PVC-Plasticizer DOP material as TMP for mammography and ultrasound dual-modality breast phantom has been evaluated using parameters of density, elasticity, x-ray attenuation, SoS, acoustic attenuation, impedance, and image. From Table 1, the density of the sample had close tissue equivalent to that of humans. For example, samples A and F were close to fat tissue with a difference of 1.16%. Meanwhile, B and F were close to the breast with a difference of 1.9 and 2.2%, respectively. Samples C, D, and E have good equivalence with the muscle tissues [24]. Based on a previous study, the density obtained was also similar to that of Jeong et al. 2017 [38], where a value of 1.02 (g cm− 3) was obtained for PVC-Plasticizer in phantom ultrasound applications. The difference in these values was due to the various concentration levels of PVC. A sample density equivalent to human tissue is the base element for estimating X-ray attenuation properties (mammography), SoS, and acoustic impedance. TMP density close to tissue can have close tissue equivalent to X-ray and ultrasound properties.
Meanwhile, Fig. 3 shows the linear attenuation coefficients of the X-ray samples that were compared with commercial breast ACR phantoms. The results between samples A, B, C, E, F, and the ACR phantoms gave promising results at 28–35 kV. However, for low voltages, the results were relatively different. From the literature data obtained by Heine et al. 2006 [39], the average X-ray linear attenuation coefficient of all samples has a difference of 1.29 ± 0.25 from that of the 50%-50% glandular-fat tissue [39]. This difference was due to variations in the quality of the X-ray tube from each manufacturer of mammography modalities. Samples A, B, C, E, and F have a close tissue equivalent with the X-ray attenuation properties of the ACR phantom. The value of the linear attenuation coefficient with a voltage obtained a quadratic relationship with a regression value of 0.957 as well as a t-test with a p-value of 0.431, which was greater compared to the set α-value of 0.05. This indicates that the fitting data does not show a significant difference with a regression value of 0.95. These results are consistent with Sato et al.2021 that the linear attenuation coefficient provides a non-linear relationship with X-ray energy [40].
Figures 4(a) and (b) show that SoS and Z obtained a trend of a constant-linear relationship to the frequency variation of 1–5 MHz with an average linear regression fitting of 0.643 for SoS and 0.675 for Z. Based on the t-test, the SoS and Z regression values obtained p-values of 0.027 and 0.036, respectively, which were smaller compared to the set α-value of 0.05. This indicates that the two data fitting results show a significant difference with a regression value of 0.95. Furthermore, the SoS and Z data do not show a linear response with frequency. These results are in line with Zell et al. 2007 [41] who stated that SoS has a very linear and constant dependence. The σ data showed a quadratic response to frequency variations with an average regression value of 0.978 and obtained a p-value equal to the set α-value of 0.05. The regression results were closer to the value of 1, indicating that the fitting were more accurate. The acoustic attenuation had an increasing trend as the frequency increased. These findings are similar to that of Villa et al. 2020 and Zell et al. 2007 [30, 41] that the attenuation coefficient increased proportionally with the frequency following the Power Law function.
Figures 4 and 5 compared the acoustic parameter measurements with human tissue data literature and other materials tissue-mimicking phantom as benchmarking, which were summarized in Table 4. The selection of this literature refers to the suitability of the frequency range used in the study. Based on SoS parameters, samples A, B, and F had close tissue equivalent to glandular with a difference of -10.4, 3.2, and − 0.8 m s− 1, respectively. The SoS of samples A, B, and F also have good equivalence with breast, soft tissue, and skin. The results also showed that D and E refer to stiff samples, such as trabecular bone at 1,886 m s− 1 [42]. When referring to the parameters of attenuation and acoustic impedance, sample A had an excellent equivalence to fat tissue with a difference of 0.03 (dB cm− 1 MHz− 1) and 0.07 (106 kg m− 2 s− 1), respectively. Sample B also was close to glandular tissue by referring to SoS values and acoustic attenuation. However, the acoustic attenuation parameter still requires further study. Sample A had an excellent close equivalence with the commercial phantom material Zerdine with a difference of 6.7 m s− 1, -0.03 (dB cm− 1 MHz− 1), and − 0.03 (106 kg m− 2 s− 1), respectively. The composition of the PVC-plasticizer material with agar and PVA materials competed with several studies’ results, which are summarized in Table 4.
The difference between TMP samples and human tissue can be caused by several factors, such as the molecular weight of the PVC-Plasticizer, which was low or exceeded during fabrication, thereby affecting X-ray attenuation and SoS, or requiring a scattering agent to increase the acoustic attenuation. Achieving tissue equivalence that satisfies both X-ray and ultrasonic waves is relatively challenging. However, the proposed TMP results were close to breast tissue, including A and B. This indicates that they satisfied the requirements as breast phantom material for dual-modality (mammography and ultrasound).
Figure 6 shows breast phantom images from samples A and B with lesions inserted for quality assurance and control. Cylindrical insertion lesions were both seen on mammography and ultrasound modalities. The gray level of breast tissue in the mammography modality obtained results that followed the breast ACR phantom. However, the ultrasound modality on layer A (PVC5%) obtained a gray level close to black. This was due to the lack of a scattering agent to increase the speckle of the ultrasound image. The adipose tissue pulse SoS factor was 1.27% lower than that of glandular tissue. This caused the contribution of the refractive effect to be small for the fat-to-glandular interface, hence, the pulse group's speed difference did not produce artifacts in the image. These findings are consistent with Carvalho et al. [27], where similar results were recorded. The results obtained have provided an important aspect in developing multimodal phantoms. This material has several other advantages, such as low-cost and reusability [37]. Improvements that can be carried out include adding scatter agents and completeness features to perform quality assurance as well as control of mammography and ultrasound modalities.