. C. & Gould, M. K. Preventing complications of central venous catheterization. N Engl J Med 348, 1123-1133, doi:10.1056/NEJMra011883 (2003).
  2. Parienti, J. J. et al. Intravascular Complications of Central Venous Catheterization by Insertion Site. N Engl J Med 373, 1220-1229, doi:10.1056/NEJMoa1500964 (2015).
  3. Quiney, N. F. Sudden death after central venous cannulation. Can J Anaesth 41, 513-515, doi:10.1007/bf03011546 (1994).
  4. Collier, P. E., Blocker, S. H., Graff, D. M. & Doyle, P. Cardiac tamponade from central venous catheters. Am J Surg 176, 212-214, doi:10.1016/s0002-9610(98)00171-8 (1998).
  5. Booth, S. A., Norton, B. & Mulvey, D. A. Central venous catheterization and fatal cardiac tamponade. Br J Anaesth 87, 298-302, doi:10.1093/bja/87.2.298 (2001).
  6. Yavascan, O., Mir, S. & Tekguc, H. Supraventricular tachycardia following insertion of a central venous catheter. Saudi J Kidney Dis Transpl 20, 1061-1064 (2009).
  7. Peres, P. W. Positioning central venous catheters--a prospective survey. Anaesth Intensive Care 18, 536-539, doi:10.1177/0310057X9001800422 (1990).
  8. Czepizak, C. A., O'Callaghan, J. M. & Venus, B. Evaluation of formulas for optimal positioning of central venous catheters. Chest 107, 1662-1664, doi:10.1378/chest.107.6.1662 (1995).
  9. Lum, P. A New Formulae-Based Measurement Guide for Optimal Positioning of Central Venous Catheters. JAVA 9, 80-85 (2004).
  10. Kim, W. Y. et al. Optimal insertion depth of central venous catheters--is a formula required? A prospective cohort study. Injury 43, 38-41, doi:10.1016/j.injury.2011.02.007 (2012).
  11. Schuster, M., Nave, H., Piepenbrock, S., Pabst, R. & Panning, B. The carina as a landmark in central venous catheter placement. Br J Anaesth 85, 192-194 (2000).
  12. Vesely, T. M. Central venous catheter tip position: a continuing controversy. J Vasc Interv Radiol 14, 527-534, doi:10.1097/01.rvi.0000071097.76348.72 (2003).
  13. Botha, R., van Schoor, A. N., Boon, J. M., Becker, J. H. & Meiring, J. H. Anatomical considerations of the anterior approach for central venous catheter placement. Clin Anat 19, 101-105, doi:10.1002/ca.20240 (2006).
  14. Ezri, T. et al. Correct depth of insertion of right internal jugular central venous catheters based on external landmarks: avoiding the right atrium. J Cardiothorac Vasc Anesth 21, 497-501, doi:10.1053/j.jvca.2006.05.011 (2007).
  15. Bannon, M. P., Heller, S. F. & Rivera, M. Anatomic considerations for central venous cannulation. Risk Manag Healthc Policy 4, 27-39, doi:10.2147/RMHP.S10383 (2011).
  16. Aslamy, Z., Dewald, C. L. & Heffner, J. E. MRI of central venous anatomy: implications for central venous catheter insertion. Chest 114, 820-826, doi:10.1378/chest.114.3.820 (1998).
  17. Uchida, Y. et al. Optimal prediction of the central venous catheter insertion depth on a routine chest x-ray. Nutrition 27, 557-560, doi:10.1016/j.nut.2010.07.005 (2011).
  18. Chin, J. H., Lee, E. H., Kim, J. I. & Choi, I. C. Prediction of the optimal depth for superior vena cava cannulae with cardiac computed tomography during minimally invasive cardiac surgery: a prospective observational cohort study. BMC Anesthesiol 17, 56, doi:10.1186/s12871-017-0347-x (2017).
  19. Ahn, J. H. et al. Transoesophageal echocardiographic evaluation of central venous catheter positioning using Peres' formula or a radiological landmark-based approach: a prospective randomized single-centre study. Br J Anaesth 118, 215-222, doi:10.1093/bja/aew430 (2017).
  20. Chu, K. S. et al. Accurate central venous port-A catheter placement: intravenous electrocardiography and surface landmark techniques compared by using transesophageal echocardiography. Anesth Analg 98, 910-914, table of contents, doi:10.1213/01.ane.0000105865.94157.4c (2004).
  21. Gebhard, R. E. et al. The accuracy of electrocardiogram-controlled central line placement. Anesth Analg 104, 65-70, doi:10.1213/01.ane.0000250224.02440.fe (2007).
  22. Ender, J., Erdoes, G., Krohmer, E., Olthoff, D. & Mukherjee, C. Transesophageal echocardiography for verification of the position of the electrocardiographically-placed central venous catheter. J Cardiothorac Vasc Anesth 23, 457-461, doi:10.1053/j.jvca.2008.12.003 (2009).
  23. Andropoulos, D. B. et al. A controlled study of transesophageal echocardiography to guide central venous catheter placement in congenital heart surgery patients. Anesth Analg 89, 65-70, doi:10.1097/00000539-199907000-00012 (1999).
  24. Weekes, A. J. et al. Central vascular catheter placement evaluation using saline flush and bedside echocardiography. Acad Emerg Med 21, 65-72, doi:10.1111/acem.12283 (2014).
  25. Blans, M. J., Endeman, H. & Bosch, F. H. The use of ultrasound during and after central venous catheter insertion versus conventional chest X-ray after insertion of a central venous catheter. Neth J Med 74, 353-357 (2016).
  26. Caruso, L. J., Gravenstein, N., Layon, A. J., Peters, K. & Gabrielli, A. A better landmark for positioning a central venous catheter. J Clin Monit Comput 17, 331-334 (2002).
  27. Tempe, D. K. & Hasija, S. Quest to determine the ideal position of the central venous catheter tip. Br J Anaesth 118, 148-150, doi:10.1093/bja/aew443 (2017).
  28. Petersen, J., Delaney, J. H., Brakstad, M. T., Rowbotham, R. K. & Bagley, C. M., Jr. Silicone venous access devices positioned with their tips high in the superior vena cava are more likely to malfunction. Am J Surg 178, 38-41, doi:10.1016/s0002-9610(99)00124-5 (1999).
  29. Fletcher, S. J. & Bodenham, A. R. Safe placement of central venous catheters: where should the tip of the catheter lie? Br J Anaesth 85, 188-191, doi:10.1093/bja/85.2.188 (2000).
  30. Lee, J. H., Bahk, J. H., Ryu, H. G., Jung, C. W. & Jeon, Y. Comparison of the bedside central venous catheter placement techniques: landmark vs electrocardiogram guidance. Br J Anaesth 102, 662-666, doi:10.1093/bja/aep046 (2009).
  31. McGee, W. T. et al. Accurate placement of central venous catheters: a prospective, randomized, multicenter trial. Crit Care Med 21, 1118-1123, doi:10.1097/00003246-199308000-00008 (1993).
  32. Sharma, D., Singh, V. P., Malhotra, M. K. & Gupta, K. Optimum depth of central venous catheter - Comparision by pere's, landmark and endocavitory (atrial) ECG technique: A prospective study. Anesth Essays Res 7, 216-220, doi:10.4103/0259-1162.118966 (2013).
  33. Gravenstein, N. & Blackshear, R. H. In vitro evaluation of relative perforating potential of central venous catheters: comparison of materials, selected models, number of lumens, and angles of incidence to simulated membrane. J Clin Monit 7, 1-6, doi:10.1007/bf01617891 (1991).