Potter Tower Calibration
The pressurised misting nozzle of the PT must be adjusted to produce a uniform spray onto the substrate positioned on the centre of the spray table (Fig. 1). Liquids pipetted into the delivery vial at the top are drawn through an atomiser by compressed air (15–20 psi). The gas pressure is checked and adjusted if necessary, and the guide screws are positioned to secure the petri dish. The first calibration step is to level the tower and the nozzle arms via the adjustable feet, using the integrated spirit level, alongside a visual assessment that the nozzle is centralised. Once the device is level, the nozzle must be centralised within the tower to apply the chemical uniformly across the spray table. Four targets (deposition samplers) are placed on the spray table, and the amount deposited on each is measured by weight using a mixture of water and glycerol (50:50) to minimise losses due to evaporation compared to water alone.
The deposition samplers rest on pieces of Bostik Blu Tack® to hold them away from the surface (Petri dish) to prevent additional wicking of liquid from the sprayed surface (Fig. 1). The amount of spray deposited onto each deposition sampler should be similar if the nozzle is correctly centralised. Uneven distribution can be resolved by adjusting the height of the arms that hold the nozzle. Where deposition is highest, the nozzle arm is too high on that side, so the nozzle is tilted away by lowering that arm or raising the two opposite arms. This manipulation of the three nozzle arms may require several attempts to achieve uniformity as it is not intuitive to adjust accurately; a four-arm system would have been more straightforward.
The initial weight of each deposition sampler is recorded, and they are positioned to form a square in the centre, with A-D always in the same position (Fig. 1). A set volume (2 ml) of the 50% glycerol solution is sprayed, after which the samplers are removed and re-weighed, recording the value next to the correct position of each (A-D) on a data sheet. The difference between the initial weight and the weight after spraying is calculated. This process is repeated until the weight difference between each cover slip is ≤ 10% of the average of the four. Table 2 summarises each step in the treatment process for both sprayers with the time taken to complete. The time taken to calibrate and centralise the nozzle can vary one of the best recorded times was 50 min and one of the worst times was a total of 2 hours and 10 mins.
The next calibration step is to confirm what volume should be pipetted into the delivery vial to deposit the required weight of spray solution onto the target surface. With the PT, most of the contents of the delivery vial is not delivered to the target surface on the spray table. Therefore, it is necessary to measure how much is actually deposited onto the target surface and then to adjust the volume of each insecticide solution needed in the delivery vial to achieve the desired spray weight [9]. The LITE method for measuring the amount of insecticide solution reaching the test surface is to spray a Petri dish containing glycerol with water; the glycerol will reduce evaporative loss of the water. The Petri dish is weighed before and after spraying to calculate the delivered spray weight. The volume aliquoted into the spray vial is adjusted until the appropriate spray deposit in weight per unit area is achieved for that spray solution (within ± 5%).
During these tests, however, the volume deposited onto on the surface was measured via fluorimetry. The fluorescent tracer is soluble providing a direct volumetric measure of deposit not impacted by evaporative losses. The amount deposited on the spray table was measured using a 9 cm diameter Petri dish as a collection surface. 1.7 ml of solution was sprayed, and the volume collected in the Petri dish was an average of 191 ± 14 µl which translates to 30 ml/m2, and correlates well with previous spray rate measurements. This volume represents 11.2% of the spray solution pipetted into the delivery vial meaning that 88.80% of applied spray solution is not deposited onto the spray target. The main route of loss of spray droplets is onto the walls of the tower, with the rest likely exhausted from the bottom of the PT.
Calibration of the PT must be repeated each time it is used: the weight of deposit should be confirmed for each spray solution, and the centralisation repeated before each spray session and each time the tower is cleaned, for example in between spraying different solutions. The time taken, particularly for the first centralisation step, varies considerably; during this study the time taken ranged from 2 hours 10 min to 50 min. The total time to calibrate, spray test surfaces and clean ranged from 4 hours 5 min to 2 hours 50 min.
Track Sprayer Calibration
The Horizontal TS cabinet contains a spray table (79 x 229 cm), enclosed within a Perspex cabinet with an opening for user accessibility that is covered with a set of doors (Fig. 2). The spray track working length is 160 cm, and the spray table is 46.2 cm below the nozzle. In this study, deposition samplers rested on Petri dishes to give a surface height of 1.2 cm, meaning that the spray surface was 45 cm below the nozzle. This is the same nozzle distance advised for the application by backpack sprayers used for IRS in the field. The nozzle traverses the entire length of the spray track, the spray zone, however, was the central 1 m of the track; 30 cm at each end is not used due to the edge effects at the start up and shutdown of spray nozzles. The spray swath width is 70 cm, but only the central 50 cm is used due to tapering of deposits at the edge of the spray pattern. The track is powered by a stepper motor governed by a control unit that offers calibrated speeds between 0.2–0.75 m/s with separations of 0.05 m/s. The travelling speed was set to 0.45 m/sec for the experiments described here, which corresponds to 1 m sprayed every 2.2 seconds as per WHO guidelines for application of IRS treatment to walls [1].
The input parameters for the TS are speed, nozzle type (FF 80 02 in this model of the TS which is the same nozzle used in the backpack sprayers during IRS operations), distance between nozzle and spray surface, and pressure. The tank was pressurised to 60 PSI, the control flow valve, however, maintains the pressure at the nozzle at 20 PSI. The settings to achieve the target application rate of 30 ml/m2, were as follows: speed setting of 0.45 m/sec with a Flat fan 80 02 (spray angle and orifice size respectively), at a distance of 45 cm from the table surface. The track sprayer requires an in-depth periodic (every 12 months or whenever any parameter changes) calibration of the following:
Swath area: To ensure no deviation of the effective swath width (0.7m), sensitive papers (Spraying System, Wheaton, Ill., 26 × 76 mm) are used that irreversibly change from yellow to blue on contact with a liquid. This test aims to identify failures due to nozzle wear. The deposition characteristics are evaluated using 10 individual sensitive papers placed width ways across the spray track, and the spray application is replicated 3 times. The TS passes this test if the sensitive papers are uniformly blue across the central (0.5m) swath.
Track speed: To confirm the speed stipulated on the track controller, a known length of the track is marked, and the time for the nozzle to traverse that length is timed to enable a calculation of speed in metres/second (m/s). An alternative method is to measure the distance travelled by the belt on one revolution of the stepper motor drive and to time the number of revolutions per second.
Acute volume check: The volume deposited per unit area should be assessed via fluorimetry using the appropriate settings. There is generally a linear response between light intensity and tracer concentration when samples are not saturated, thus allowing a calculation of volume from calibration standards. If the dosage applied to the samples is incorrect, settings must be modified accordingly. The speed of the TS is the primary parameter to adjust; it can be increased and decreased to reduce and increase the volume applied, respectively.
The set up for the Track sprayer is measuring out the fluorescent tracer and the insecticide to the tank closing the tank and pressurizing to 60 PSI. Full calibration of the TS takes 1 hr 30 min, and this should be repeated every 6 months. Otherwise, only a routine check of sprayer settings is required, which takes 15 min. The time taken for calibration with the TS can range from 15 min to 1hr 30 min. The total time to calibrate spray and clean ranged from 2 hours 50 min to 4 hours 5 min.
Table 2
Comparative assessment of the time taken to achieve various goals.
|
|
Time Taken (minutes)
|
Activity
|
Potter Tower best-case
Calibration
|
Potter Tower worst-case
Calibration
|
Track Sprayer check Calibration
|
Track Sprayer full
Calibration
|
Set up for calibration stage 1
|
10
|
10
|
10
|
10
|
PT calibration
|
Calibration stage 1: centralisation of the potter tower (samplers < 10% of the mean weight)
|
20
|
90
|
N/A
|
N/A
|
Calibration Stage 2: volume required for the desired spray weight
|
20
|
20
|
N/A
|
N/A
|
TS calibration
|
Calibration TS − 3 reps carried out at 0.45 m/s for the full calibration 15 min to check settings
|
N/A
|
N/A
|
15
|
90
|
Time to spray
|
20
|
20
|
10
|
10
|
Time taken for fluorometric analysis
|
60
|
60
|
60
|
60
|
Time to clean
|
60
|
60
|
75
|
75
|
Total
|
190
|
260
|
170
|
245
|
Table 2 Comparative assessment of the time taken to achieve various goals. Note that the PT requires calibration every time it is used whereas the TS only once every six months. The PT calibration time is extremely variable, and this is reflected with best- and worst-case examples.
Sprayer Cleaning
Cleaning of PT: Water is added to the atomiser reservoir and sprayed through three times. The nozzle is removed and cleaned with distilled water flushing through the feeding tube. The inside of the PT is wiped with Decon 90 on absorbent paper towel; this is repeated three times. Distilled water is then sprayed through the PT.
Time taken actively cleaning the Potter tower: 1 hour
Cleaning the TS: The exterior of the TS and associated electronic parts require minimal cleaning. All spray tubes are purged until clear; the tank is depressurised, and any remaining insecticide discarded. The tank is refilled with water and 5% solution of Decon 90 re pressurised and rinsed through to decontaminate followed by a rinse with water alone.
Time taken to clean Track sprayer tubing: 1 hour 15 minutes every time a distinct compound is used.
To clean the Perspex chamber the cabinet is first washed down with water and drained. Then 5% Decon 90 is sprayed over the surface and left for 60 minutes to decontaminate, and the cabinet is wiped down. This is followed by two more water rinses, and lastly the cabinet is sprayed with 70% ethanol which is left to evaporate.
Time taken to clean Track sprayer chamber: 1 hour 15 minutes (including a 60-minute waiting period for Decon decontamination) at the end of a treatment period.
Uniformity of the deposited volume
Both the PT and the TS were calibrated to deposit 30 ml/m2 and, on average, both sprayers deposited 31 ml/m2; the difference between the two sprayers was in the variability in the deposits (Tables 4 and 5). Between treatments the PT had a standard deviation of 11 vs a standard deviation of 3 for the TS. Within treatments the standard deviations were < 15 for the PT and < 4 for the TS.
Table 3 Spray uniformity for the Potter Tower and the Track Sprayer.
Table 3 Spray uniformity for the Potter Tower and the Track Sprayer. Spray uniformity is calculated as a total for uniformity between replicates and uniformity across single treatment tiles. Data presented as average ml/m2, and standard deviation
If a calibration metric of < 10% deposition deviation from the mean of all samplers is applied as a metric of pass or fail, it is clear that the PT immediately fell out of calibration for uniformity of deposits. However, in all but three instances, the TS passed this calibration metric.
Table 4 Deviation from the mean of spray from the Potter Tower and the Track Sprayer.
Table 4 Deviation from the mean of spray from the Potter Tower and the Track Sprayer. The deviation from the mean (< 10%) is calculated as a metric of calibration for the potter tower was assigned to the data for both sprayers
Deposition of active ingredient
HPLC Analysis
The HPLC analysis (Jan 2021) was run alongside the fluorimetry. The fluorimetry showed that the volume of liquid deposited on the surface was as expected, but HPLC analysis showed very different numbers for the active ingredient.
The deposition of spray droplets measured via fluorimetry for Actellic 300 CS, a microencapsulated suspension, and K-Othrine WG250, a wettable granule formulation, showed that the calibrated volume of formulated product made up as a suspension in water (30 ml/m2) was deposited onto the surface.
-
The TS deposited an average of 28.3 ± 2.4 ml/m2 of Actellic 300CS spray solution and 30.8 ± 3.8 ml/m2 of K-Othrine WG250 spray solution
-
The PT deposited and average of 30.2 ± 11.8 ml/m2 of Actellic 300CS spray solution and 29.7 ± 5.0 ml/m2 of K-Othrine WG250 spray solution
In contrast, the HPLC data showed that significantly more insecticide had been deposited using the PT and significantly less using the TS. The desired application rate of Actellic 300CS on the surface was 1,000 mg AI/m2, and the desired concentration for K-Othrine WG250 was 25 mg AI/m2.
-
The PT deposited an average of 2,137 ± 607 mg AI/m2 for Actellic 300CS and an average of 75.5 ± 8.7 for K-Othrine WG250
-
The TS deposited an average of 92.4 ± 13.9 mg AI/m2 for Actellic and 8.4 ± 0.9 mg AI/m2 for K-Othrine WG250
The PT deposited approximately 2x and 3x the intended dose of Actellic 300CS and K-Othrine WG250, quantified by the amount of AI recovered and measured by HPLC, respectively, whilst the TS deposited a 10th and a 3rd of the intended dose.
Tank samples (spray solution) throughout testing with K-Othrine WG250 were analysed, which flagged a potential flaw with the TS; there was a sudden drop in the concentration from the tank sample taken post spray (Fig. 3). Visual inspection of the TS showed signs of the active ingredient sedimenting out in the lines. The suspension of diluted product in the tank can be maintained by constant agitation. The sedimentation of the formulation can be addressed by flushing out the lines after any significant stoppage in spray applications alongside regular agitation of the spray tank. At the same time, samples were taken from the stock solution of K-Othrine WG250 before and after spraying with the PT, and directly from the nozzle, which confirmed that the active ingredient content remained constant, and no significant sedimentation was occurring (Fig. 3).
With sedimentation of the IRS formulation identified as the likely cause of the observed under spray with the TS, a follow up set of tests were run to compare K-Othrine WG250 with another formulation of deltamethrin with more favourable suspension characteristics (Suspend PolyZone, Bayer USA): the recommended application rate of both formulations is 25 mg AI/m2. TS spray tank was agitated immediately prior to pumping to target pressure and applied promptly to spray surfaces to prevent sedimentation within the delivery tubes during set up.
Measurement of the volume of spray deposited via fluorimetry from K-Othrine WP250 and Suspend PolyZone applications showed that the calibrated volume of formulated product (30 ml/m2) was deposited to the surface.
-
The TS deposited an average of 31 ± 1.6 ml/m2 for K-Othrine WG250 and 30 ± 1.9 ml/m2 for Suspend PolyZone.
-
The PT deposited an average of 31 ± 2.8 ml/m2 for K-Othrine WG250 and an average of 32 ± 2.3 ml/m2 (SD 2.3) for Suspend PolyZone.
The HPLC analysis showed that the improved agitation of the TS system to stop sedimentation of active ingredient returned the desired concentrations of deltamethrin sprayed in the two formulations. The PT deposited approximately 20% more than the target application rate for both formulations.
Both formulations were still over-applied when using the PT, but not to the same degree as seen previously. Using a formulation with improved suspension qualities showed some improvement, meaning the over-application could result from the active ingredient partitioning during the descent of droplets through the tower. Ultimately, however, the reason for the over-application is still unknown. What can be seen is that the deviation in the deposition of active ingredient from the target dose exceeds the uniformity metric of 10% when using the PT (15%) but not when using the TS (%)
Tarsal contact bioassay
The bio-efficacy data from the first trials showed that Actellic 300CS applied using the PT and TS at the recommended label rate and one-quarter of this rate delivered a deposit of pirimiphos-methyl which remained bioactive (≥ 80% mortality in a WHO cone test) for at least 3 months (Fig. 4). At one quarter label rate (4a), the TS sprayed tiles gave 82% mortality and the PT sprayed tiles 62% mortality at 6 months post-application, but both sets of sprayed tiles at label rate returned < 80% mortality in bioassays at 7- and 8-months post-applications. For Actellic 300CS applied at the full label rate (4b), the PT sprayed tiles failed (< 80% mortality) at 6 months, whereas the TS maintained residual efficacy through 7 months before failing at 8 months.
The second round of testing applied Actellic 300CS and K-Othrine WG250 at the recommended application rate. The active ingredient concentration on all surfaces was measured by HPLC analysis. K-Othrine WG250 remained active, killing 100% of mosquitoes in a cone test for 3 months. Actellic 300CS had reduced efficacy at two months when applied using both the PT and TS; at 3 months mortality rebounded up to 100% with the PT sprayed surfaces but further declined with the TS sprayed surfaces. This matches other observations that the TS routinely shows logical progressions, whereas the data associated with the PT tends to be more erratic, possibly due to the more variable spray deposition profiles. HPLC analysis showed that the PT deposited significantly higher rates of active ingredient compared to the TS which is at odds with the bioassay results where mortality was not strikingly different between the TS and PT treated tiles.