Does the Expensive IOFix implant Improve Union Rates?

DOI: https://doi.org/10.21203/rs.3.rs-22714/v1

Abstract

BACKGROUND

First metatarso-phalangeal joint fusion is the current gold standard for severe hallux rigidus. Data regarding the union rate and the re-operation rate when IOFix (an Intra-osseous fixation device, Extremity medical, New Jersey, USA) is used for hallux rigidus fusion is limited but promising. The aim of this study was to review our outcomes with the IOFix implant.

METHODS

We have conducted a retrospective chart review, following the approval of the hospital IRB committee. Exclusion criteria included bilateral operations on the same patient, multiple surgeries, charcot foot or other structural foot abnormalities (except hallux valgus), rheumatoid arthritis and a recent foot trauma. We collected demographic data, physical examination documentation, functional score evaluations (AOFAS), and X-ray radiographic studies.

RESULTS

Thirty patients were included in the study. The mean age was 60.36 ± 9.12 (range 36 to 77) years, 18 (60%) female patients and 12 (40%) male. Fourteen (53.33%) were left side pathologies. The average follow up period was 36.2 ± 12.31 (range 12 to 54) months. Union was obtained in 28 (93.33%) patients, of whom none had requested a hardware removal due to a prominent hardware during a minimum of 2 year follow up period. The mean postoperative AOFAS score was 80.5 ± 10.87 (range 35 to 90). A more stringent inclusion criteria and fusion definitions would have led to an exclusion of two more patients and a dropout of two patients from the “fused” group, which would have led to a fusion rate of 85.71%.

CONCLUSIONS

This is the largest series of hallux rigidus patients that were operated with an IOFix device. The rates of fusion and hardware removal in MTPJ1 arthrodesis performed with an IOFix implant were found to be similar at most when compared to previously described rates that were obtained with other cheaper and more simple fixation devices.

Background

Hallux rigidus (HR) is a common foot pathology that is associated with degenerative changes of the first metatarso-phalangeal joint (MTPJ1) and causes a limitation in hallux dorsiflexion.[1, 2] To date the surgical options to treat HR include cheilectomy, excisional arthroplasty, interposition arthroplasty, phalangeal osteotomy, first metatarsal osteotomy, implant arthroplasty, and arthrodesis.[3] MTPJ1 fusion, which has been described in various techniques, allows a definitive resolution and is the current gold standard for severe MTPJ1 osteoarthritis.[1, 2] During arthrodesis the joint is prepared with flat or conical cuts, often in a ball and socket configuration which are performed with reamers in order to obtain optimal bone contact and stability. A few fixation devices were described in order to obtain additional stability in the fused joint: a lag screw, a dorsal plate, crossed screws, staples and an intraosseous device.[4] The union rate with various implants was found to be consistently around 91–100%[512] and the need for reoperation due to a prominent hardware was usually between 0–13%[412] (with the exception of 78% in one study[13]). These reports imply a relative success in obtaining union, but in some cases, as in a single screw implant or a dorsal plate, the secondary procedure rates seem to be too high.

The IOFIX (an Intra-osseous fixation device, Extremity medical, New Jersey, USA) is a fixed or variable angle intraosseous device that is consisted of an “X-post” that is inserted proximal and parallel to the joint, and a compression screw that passes through a hole in the post, crosses the joint and engages on the proximal phalanx. Data regarding the fusion rate and the re-operation rate when IOFIX is used is limited but promising.[4] In a recent biomechanical study[14] it was found that the intramedullary device demonstrated the highest initial compression force when compared to a plantar locking plate and a dorsal locking plate. Nevertheless, it was also found to be the most susceptible to failure. The authors concluded that further research with clinical data is necessary in order to further analyze the outcomes of this device. The aim of this study was to review the fusion rate of the IOFix and the removal rate following fusion due to prominent hardware based our experience.

Methods

We have conducted a retrospective chart review, following the approval of our hospital IRB committee (the full name will be revealed for the unblinded version of the manuscript). We included patients aged 18 to 99 who were operated for a moderate to severe MTPJ1 osteoarthritis (OA) related HR[2] in a single medical center by a single surgeon (name initials will be added in the un-blinded manuscript) between the years 2015–2017 and that had a pre-operative weight baring X-ray radiographs. Exclusion criteria were bilateral operations on the same patient, multiple foot surgeries, charcot foot or other structural foot abnormalities (except hallux valgus), rheumatoid arthritis and a recent foot trauma. We collected demographic data, physical examination documentation, pre-operative and the latest postoperative (36.2 ± 12.31 (range 12 to 54) months) functional score evaluations (AOFAS)[15]. X-ray radiographic studies were taken at the following postoperative time points: 6 weeks, 3-6-12 months, and at 24 months when applicable. The consecutive radiographs were used in order to evaluate the union rate.[6] When reoperations were conducted, the reasons that have led to these procedures, as well as the surgical records were studied.

A flow chart that demonstrates our treatment protocol is presented in Fig. 1. Each patient was classified clinically and radiographically into one of 3 categories: mild, moderate or severe HR.[2] We did not offer surgical treatment to mild OA patients. When OA was considered to be moderate, surgical treatment was advised. In these cases we have accepted the patients’ informed consent for both a cheilectomy procedure and an MTPJ1 fusion.[6] The specific procedure was then chosen during the operation and was based on the remnant cartilage state as was evaluated intraoperatively. If sufficient cartilage was seen, which was estimated as > 50% of the joint surface, a cheilectomy was performed. If the remnant cartilage was found to be insufficient or lacking (Fig. 2), or if the HR grade was pre-operatively classifies as severe Fig. 3), an MTPJ1 fusion was performed. Only patients who have underwent MTPJ1 fusion and have answered our inclusion criteria were included in this study.

The MTPJ1 fusion was conducted with the patient in a supine position, under an ankle block anaesthesia and an above ankle tourniquet. An image intensifier was used throughout the procedure. A standard dorsal incision was used. After the joint was revealed we used conical reamers and stabilized the joint with 1.6 K-wires. At this point the IOFix was inserted using the standard surgical technique.[16] Following surgery we restricted weight bearing to the heels only, using a firmed sole shoe and crutches. Patients were invited for a follow up examinations on postoperative weeks 2, 4, 6, 12, 24 and 52.

Union was defined both clinically and radiographically, in concordance with previous publications.[4] Clinically, we expected the patients to be able to fully bear weight on their feet without pain, and to deny pain when applying external force on the 1st MTPJ. Radiographically we searched for a complete callus formation or trabeculae on 3 out of 4 cortices. If at 6 months follow up no union was seen on an X-ray radiograph and there was a mild local discomfort, a 6 months therapy course with MELMAKTM, a low intensity pulsed ultrasound (LIPUS) device, (BTT Melmak Development & production GmbH. Raisting Germany) was indicated in order to encourage union.

Results

Thirty patients were included in the study. (Table 1) The mean age was 60.36 ± 9.12 (range 36 to 77) years, 18 (60%) female and 12 (40%) male. Fourteen (53.33%) had a left side pathology. The average follow up period was 36.2 ± 12.31 (range 12 to 54) months. Union was achieved in 28 (93.33%) patients, of whom none had requested a hardware removal due to a prominent hardware during a minimum of 2 year follow up period. The mean time to union was 8.46 ± 2.93 (range 6 to 20) weeks. The mean postoperative AOFAS scores were found to be 80.5 ± 10.87 (range 35 to 90) points. Due to delayed union signs, 2 patients were treated with a six months course of MELMAK™. Following this treatment union was obtained in both patients. Two patients reported a discomfort due to a MTPJ1 rigidity. In one patient the hardware could be palpated by the surgeon at a follow up visit, even though the patient could not feel it and was not symptomatic. One patient report a mild pain that was regarded to a post protrusion (Fig. 4). Two patients were lost to follow up at 6 months, but we do know that at their last visit union was seen on an X-ray radiograph and that at 2 years they did not undergo a repeat surgery at any of the national public health system hospitals or at the largest private medical care facility at our area.

Two patients had a non-union (Fig. 5). One was re-operated 8 months after his primary surgery due to a symptomatic nonunion that was resistant to a treatment with MELMAK™. We have performed a second fixation with plate and screws but union was still lacking. Eventually the patient was still mildly symptomatic and did not want a third surgery. The second patient presented with a non-symptomatic nonunion. Even at a 42 months postoperatively X-ray radiograph have lacked any signs of fusion. Despite the radiographic appearance we did not detect painful movements in the involved joint, and no further surgery was indicated. Except for these two patients, all patients were satisfied with their results.

A more stringent approach would have been to regard the 2 patients who were treated with MELMAK™ as “failures”, and to exclude the 2 patients who were lost to follow up (even though a fusion was defined for both). This would have led to a 28 patient cohort and a fusion rate of 85.71% (24 patients).

Discussion

In the current study we present a series of 30 patients who suffered from HR and underwent an MTPJ1 fusion with an IOFix device. Using this technique a very high fusion rate was achieved, although it wasn’t superior, and might have been even inferior to fusion rates with other implants. None of our patients had requested a hardware removal due to a prominent hardware during a minimum two years follow up. For our knowledge this is the largest series that was thus far published on HR patients who were operated by this implant. In a previous similar study on 21 patients by Singhal et al[4], the use of IOFix allowed a 95% union rate. They too did not report on patients who requested an elective hardware removal. In a former preliminary series of 12 patients who were operated by an IOFix implant, the union rate was reported to be 91.67%. These two studies along with the current indicate reproducible union rates.

In the united states HR with arthritic changes is diagnosed in 2.5% of the population that is older than 50 years.[1] The pathophysiology of the disease is unclear. There are reports of familial history,[6] but trauma,[17] improper shoe wear, [18] a tight Achilles tendon,[18] or an elevated 1st metatarsus[6] were also proposed as possible triggers for the disease. In a series of 114 HR patients almost four of every five patients have developed a bilateral disease, and 95% percent of the patients with a family history have eventually developed symptoms in the contralateral foot.[6] The most common presenting symptoms were found to be a local pain, and a decreased range of motion.[6] Physical examination often revealed a local tenderness, a dorsal bump, and a decreased range of motion. [6] A grind test exacerbated pain.[6] Standing radiographs can reveal degenerative changes that include metatarsal head flattening and widening, subchondral sclerosis, osteophytes and subchondral cysts, joint space narrowing, and eventually a joint destruction.[6]

On the basis of the aforementioned clinical and radiographical findings Coughlin and Shurnas[6] have published a grading system according to which a treatment scheme was proposed. In their protocol non-surgical treatment was offered only for those patients who did not have any radiographical findings (grade 0). We on the contrary have offered a primary non-surgical treatment to all of our patients as the first line of treatment for a period of three months. This treatment included Local anti-inflammatory drugs, activity modification (avoiding tip toe standing, avoiding use of flexible sole shoes), and insoles with a morton extension. The data on the patients who have had radiographic signs of HR and were satisfied with a non-surgical treatment is beyond the scope of this article. Coughlin and Shurnas offered a cheilectomy procedure to all grade 1–3 patients, and added an arthrodesis as an option for grade 3 patients. Grade 4 patients were treated by an arthrodesis. We have included grades 1–3 into the same “Moderate” group and have made the decision of whether to conduct a cheilectomy or an arthrodesis during operation according to the cartilage appearance. This was similar to the way Coughlin and Shurnas treated grade 3 patients. Severe cases, which were equivalent to Coughlin and Shurnas grade 4 were treated by an arthrodesis procedure using an IOFix implant. We found this simplification of their grading system to be more applicable to the daily practice and similarly useful when treatment decisions were done.

A few modalities have been used in order to fuse the MTPJ1. With the use of plate and screws (cost ~ 800 to 1200 Euro) in 233 feet, surgeons have obtained union in 230 (98%) joints.[7] Three (1.3%) of these 230 patients have suffered from a prominent hardware and underwent a secondary surgery for hardware removal. 26 (11.3%) patients have had minor complications (superficial wound infection, hematoma/seroma or mild wound dehiscence). With the use of staples (cost ~ 600 Euro per 1 staple. At least two screws are usually needed), union was achieved in 29 (96.7%) out of 30 feet.[12] None of the 29 implants was removed. When two parallel screws (cost ~ 200 Euro per 1 headless screw) were used in 60 feet a union rate of 100% was achieved and no hardware removal was reported.[8] Two crossed screws have led to a 93.3% union rate.[10] When only one intramedullary screw was used in 109 feet, union was achieved in 104 (95.4%) joints, but 85 (78%) of these patients have undergone a hardware removal due to a prominent hardware. This removal rate was exceptional, and alarmingly high. The last article was later criticized both for the questionable surgical technique and the authors’ determination for “fusion”.[11] Altogether, it seems that all fixation devices allowed similar fusion rates[11], and despite one report the removal rate was very low. Since the IOFix (cost ~ 2000 Euro) was more expensive, and did not allow superior results, its cost-effectiveness is questionable.

Biomechanically, the IOFix device was previously found to be superior over the crossed pins and the dorsal plate device.[9] In vitro it was shown to sustain higher loads, to be stiffer and to sustain a more narrow inter-fragmentary space.[9] In another biomechanical study[14] the IOFix was found to be superior in producing primary inter-fragmentary compression forces but inferior in failure susceptibility when compared to dorsal or plantar plates.

The main weakness of this study is its retrospective and non-comparative fashion. Two patients were lost to follow-up. Although this is the largest series of its kind thus far, it is still a comparatively small patient group. Therefore, we might have had too few patients than the “number needed to harm”, when the need for reoperation was studied.

Conclusions

The rates of fusion and hardware removal in MTPJ1 arthrodesis performed with an IOFix implant were found to be similar at most when compared to previously described fusion rates that were obtained with other cheaper and simpler fixation devices.

abbreviations

HR: Hallux rigidus; MTPJ1: First metatarsophalangeal joint; AOFAS: American orthopedic foot and ankle society;

Declarations

References

  1. Ho B, Baumhauer J. Hallux rigidus. EFORT Open Rev. [Internet]. 2017 [cited 2019 Sep 30];2:13–20. Available from: http://www.ncbi.nlm.nih.gov/pubmed/28607766
  2. Coughlin MJ, Shurnas PS. Hallux Rigidus: Demographics, Etiology, and Radiographic Assessment. Foot Ankle Int. [Internet]. 2003 [cited 2019 Sep 21];24:731–43. Available from: http://www.ncbi.nlm.nih.gov/pubmed/14587987
  3. Coughlin MJ, Shurnas PS. Hallux rigidus. Grading and long-term results of operative treatment. J. Bone Joint Surg. Am. [Internet]. 2003 [cited 2019 Oct 6];85:2072–88. Available from: http://www.ncbi.nlm.nih.gov/pubmed/14630834
  4. Singhal R, Kwaees T, Mohamed M, Argyropoulos M, Amarasinghe P, Toh EM. Result of IOFIX (Intra Osseous FIXation) device for first metatarsophalangeal joint arthrodesis: A single surgeon’s series. Foot Ankle Surg. [Internet]. 2018 [cited 2019 Sep 21];24:466–70. Available from: https://linkinghub.elsevier.com/retrieve/pii/S1268773117301066
  5. Mohammed R, Gadgil A. Molded arthrodesis of the hallux metatarsophalangeal joint using the crossed-screw technique: Surgical technique, results and functional outcomes. Foot Ankle Surg. [Internet]. 2012 [cited 2019 Sep 30];18:132–5. Available from: http://www.ncbi.nlm.nih.gov/pubmed/22444002
  6. Coughlin MJ, Shurnas PS. Hallux rigidus. Grading and long-term results of operative treatment. J. Bone Joint Surg. Am. [Internet]. 2003 [cited 2019 Sep 30];85:2072–88. Available from: http://www.ncbi.nlm.nih.gov/pubmed/14630834
  7. Bennett GL, Sabetta J. First Metatarsalphalangeal Joint Arthrodesis: Evaluation of Plate and Screw Fixation. Foot Ankle Int. [Internet]. 2009 [cited 2019 Oct 4];30:752–7. Available from: http://www.ncbi.nlm.nih.gov/pubmed/19735631
  8. Brodsky JW, Passmore RN, Pollo FE, Shabat S. Functional outcome of arthrodesis of the first metatarsophalangeal joint using parallel screw fixation. Foot ankle Int. [Internet]. 2005 [cited 2019 Oct 5];26:140–6. Available from: http://journals.sagepub.com/doi/10.1177/107110070502600205
  9. Knutsen AR, Fleming JF, Ebramzadeh E, Ho NC, Warganich T, Harris TG, et al. Biomechanical Comparison of Fixation Devices for First Metatarsocuneiform Joint Arthrodesis. Foot Ankle Spec. [Internet]. 2017 [cited 2019 Oct 5];10:322–8. Available from: http://www.ncbi.nlm.nih.gov/pubmed/27881829
  10. Asif M, Qasim SN, Kannan S, Bhatia M. A Consecutive Case Series of 166 First Metatarsophalangeal Joint Fusions Using a Combination of Cup and Cone Reamers and Crossed Cannulated Screws. J. Foot Ankle Surg. [Internet]. 2018 [cited 2019 Oct 5];57:462–5. Available from: http://www.ncbi.nlm.nih.gov/pubmed/29503133
  11. Treadwell JR. First Metatarsophalangeal Joint Arthrodesis; What is the Best Fixation Option? Clin. Podiatr. Med. Surg. [Internet]. 2013 [cited 2019 Sep 30];30:327–49. Available from: http://www.ncbi.nlm.nih.gov/pubmed/23827491
  12. Choudhary RK, Theruvil B, Taylor GR. First metatarsophalangeal joint arthrodesis: A new technique of internal fixation by using memory compression staples. J. Foot Ankle Surg. [Internet]. 2004 [cited 2019 Oct 4];43:312–7. Available from: http://www.ncbi.nlm.nih.gov/pubmed/15480407
  13. Wassink S, van den Oever M. Arthrodesis of the First Metatarsophalangeal Joint Using a Single Screw: Retrospective Analysis of 109 Feet. J. Foot Ankle Surg. [Internet]. 2009 [cited 2019 Sep 30];48:653–61. Available from: http://www.ncbi.nlm.nih.gov/pubmed/19857821
  14. Burchard R, Massa R, Soost C, Richter W, Dietrich G, Ohrndorf A, et al. Biomechanics of common fixation devices for first tarsometatarsal joint fusion-a comparative study with synthetic bones. J. Orthop. Surg. Res. [Internet]. BioMed Central Ltd.; 2018 [cited 2020 Feb 26];13:176. Available from: https://josr-online.biomedcentral.com/articles/10.1186/s13018-018-0876-0
  15. Vosoughi AR, Roustaei N, Mahdaviazad H. American Orthopaedic Foot and Ankle Society ankle–hindfoot scale: A cross-cultural adaptation and validation study from Iran. Foot Ankle Surg. [Internet]. 2018 [cited 2019 Sep 13];24:219–23. Available from: http://www.ncbi.nlm.nih.gov/pubmed/29409212
  16. IO FiX Plus (Classic) | Extremity Medical [Internet]. [cited 2019 Sep 30]. Available from: https://www.extremitymedical.com/lower-extremity/io-fix/
  17. Bryant A, Tinley P, Singer K. A comparison of radiographic measurements in normal, hallux valgus, and hallux limitus feet. J. Foot Ankle Surg. [Internet]. [cited 2019 Oct 5];39:39–43. Available from: http://www.ncbi.nlm.nih.gov/pubmed/10658949
  18. BINGOLD AC, COLLINS DH. Hallux rigidus. J. Bone Joint Surg. Br. [Internet]. 1950 [cited 2019 Oct 5];32-B:214–22. Available from: http://www.ncbi.nlm.nih.gov/pubmed/15422020

Tables

Table 1: Patients who underwent a first metatarsophalangeal joint arthrodesis by an IOFix implant for hallux rigidus

No

Age range

Year of surgery

Follow up (months)

Side

Time to union (months)

AOFAS score

Complications

1

55-59

2016

36

L

5

77

Scar anaesthesia

2

55-59

2016

36

L

8

83

Mild plantar flexion position

3

65-69

2016

37

L


85


4

75-79

2016

40

R

4

87


5

60-64

2015

48

R


67

A mild local discomfort, but no pain or signs of non-union

6

65-69

2015

48

L

9

35

Non-union

7

65-69

2015

46

L

8

80


8

55-59

2015

45

R

7

83


9

70-74

2018

14

R

6

90


10

35-39

2015

54

L

20

87

Used Melmac for 8 months postoperatively

11

55-59

2017

24

R

-

83

Non-union

12

65-69

2015

52

L

12

80


13

50-54

2015

42

R

6

87


14

50-54

2015

40

L

9

75


15

60-64

2015

38

R

6

83


16

40-44

2017

25

L

9

87

Mild wound dehiscence

17

70-74

2018

14

L

8

69

A K-wire was inserted during the primary operation in order to add stability

18

55-59

2016

43

L

9

87


19

55-59

2017

28

R

8

75


20

55-59

2015

42

R

-

87

Non-symptomatic non-union

21

65-69

2016

38

R

10

87


22

65-69

2016

39

L

8

87


23

55-59

2015

53

L

6

85


24

70-74

2017

25

R

8

87


25

60-64

2016

38

R

10

69


26

65-69

2017

26

L

8

85


27

55-59

2015

52

R

8

87


28

60-64

2016

38

L

9

80


29

45-49

2016

12

R

10

Lost to follow up


30

65-69

2015

13

L

9

Lost to follow up


Mean

60.36±9.12 (range 36 to 77) years


36.2±12.31 (range 12 to 54) months


8.46±2.93 (range 6 to 20) months

80.5±10.87 (range 35 to 90) points


Table 1 Patients who underwent a first metatarsophalangeal joint arthrodesis by an IOFix implant for hallux rigidus. R: Right, L: Left.