X-ray topography detects crystallinity information from X-ray images diffracted by a sample using Bragg or Laue-Case X-ray diffraction, as shown in Fig. 6a. Since the intensity of the diffracted X-rays in each region of the sample depends on the crystal defects and distortions, these defects and distortions can be visualized from the intensity changes. However, the intensity change is integrated along the optical path, which means the depth direction information cannot be obtained. Conventional section topography has therefore been utilized to detect the three-dimensional location of defects and distortions in a sample by means of multiple topograms acquired using sheet-shaped X-rays while scanning the sample, as shown in Fig. 6b.
The depth resolution of the section topography depends mainly on the height h of the incident X-ray beam at the irradiated position. The sheet-shaped X-ray beam is usually formed by an X-ray slit with an opening aperture D, and the beam height h' at the irradiated position on the sample is given by the summation of D and the beam broadening due to diffraction. Therefore, h' cannot be infinitely small due to the diffraction limit, and the minimum value is theoretically calculated from D, the X-ray wavelength λ, and the distance x between the slit and the irradiation position. For example, h' is calculated as 3 µm for D = 1 µm, λ = 0.1 nm, and x = 10 mm. In addition, the slit cuts the beam size down to 1/1000 or less, which reduces the X-ray intensity by the same ratio. Thus, even for very strong X-rays such as synchrotron radiation, a long measurement time is required. Another problem is that it is technically quite difficult to make such a narrow slit with an opening aperture of a few microns. In principle, it is impossible for conventional sectional topography to achieve a depth resolution of less than 1 µm.
This limitation can be circumvented by using a sheet-shaped beam focused in one dimension by an X-ray focusing device. The X-ray beam is focused separately in the vertical and horizontal directions by two total reflection mirrors in a Kirkpatrick-Baez (KB) optical configuration. Therefore, one-dimensional focused X-ray is obtained by retracting one mirror from the optical beam path, as shown in Fig. 6c (in this case, the horizontal focusing mirror was retracted). The beam size at the focal point can be focused at less than 1 µm by using the latest X-ray focusing system and mirror, and µ-XRT with a depth resolution below 1 µm is expected to be easily achieved. Although the divergence angle of the X-ray beam will be wider than that of the X-ray beam cut by a slit in conventional section topography, it can be suppressed to less than sub-mrad by utilizing an X-ray slit upstream of the mirror. Note that if the sample is scanned up and down vertically, as in conventional section topography, the focal point and the sample irradiation position will be misaligned. Therefore, the sample has to be scanned parallel to the surface so that the focal point of the X-ray beam and the sample irradiation point always coincide, as shown in Fig. 6c.
We designed the 3D µ-XRT system using the microbeam system20 installed at the beamline BL16XU of the SPring-8 in Japan. As shown in Fig. 7a, the system consists of a focusing mirror, a sample positioner, and an X-ray micro-imager. The X-ray passes through a 4D slit installed upstream as a virtual light source and is focused on the sample by the total-reflection elliptical mirror in vertical directions. Note that a second focusing mirror for the horizontal direction is retracted from the optical beam path to form a sheet-shaped X-ray beam in this optical configuration. The incidence angle of the mirrors is 5 mrad, and the distances from the 4D slit to the center of the mirrors and from the center of the mirrors to the focal point are 5,050 mm and 250 mm, respectively. Hence, the reduction ratio is approximately 1/20, and the X-ray beam can be focused to 1 µm by setting the 4D-slit vertical aperture to 20 µm.
The sample positioner consists of a swivel table that adjusts the incident angle of the X-ray to satisfy the Bragg diffraction condition, a Y-axis linear table that scans the sample parallel to the sample surface to keep the X-ray focal point and the irradiation point of the sample at the same position, and Z-axis and X-axis tables that adjust the vertical and horizontal positions of the sample. All tables are driven by a stepping motor and controlled remotely.
A lens-coupled X-ray imager (Rad device, Xsight Micron™) is used for the X-ray micro imager. The incident X-rays are converted to visible light by a phosphor and then imaged onto an sCMOS visible camera (Andor Zyla, pixel size: 6.5 µm, 2048 × 2048 pixels) by the visible light lens system. We utilize a 5× objective lens with the pixel size of 1.3 µm and the field of view of 2.6 mm2. The imager is mounted on the 2θ arm of the X-ray diffractometer of the microbeam system. To reduce image blurring, we limit the distance between the X-ray imager and sample to within 5 mm using the height table of the 2θ arm.
Obtained topograms are shifted when scanning the sample along the Y-axis (as shown in Fig. 7b), so we reconstruct a standard 3D topogram with a ratio of 1:1:1 and orthogonally intersect each axis by shifting, stacking, and rotating each topogram using SAKAS-Viewer21 and Image J.