Hybrid pedicle screw and modified cortical bone trajectory technique in transforaminal lumbar interbody fusion at L4-L5 segment: finite element analysis

DOI: https://doi.org/10.21203/rs.3.rs-2485180/v1

Abstract

Background:: Investigate the biomechanical properties of the hybrid fixation technique with bilateral pedicle screw (BPS) and bilateral modified cortical bone trajectory screw (BMCS) in L4-L5 transforaminal lumbar interbody fusion (TLIF).

Methods: Three finite element (FE) models of the L1-S1 lumbar spine, including seven ligaments and facet joint cartilage were established according to the four human cadaveric lumbar specimens. BPS-BMCS (BPS at L4 and BMCS at L5), BMCS-BPS (BMCS at L4 and BPS at L5), BPS-BPS (BPS at L4 and L5), and BMCS-BMCS (BMCS at L4 and L5) were implanted into the L4-L5 segment of each FE model. The range of motion (ROM) of the L4-L5 segment, the von Mises stress of the fixation, intervertebral cage, and rod were compared under a 400-N compressive load with 7.5 N/m moments in flexion, extension, bending, and rotation.

Results: BPS-BMCS technique has the lowest ROM in extension and rotation, and BMCS-BMCS has the lowest ROM in flexion and lateral bending. The BMCS-BMCS group showed maximal cage stress in flexion and lateral bending, and the BPS-BPS group in extension and rotation. Compared to the BPS-BPS and BMCS-BMCS group, BPS-BMCS presented a lower risk of screw breakage and BMCS-BPS presented a lower risk of rod breakage.

Conclusion: The results of this study support that the use of the BPS-BMCS and BMCS-BPS techniques in TLIF surgery for offering the superior stability and a lower risk of cage subsidence and instrument-related complication.

1 Introduction

The number of spinal fusion surgery performed for degenerative diseases is gradually rising [1]. transforaminal lumbar interbody fusion (TLIF) has become a commonly used surgical approach in spinal fusion surgery Since Harms and Rolinger introduced it as an alternate technique In 1982 [2]. Previous studies have shown that TLIF has many advantages over posterior lumbar interbody fusion (PLIF), including preventing nerve root injury and allowing a large intervertebral cage through a small incision [3].

Traditional trajectory (TT) screw has been the gold standard in spine surgery [4]. It was widely used in TLIF and PLIF procedures. However, in recent years, due to complications such as decreased stability of the fixation unit caused by screw loosening, adjacent segment degeneration (ASD), and destruction of the facet joints, fixation in some patients cannot obtain a superior therapeutic effect and require revision, which is more common in elderly patients [5]. The main reason for screw loosening is that pedicle screws are mainly fixed by cancellous bone, and it is often compromised in patients with osteoporosis. Corticle bone trajectory (CBT) technique proposed by Santoni et al. [5] in 2009 could compensate for the shortcomings of p, but it failed to utilize cortical bone throughout its trajectory. To compensate for this shortcoming of CBT and TT, we have previously proposed a modified corticle bone trajectory (MCBT) technique (Fig. 1) and it's hybrid fixation (BPS-BMCS and BMCS-BPS) with TT [69]. The hybrid technique with MCBT and TT showed superior biomechanical properties in the L4-L5 segment without fusion in our previous study [9]. However, the hybrid technique in the TLIF model remains unclear. In this study, the biomechanical properties of the hybrid BPS-BMCS and BMCS-BPS techniques were investigated in the L4-5 TLIF model using finite element (FE) analysis.

2 Materials And Methods

2.1 Development of an intact L1-S1 finite element model

Three specimens were from the Anatomy Teaching-Research Office of Xinjiang Medical University. The mean age of them was 71, ranging from 64 to 77 years. Bone mineral density test indicating osteoporosis (BMD T<-0.25 SD), with the exclusion of previous lumbar spine surgery, infection, and tumor. High-resolution computed tomography (CT) data (AQUIRRON 16, PHILIPS, Netherlands) of L1-S1were collected and, Image data was stored in DICOM format. Import CT data of L1-S1 vertebrae into Mimics 20.0 (Materialize, Leuven, Belgium) software, and define the orientations. Select the "bone" threshold selection tool, select the appropriate range of vertebral body, fill in the vacancy, remove the redundant area with the brush tool, process the CT layer by layer, and reconstruct the L1-S1 vertebral body. Afterward, the smoothed 3D model was saved as an STL format and imported into 3-Matic 11.0 software (Materialise, Leuven, Belgium) for further processing. First, model surfaces were optimized. Second, the model was re-meshed to reduce the risk of distortion during the subsequent smoothing process, and the details of the model were refined by the sandpaper tool, with particular attention to the posterior structures of the vertebral body, such as the upper and lower facet joints and the isthmus. The intact L1-S1 FE model was composed of 5 lumbar vertebrae with the sacrum (Fig. 2A, B, C), facet joint cartilages at each segment (Fig. 2D), 7 ligaments (Fig. 2E), 5 intervertebral discs with endplates above and below the disc (Fig. 2F, G). The thickness of cortical bone (Fig. 2C) and endplate were defined as 0.5-1mm [10] and 1 mm [11] respectively. The nucleus pulposus was modeled as an incompressible fluid-filled cavity with very low stiffness occupying 44% of the disc volume [12], The contact between the facet cartilages was defined as "soft frictionless contact" with an initial gap of 0.5 mm [10]. The anterior longitudinal ligament (ALL), posterior longitudinal ligament (PLL), intertransverse ligament (ITL), ligamentum flavum (LF), capsular ligament (CL), interspinous ligament (ISL), and supraspinous ligament (SSL) were represented and assigned nonlinear material properties (Fig. 2E). At last, the meshed models were processed using ANSYS Workbench 19.1 (ANSYS, Inc., Canonsburg, PA, USA) to set the material properties [1317] (Table 1).

2.2 Model validation

The validation of the intact FE models includes two steps. First, the mesh convergence test was performed for 1 of the 3 intact FE models. For this FE model, 3 different mesh resolutions (mesh 1, mesh 2, and mesh 3) were generated consecutively. Among the three mesh resolutions, mesh 3 has the highest number of elements and nodes, and mesh 1 has the least. Ayturk et al. [18] demonstrated that axial rotation is the most sensitive to the mesh resolution of the FE model. The rotating conditions were simulated at a torque of 7.5 Nm and the von Mises stresses of different components were compared, and the meshes were considered converged when the difference between the predictions obtained by two successive mesh resolutions was less than 5%. In this study, the differences between the von Mises stresses of the three meshes were compared by the same method. Second, a 400 N compressive load and 7.5 N-m moments were applied to simulate flexion, extension, lateral bending, and rotation, and the ROM (range of motion) of each segment was compared with those of Yamamoto et al. [19], Shim et al. [20], Huang et al. [21], Lo et al. [22].

2.3 Construction of surgical models

TLIF procedures were performed by removing the facet joint at one side and part of the lamina. The TT screw with a diameter of 6.0 mm and length of 45 mm, the MCBT screw with a diameter of 5.0 mm and length of 40 mm, and the titanium-based rod measured 5.0 mm in diameter. Four different fixation groups were established: (1) BPS-BPS group, TT at the L4 and L5 (Figs. 3A, E) ; (2) BMCS-BMCS group, MCBT at the L4 and L5 (Figs. 2B, F); (3) BPS-BMCS group, TT at the L4, and MCBT at the L5 (Figs. 3C, G); and (4) BMCS-BPS group, MCBT at the L4, and TT at the L5 (Figs. 3D, H). Four FE models were established for each specimen. Finally, the constructed FE models were imported into ANSYS Workbench 19.1 (ANSYS, Inc., Canonsburg, PA, USA) for further biomechanical analysis.

2.4 Boundary and loading conditions

The sacrum was completely fixed and constrained to avoid displacement and rotation. Established a reference point in the center of the upper endplate of L1, and the contact between the reference point and endplate was using the “contact” constraint. 400 N compressive load and 7.5 Nm torque were applied for simulating flexion, extension, lateral bending, and rotation respectively (Fig. 4). The ROM of the L4-L5 segment, von Mises stress of the screw, intervertebral cage, and rod were recorded and discussed.

2.5 Statistical analysis

SPSS 26.0 software was used for data analysis. The data were expressed as the mean ± standard deviation. One-way analysis of variance (ANOVA) was used for the analysis of differences. Post hoc tests were performed using the LSD method when differences were statistically significant. P < 0.05 was considered statistically significant.

3 Results

3.1 Model validation

Figure 5 shows the percentage difference in von Mises stress between mesh 1 and mesh 3 and between mesh 2 and mesh 3. The greatest difference in predicted von Mises stresses between mesh 1 and mesh 3 was found in the cortical bone (4.06%). The differences in von Mises stress of each component between mesh 2 and mesh 3 was the least. Therefore, mesh 2 was selected. The ROM of the intact model at each segment was similar to the results and trends of Yamamoto et al. [19], Shim et al. [20], Huang et al. [21], Lo et al. [22] (Fig. 6), indicating that the intact L1-S1 FE model in this study was successfully constructed and can be used for further analysis.

3.2 ROM of the L4-L5 segment

Compared to the BPS-BPS group, the BMCS-BMCS group showed 1.9%, 5.7%, 5.8%, and 12.2% lower ROM in four motions. The BPS-BMCS group showed 3.7%, and 24.8% higher ROM in flexion and lateral bending, compared to the BPS-BPS group, and has 5.7%, and 33.3% higher ROM, compared to the BMCS-BMCS group. BPS-BMCS was decreased by 9.8%, 17.3%, 4.3%, and 5.8% in extension and rotation, respectively, compared to the BMCS-BMCS group. The BMCS-BPS group increased by 21%, 1.7%, and 7.6% in flexion, extension, and lateral bending, and decreased by 7.1% in rotation compared to the BPS-BPS group. The BPS-BMCS showed a 23.3%, 7.9%, 15%, and 5.8% increase than the BMCS-BMCS group in four motions, and a 16.7%, 12.7%, and 12.3% increase than the BPS-BMCS group in flexion, extension, and rotation, respectively, and 13.8% decrease in lateral bending (Fig. 7A). The BPS-BMCS group showed superior stability in extension and the BMCS-BMCS group in flexion and lateral bending. Only the difference between the BPS-BMCS and BMCS-BMCS in lateral bending was significant (P༜0.05).

3.3 Von Mises stress of the intervertebral cage

Compared to the BPS-BPS group, the BMCS-BMCS group showed 6.9% and 11.1% higher cage stress in flexion and lateral bending, and 3.3% and 0.5% lower in extension and rotation. The BPS-BMCS group showed 5.4% and 7.2% higher cage stress in flexion and lateral bending than the BPS-BPS group, and 8.7% and 2.9% lower cage stress in extension and rotation. Compared to the BMCS-BMCS, the BPS-BMCS showed reductions of 1.4%, 5.6%, 3.5% and 2.6% in all conditions. The BMCS-BPS group showed the same trend as the BPS-BMCS group, with an increase of 0.9% and 6.6% in flexion and lateral bending, and a decrease of 11.1% and 10.5% in extension and rotation, compared to the BPS-BPS group. Compared to the BMCS-BMCS group, the reductions were 5.6%, 8.1%, 4.0%, and 10.1% in four motions. compared to the BPS-BMCS group, the reductions were 4.2%, 2.7%, 0.5%, and 7.8% in four motions (Fig. 7B). The difference between each group was not significant (P༞0.05).

3.4 Von Mises stress of the screw

Compared to the BPS-BPS group, the BMCS-BMCS group showed a 2.0%, 2.0%, and 5.6% decrease in screw stress in flexion, extension, and lateral bending, respectively, and a 7.2% increase in rotation. The BPS-BMCS group showed a 21%, 8.7%, 5.6%, and 24.3% decrease in four motions, respectively. Compared to the BMCS-BMCS group, the BPS-BMCS group showed an 18.5%, 6.8%, and 29.4% decrease in flexion, extension, and rotation, and 2.4% increases in lateral bending. Compared to the BPS-BMCS group, the BMCS-BMCS group showed a 12.7% and 10.5% decrease in screw stress in flexion and extension and a 0.5% and 21.5% increase in lateral bending and rotation. This trend was similar to that of the BMCS-BMCS group, where the BMCS-BPS group decreased by 11% and 8.6% in flexion-extension, and increased by 9.0% and 13.4% in lateral bending and rotation. Compared to the BPS-BMCS group, there were increases of 9.2%, 6.5%, and 60.6% in flexion, lateral bending, and rotation, while in 1.9% increase in extension (Fig. 7C). The BPS-BMCS group was superior to its counterparts in terms of screw stress, but the difference between each group was not significant (P༞0.05). As shown in Fig. 8, peak von Mises stress of the screw was concentrated in the screw tail connecting with the screw hub.

3.5 Von Mises stress of the rod

The rod stress was shown in Fig. 9. Compared to the BPS-BPS group, the screw stress of the BMCS-BMCS group increased by 34.4%, 99%, 38.6%, and 120% in four motions, respectively, and the BPS-BMCS group decreased by 11% in flexion and increased by 13.25%, 1.9%, and 4.4% in the remaining motions. Compared to the BMCS-BMCS group, the BPS-BMCS group showed reductions of 34%, 43%, 26.5%, and 52.7% in all conditions. The BMCS-BPS group showed decreases of 3.1%, 0.3%, and 6.6% in flexion, extension, and rotation, and an increase of 12.3% in lateral bending compared to the BPS-BPS group (Fig. 7D). The BMCS-BPS group was superior in extension and rotation, and the BPS-BMCS group was superior in flexion and lateral bending, but the difference between each group was not significant (P༞0.05).

4 Discussion

The superior stability of the hybrid technique with MCBT and TT at the L4-L5 segment without fusion technique was previously demonstrated by our team [9]. However, patients with lumbar spinal stenosis were more common in clinical practice and often required thorough decompression and fusion, thus the L1-S1 lumbar spine models were established and TLIF procedures were performed at the L4-L5 segment. The selection of the fixation type is correlated with the occurrence of postoperative complications such as screw loosening and breakage, especially in patients with osteoporosis. As known, the CBT technique increased the pullout load by 30%, and torque by 1.7 times [5]. As previously shown by our team that the MCBT technique provides superior biomechanical stability compared to the pedicle screw technique and the CBT technique [6, 7]. Fujiwara et al. [23] demonstrated that increasing the medio-lateral angle and decreasing the distance to the anterior lateral edge of the upper endplate increases the insertional torque of the screw. The MCBT technique compensates for the deficiencies of the CBT technique. However, we also found that the MCBT technique still has deficiencies, such as limited decompression of the lateral recess and intervertebral foramen, splitting or fracture of the screw insertional point during decompression, and pars fracture are contraindications to the CBT technique [24]. If the BMCS-BMCS was implanted first and then proceed with the decompression, the decompression of the lateral recess might be incomplete. While the decompression proceeded first the screw insertion point might be damaged. It is advisable to use hybrid BMCS-BPS and BPS-BMCS fixation techniques according to the patient's preoperative radiographic data.

In this study, the stability of the BMCS-BMCS group in flexion and lateral bending, and the BPS-BMCS group in extension and rotation were superior to that of the BPS-BPS and BMCS-BPS groups. Perez-Orribo et al. [25] demonstrated that the stability of the CBT technique was superior in flexion and extension, but inferior in lateral bending and rotation and the differences between this study and the currenct study may be due to the greater MCBT screw diameter of 5.0 mm and length of 40 mm compared to the CBT screw with a diameter of 4.5 mm and length ranging from 25 to 35 mm in Perez-Orribo et al. [25] and contact between the MCBT screw and the cortical bone of the lateral edge of the upper endplate and medial wall of the pedicle. BPS-BMCS group offered superior stability in extension and rotation, but in flexion and lateral bending. Huang et al. [26] demonstrated that the inter-screw angle of 15° between the screw at the cranial and caudal level resulted in the lowest L4-L5 segmental ROM in flexion, extension, and rotation. In this study, the inter-screw angle was 8° in the BPS-BMCS group and 33° in the BMCS-BPS group. The morphology of the L5 vertebral arch is unique in that the superior edge of the vertebral body is almost at the same level as the superior edge of the vertebral arch, but the distance between the superior edge of the L4 vertebral arch and the vertebral body is somehow different from that of the L5 vertebral body, which has an "/\" pattern and moves more obliquely to the left and right, making the cranio-caudal angle of the screw in the sagittal plane more constant and milder than that of the L4 vertebral body. Therefore, when using the MCBT technique in the L5 vertebral body, the cranio-caudel angle should be reduced (around 25°) to keep the screw from penetrating the upper endplate of the vertebral body, while the L4 vertebral body has a larger cranio-caudel angle (22°- 35°) [6]. Although the inter-screw angle at a sagittal plane in the BPS-BMCS group was different from that of Huang et al. [26], the variation rule of ROM in extension and rotation was similar. Except for the inter-screw angle at the sagittal plane, which in the axial plane was also different.

The peak von Mises stress of the cage was lower in the BPS-BMCS group and BMCS-BMCS group than in the BPS-BPS group and BMCS-BMCS group in extension and rotation. In flexion and lateral bending, the cage stress of the BMCS-BMCS group was the greatest, The greater the cage stress, the higher the subsidence rate [27]. Numerous scholars have concluded that different types of fixation were essential to maintain the stability of the surgical segment and to reduce cage subsidence [28], but there is still no consensus regarding the ideal instruments. Different fixation methods have been reported previously, but the differences among the BPS-BPS, BMCS-BMCS, BPS-BMCS, and BMCS-BPS have not been reported. The previous study showed that the BMCS-BPS group had the lowest intervertebral disc stress in rotation. The BMCS-BPS group in the current study showed the least cage stress (51.7 ± 23.3 MPa) in rotation, which was comparable to the tendency in the previous model without fusion technique [9]. In the biomechanical analysis of Xu et al. [28], the largest von Mises stress was found in lateral bending, and the peak von mises stress of the single-cage (77.23 MPa) was greater than paired-cage (49.77 MPa) in TLIF model with BPS-BPS technique [29]. In this study, the peak von Mises stress was found in lateral bending in the BMCS-BMCS group (72.66 ± 12.4 MPa), but in the BPS-BPS group, The peak von Mises stress was found in extension (66.27 ± 18.85 MPa), variation rule of BPS-BPS group was inconsistent with Xu et al. [28]. This may be due to the different screw diameters, lengths, and cage sizes. The mean cage stress in the BMCS-BPS group was lowest in extension and rotation and lower in flexion and lateral bending, but did not reach a statistical difference (p > 0.05). In the RCT (randomized control trial) of Lee et al. [29], the fusion rates of CBT and pedicle screw technique were comparable (CBT 94.5% versus PS 91.4%, p > 0.05) at 24 months postoperatively, but the MCBT technique has not been used in the clinic. There was no statistical difference in the stability between the single and paired cage in the TLIF model [11]. Single cage implantation was preferable because of its convenience, and lower cost. As for the hybrid fixation with the BMCS and BPS, the different effects of single and paired cages on the stability of the fixation and fusion rate need further biomechanical study.

Increasing the medio-lateral angle of the pedicle screw reduced the risk of screw loosening and breakage [30]. Peak von mises stress of the screw in the hybrid fixation technique with CBT and pedicle screw, in Su et al. [31], were lower than that of the BPS-BPS technique in flexion, extension, and lateral bending. In this study, As for the BPS-BMCS group, the von Mises stress of the screw was lower than those of the BPS-BPS group in motion. As for the BMCS-BPS group, the von Mises stress of the screw was lower than that in the BPS-BPS group in flexion, extension, and rotation and the stress variation rules were comparable to those of Su et al. [31], BPS-BMCS group and BMCS-BPS group may reduce the incidence of screw breakage. The lowest screw stress was found in the BPS-BMCS (30.12 ± 5.73 MPa) and BMCS-BPS (32.79 ± 4.63 MPa) group in flexion. As for BCS-BCS (CBT at the L3 and L4), the von mises stress of the screw was greater than the counterparts in flexion and rotation [31], However, the opposite result was found in the present study, which may be related to the greater medio-lateral angle of the MCBT technique than that of the CBT technique as mentioned by Newcomb et al. [30]. Huang et al. [26] suggest that the longer rod length may decrease the screw stress. In this study, the BMCS-BPS group had the shortest rod length and the highest screw stress in lateral bending (200.34 ± 91.11 MPa) and rotation (193.35 ± 68.09 MPa), while the BPS-BMCS group had the longest rod length and the lowest von mises stress of the screw stress in flexion (60.81 ± 23.70 MPa) and rotation (120.37 ± 21.98 MPa) (p > 0.05). Results only in Rotation were consistent with Newcomb et al. [30], The reasons were probably due to the different sample sizes, different material properties, and inconsistent screw and rod sizes in the two studies. As shown in Fig. 8, the maximum stress in the screw occurs near the screw hub, that is, the BPS-BMCS group may reduce the risk of screw breakage.

Commonly, the junction of the screw and the connecting rod is the most common site for breakage. In this study, the von Mises stress of the rod in the BMCS-BMCS group was greater in all motions than its counterparts, with the maximum stress occurring in lateral bending (119.35 ± 43.20 MPa) at the lower end of the junction between the screw and the connecting rod (Fig. 8). The rod stress of the BMCS-BPS group was lower than that of the counterparts in extension and rotation, and the BPS-BMCS group also showed lower rod stress than the counterparts in flexion, and approximately similar to that of the BPS-BPS group in lateral bending. It can be deduced that the BPS-BMCS and BMCS-BPS groups may reduce the risk of rod breakage. Wang et al. [32] showed that the CBT-CBT group (CBT screw: 5.0 mm in diameter, 35 mm in length) had the highest rod stress in almost all working conditions while the UP-CBT group (TT screw: 6.5 mm in diameter, 45 mm in length), similar with the BMCS-BPS group, had the lowest rod stress in all motions. Although the length of the MCBT screw in this study was slightly longer and the TT screw was slightly shorter, the variation rules were consistent with the Wang et al. study [32]. Xiao et al. [33] demonstrated that the Dynesys hybrid fixation technique combined the advantages of dynamic stabilization and rigid fusion, using a "soft" rod in the upper segment to reduce the incidence of ASD (adjacent segment degeneration), further manifesting the principle of "overcoming rigidity with flexibility". By analogy, the effect on the adjacent segment of the "rigid" fixed segment using the hybrid fixation techniques discussed in this study and connecting with the "soft" rod needs to be further investigated.

Lumbar musculature plays an important role in spinal stability [34]. A previous study showed that S1 screw loosening was closely related to the degeneration of paraspinal muscles [35]. CBT screw reduces the postoperative serum creatinine phosphokinase concentration [36]. Since the insertion point of the MCBT technique is closer to the midline than the CBT technique [6, 7], the hybrid technique can further reduce the damage to the facet joint [9] and iatrogenic paravertebral muscle injury. Compared with the traditional BPS-BPS and CBT-CBT fixation techniques, the hybrid BPS-BMCS and BMCS-BPS techniques can not only overcome the embarrassing situation of incomplete decompression of the lateral recess after implantation of the screw but also reduce the surgical incision to reflect the principle of "minimally invasive surgery" and avoid the degeneration of the facet joint.

Perez-Orribo et al. [25] demonstrated that cortical screw systems offered greater rigidity in flexion and extension. However, it is worth noting that 20% of CBT screws may cause medial pedicle wall fractures resulting in nerve injury [5]. The medio-lateral angle of the MCBT technique is greater than that of the CBT technique, and it is extraordinarily difficult to complete the high precision screw placement with freehand in high accuracy, therefore, a 3D (3-dimensional) printed template or a robotic technique is required to assist in the nail placement process. Our research team has previously investigated the 3-dimensional printed template of MCBT techniques with in vitro experiments and demonstrated that it improves the accuracy of the MCBT screw placement [37].

There were some limitations in this study. First, the muscle tissues were not reconstructed, which may affect the stability of the lumbar spine. Second, this study did not analyze the effect of the hybrid technique on the adjacent segments in the TLIF model. Third, screw sizes of different diameters and lengths were not discussed in this study. Fourth, the sample size of this study is only three which needs to be increased in further study.

5 Conclusion

This study used finite element analysis to evaluate the biomechanical effects of four different fixation techniques in the L1-S1 lumbar spine with the TLIF model at the L4-L5 segment. The BMCS-BMCS group and the BPS-BMCS group provide superior stability than the BPS-BPS and BMCS-BPS groups. BPS-BMCS group presented a lower risk of screw breakage and BMCS-BPS group presented a lower risk of rod breakage. The BPS-BMCS and BMCS-BPS groups may reduce the cage subsidence rate. The results of this study support that the use of the BPS-BMCS and BMCS-BPS techniques in TLIF for offering superior stability and a lower risk of cage subsidence and instrument-related complication.

Abbreviations

TLIF

Transforaminal lumbar interbody fusion

PLIF

Posterior Lumbar Interbody Fusion

TT

Traditional trajectory

CBT

Cortical bone trajectory

BPS

Bilateral Pedicle Screw

BMCS

Bilateral modified cortical screw

FE

finite element

ROM

Range of motion

ALL

Anterior longitudinal ligament

PLL

Posterior longitudinal ligament

ITL

Intertransverse ligament

LF

Ligamentum flavum

CL

Capsular ligament

ISL

Interspinous ligament

SSL

Supraspinous ligament

BMD

Bone mineral density

Declarations

Ethics approval and consent to participate

The study design was approved by the Ethics Committee of Xinjiang Medical University before data collection and analysis (Approval No.20220930-01). All methods were carried out in accordance with relevant guidelines and regulations. Informed consent was obtained from the donor or next of kin . 

Consent for publication

Not applicable.

Availability of data and materials

The raw data supporting the conclusion of this article will be made available by the Corresponding author, without undue reservation.

Competing interests

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Funding

This study was supported by Central-class Public Welfare Research Institute Project of Chinese Academy of Medical Sciences (Grant No. 2022-JKCS-19).

Authors' contributions

AK, RZ, HLG, and PR conceived and designed the study and the interpretation of data, supervised the research, and assisted in reviewing and editing the manuscript. AK, YXW, and TZ conducted the experiment, completed the analysis and interpretation of data from the FE models, and wrote the initial draft of the article. RZ, AM, and HPL completed the acquisition of data and the analysis and interpretation of data. DSL and WJS provided clinical guidance and analyzed data. All authors contributed to the article and approved the submitted version.

Acknowledgements

Not applicable

References

  1. Deyo RA, Mirza SK, Martin BI, Kreuter W, Goodman DC, Jarvik JG. Trends, major medical complications, and charges associated with surgery for lumbar spinal stenosis in older adults.JAMA. 2010 Apr7;303(13):1259–65.
  2. Harms J, Rolinger H. [A one-stager procedure in operative treatment of spondylolistheses: dorsal traction-reposition and anterior fusion (author’s transl)]. Z Orthop Ihre Grenzgeb. 1982 Jun;120(3):343–7.
  3. Faizan A, Kiapour A, Kiapour AM, Goel VK. Biomechanical analysis of various footprints of transforaminal lumbar interbody fusion devices. J Spinal Disord Tech. 2014 Jun;27(4):E118–127.
  4. Verma K, Boniello A, Rihn J. Emerging Techniques for Posterior Fixation of the Lumbar Spine. J Am Acad Orthop Surg. 2016 Jun;24(6):357–64.
  5. Santoni BG, Hynes RA, McGilvray KC, Rodriguez-Canessa G, Lyons AS, Henson M. Cortical bone trajectory for lumbar pedicle screws. Spine J Off J North Am Spine Soc. 2009 May;9(5):366–73.
  6. Rexiti P, Abudurexiti T, Abuduwali N, Wang S, Sheng W. Measurement of lumbar isthmus parameters for novel starting points for cortical bone trajectory screws using computed radiography. Am J Transl Res. 2018;10(8):2413–23.
  7. Rexiti P, Aierken G, Wang S, Abudurexiti T, Abuduwali N, Deng Q, et al. Anatomical research on strength of screw track fixation in novel cortical bone trajectory for osteoporosis lumbar spine. Am J Transl Res. 2019;11(11):6850–9.
  8. Rexiti P, Aierken A, Sadeer A, Wang S, Abuduwali N, Deng Q, et al. Anatomy and Imaging Studies on Cortical Bone Screw Freehand Placement Applying Anatomical Targeting Technology. Orthop Surg. 2020 Dec;12(6):1954–62.
  9. Kahaer A, Maimaiti X, Maitirouzi J, Wang S, Shi W, Abuduwaili N, et al. Biomechanical investigation of the hybrid modified cortical bone screw-pedicle screw fixation technique: Finite-element analysis. Front Surg. 2022;9:911742.
  10. Zander T, Rohlmann A, Bergmann G. Influence of different artificial disc kinematics on spine biomechanics. Clin Biomech Bristol Avon. 2009 Feb;24(2):135–42.
  11. Xu H, Ju W, Xu N, Zhang X, Zhu X, Zhu L et al. Biomechanical comparison of transforaminal lumbar interbody fusion with 1 or 2 cages by finite-element analysis.Neurosurgery. 2013 Dec;73(2 Suppl Operative):ons198-205; discussion ons205.
  12. Wang Md K, Jiang PhD C, Wang PhDL, Wang Md H, Niu PhD W. The biomechanical influence of anterior vertebral body osteophytes on the lumbar spine: A finite element study. Spine J Off J North Am Spine Soc. 2018 Dec;18(12):2288–96.
  13. Kiapour A, Ambati D, Hoy RW, Goel VK. Effect of graded facetectomy on biomechanics of Dynesys dynamic stabilization system. Spine. 2012 May 1;37(10):E581-589.
  14. Lu YM, Hutton WC, Gharpuray VM. Do bending, twisting, and diurnal fluid changes in the disc affect the propensity to prolapse? A viscoelastic finite element model.Spine. 1996 Nov15;21(22):2570–9.
  15. Natarajan RN, Watanabe K, Hasegawa K. Biomechanical Analysis of a Long-Segment Fusion in a Lumbar Spine-A Finite Element Model Study. J Biomech Eng. 2018 Sep 1;140(9).
  16. Kim HJ, Chun HJ, Lee HM, Kang KT, Lee CK, Chang BS, et al. The biomechanical influence of the facet joint orientation and the facet tropism in the lumbar spine. Spine J Off J North Am Spine Soc. 2013 Oct;13(10):1301–8.
  17. Fan W, Guo LX, Zhao D. Stress analysis of the implants in transforaminal lumbar interbody fusion under static and vibration loadings: a comparison between pedicle screw fixation system with rigid and flexible rods. J Mater Sci Mater Med. 2019 Oct;18(10):118.
  18. Ayturk UM, Puttlitz CM. Parametric convergence sensitivity and validation of a finite element model of the human lumbar spine. Comput Methods Biomech Biomed Engin. 2011 Aug;14(8):695–705.
  19. Yamamoto I, Panjabi MM, Crisco T, Oxland T. Three-dimensional movements of the whole lumbar spine and lumbosacral joint. Spine. 1989 Nov;14(11):1256–60.
  20. Shim CS, Park SW, Lee SH, Lim TJ, Chun K, Kim DH. Biomechanical evaluation of an interspinous stabilizing device,Locker. Spine. 2008 Oct15;33(22):E820-827.
  21. Huang YP, Du CF, Cheng CK, Zhong ZC, Chen XW, Wu G, et al. Preserving Posterior Complex Can Prevent Adjacent Segment Disease following Posterior Lumbar Interbody Fusion Surgeries: A Finite Element Analysis. PLoS ONE. 2016;11(11):e0166452.
  22. Lo HJ, Chen HM, Kuo YJ, Yang SW. Effect of different designs of interspinous process devices on the instrumented and adjacent levels after double-level lumbar decompression surgery: A finite element analysis. PLoS ONE. 2020;15(12):e0244571.
  23. Fujiwara S, Ohnishi Y, Iwatsuki K, Kishima H. Cortical bone trajectory fixation cause low compression force in anterior vertebral column. North Am Spine Soc J. 2022 Jun;10:100113.
  24. Kim KT, Song MG, Park YJ, Lee DY, Kim DH. Cortical Trajectory Screw Fixation in Lumbar Spine Surgery: A Review of the Existing Literature. Asian Spine J. 2022 Feb;16(1):127–40.
  25. Perez-Orribo L, Kalb S, Reyes PM, Chang SW, Crawford NR. Biomechanics of lumbar cortical screw-rod fixation versus pedicle screw-rod fixation with and without interbody support.Spine. 2013 Apr15;38(8):635–41.
  26. Huang ZB, Nie MD, Zhang NZ, Liu S, Yuan JB, Lin XM, et al. Biomechanical evaluation of a short-rod technique for lumbar fixation surgery. Front Bioeng Biotechnol. 2022;10:959210.
  27. Guo HZ, Tang YC, Guo DQ, Liang D, Zhang SC. Biomechanical evaluation of four different posterior instrumentation techniques for single-level transforaminal lumbar interbody fusion: a finite element analysis. Am J Transl Res. 2020;12(10):6160–9.
  28. Xu DS, Walker CT, Godzik J, Turner JD, Smith W, Uribe JS. Minimally invasive anterior, lateral, and oblique lumbar interbody fusion: a literature review. Ann Transl Med. 2018 Mar;6(6):104.
  29. Lee GW, Ahn MW. Comparative Study of Cortical Bone Trajectory-Pedicle Screw (Cortical Screw) Versus Conventional Pedicle Screw in Single-Level Posterior Lumbar Interbody Fusion: A 2-Year Post Hoc Analysis from Prospectively Randomized Data. World Neurosurg. 2018 Jan;109:e194–202.
  30. Newcomb AGUS, Baek S, Kelly BP, Crawford NR. Effect of screw position on load transfer in lumbar pedicle screws: a non-idealized finite element analysis. Comput Methods Biomech Biomed Engin. 2017 Feb;20(2):182–92.
  31. Su KC, Chen KH, Pan CC, Lee CH. Biomechanical Evaluation of Cortical Bone Trajectory Fixation with Traditional Pedicle Screw in the Lumbar Spine: A Finite Element Study. Appl Sci-Basel. 2021 Nov;11(22):10583.
  32. Wang TN, Wu BL, Duan RM, Yuan YS, Qu MJ, Zhang S, et al. Treatment of Thoracolumbar Fractures Through Different Short Segment Pedicle Screw Fixation Techniques: A Finite Element Analysis. Orthop Surg. 2020 Apr;12(2):601–8.
  33. Xiao X, Chen G, Wang S, Liu J, Lin E, Chen K, et al. Efficacy of the Dynesys Hybrid Surgery for Patients with Multi-Segmental Lumbar Spinal Stenosis. Front Surg. 2022;9:849679.
  34. Panjabi MM. Clinical spinal instability and low back pain.J Electromyogr Kinesiol. 2003Aug;13(4):371–9.
  35. Kim JB, Park SW, Lee YS, Nam TK, Park YS, Kim YB. The Effects of Spinopelvic Parameters and Paraspinal Muscle Degeneration on S1 Screw Loosening. J Korean Neurosurg Soc. 2015 Oct;58(4):357–62.
  36. Ohkawa T, Iwatsuki K, Ohnishi YI, Ninomiya K, Yoshimine T. Isthmus-guided Cortical Bone Trajectory Reduces Postoperative Increases in Serum Creatinine Phosphokinase Concentrations. Orthop Surg. 2015 Aug;7(3):232–8.
  37. Shi W, Aierken G, Wang S, Abuduwali N, Xia Y, Rezhake R, et al. Application study of three-dimensional printed navigation template between traditional and novel cortical bone trajectory on osteoporosis lumbar spine. J Clin Neurosci. 2021 Mar;85:41–8.

Tables

Table 1 is available in the Supplementary Files section.