This paper demonstrated how PixelPrint can be utilized to create patient-specific 3D printed bone and soft tissue CT phantoms using one filament. Our approach provides economical and efficient means of producing high resolution CT phantoms, exhibiting excellent accuracy in HU and image texture characteristics in CT scans. These phantoms are useful for a wide range of academic research and clinical evaluation of CT performance.
In contrast to prior studies of image-based 3D printed bone phantoms using slices of the human head/skull [13], chest/thoracic cage [15], pelvis [14] and femoral shaft [6], this study printed the human cervical vertebrae with surrounding soft tissue. Human vertebrae particularly present a challenging task for 3D printing, as they contain intricate details and are comparatively smaller in size. Nevertheless, these areas, especially in combination with the adjacent tissues, are not only fundamental in clinical diagnostic applications, such as the assessment of severe fractures or degenerative diseases, but also crucial in surgical interventional planning. Our phantoms possess the potential to be utilized for those applications, such as optimizing CT protocols for the assessment of bone mineral density [37] among others. Here, only human cervical vertebrae and knee joint phantoms were printed, but the approach can be extended to any bone structure. With StoneFil filament, a range of approximately − 227 HU to 851 HU for a CT scans with a tube voltage of 120 kVp can be reliably printed using PixelPrint, with a deviation of less than 15 HU compared to patient data. This range covers most tissue types in the human body and is applicable to various research applications.
Continuing our previously published research on the PixelPrint lung phantom [18], [19], this study not only extended the types of human tissue printed, but also enhanced the resolution and stability of PixelPrint. Filament line spacing was reduced from 1.0 to 0.5 mm, potentially doubling the resolution capabilities of the printed phantoms. Phantoms produced using this approach can have greater filament coverage and finer details in a given area, serving as valuable tools to evaluate the efficacy of novel higher resolution CT systems such as photon-counting CT [38]–[40]. Printing finer lines with PLA/Stone filament poses more challenges to printer stability control and requires finer system tuning. By optimizing extrusion rate, printing speed, nozzle temperature, and acceleration speed, PixelPrint can still produce highly accurate patient phantoms in reliable stability as demonstrated by qualitative and quantitative evaluation. Additionally, micro-CT acquisitions revealed that filament lines and underlying structure can be generated with high degree of consistency.
With the growing popularity and accessibility of 3D printing technology, a variety of printing filaments are now available for printing human bone and soft tissue. Several studies have discussed materials for 3D-printed phantoms in CT [24]–[26]. Novel filament materials composed of hydroxyapatite and biocompatible, biodegradable polymers, such as CT-Bone (Xilloc Medical Int., Sittard-Geleen, the Netherlands), can be utilized for printing synthetic bone implants that rapidly induce bone regeneration and growth [41], [42]. Filaments made from composites of fatty acids and ceramic powders have also been explored [28]. However, bone-like filaments available in the general market (FibreTuff, Toledo, OH, USA), suitable for medical surgery purposes [29], [30], do not necessarily have high radiometric densities and are not capable to reach much higher than 400 HU in CT scans. While cancellous bone is only about 300 to 400 HU in CT images, cortical bone can range from 500 HU and up to over 1900 HU [43]. By contrast, materials such as vinyl and PLA with stone (PLA/stone) can offer up to nearly 1000 HU at 96.9% infill ratio at tube voltage of 120 kVp, as they exhibit relatively higher X-ray absorption. Additionally, considering materials for spectral CT phantoms, high impact polystyrene (HIPS) based filaments may be suitable for mimicking CT numbers in applications where energy dependence is important [26], because they show similar spectral profiles as the human body. In this study, we employed StoneFil filament, one type of PLA/stone filament. Unlike normal PLA, StoneFil filament is gravimetrically filled with 50% powdered stones, resulting in significantly higher material density and enabling denser printed objects. Carbonate calcium-containing limestones exhibit a similar X-ray response in CT to that of human bone, whose density can be attributed to hydroxyapatite. This property was reflected in the the spectral response of the printed vertebrae with its similarity to that of hydroxyapatite.
This study has a few limitations: (i) The filament used in our study did not encompass the entire range of Hounsfield Units (HU) required for bone structures. Future research should focus on the development of next-generation filaments that cover the full HU range while preserving spectral capabilities. (ii) The calcium-based material used in the printing process was applied to the entire print, including soft tissue regions. While this approach does not severely impact performance in conventional CT applications, it may have an influence on the evaluation with spectral CT. To achieve the full dynamic range with spectral characterization for both soft tissue and bone, further development of multiple print head systems will be required. (iii) The printed phantoms were limited to a specific field of view. Future studies should explore the potential to print larger anatomical regions, such as the entire chest or abdomen.