Background : Hyperinflation of laryngeal mask cuffs may carry the risk of airway complications. The manufacturer recommends inflating cuff until the intracuff pressure reaches 60 cmH 2 O, or inflate with the volume of air to not exceed the maximum recommended volume. We prospectively assessed the correlation of cuff inflating volumes and pressures, and the appropriated the cuff inflating volumes to generate an intracuff pressure of 60 cmH 2 O in the adult laryngeal masks from different manufacturers. Methods : Two groups of eighty patients requiring laryngeal mask size 3 and 4 during general anesthesia were randomized into 4 subgroups for each size of the laryngeal mask: Soft Seal ® (Portex ® ), AuraOnce TM (Ambu ® ), LMA-Classic TM (Teleflex ® ) and LMA-ProSeal TM (Teleflex ® ). After insertion, the cuff was inflated with 5-ml increments of air up to the maximum recommended volume. After each 5-ml intracuff pressure was measured, the volume of air that generated the intracuff pressure of 60 cmH 2 O was recorded . Results : Mean(SD) volume of air required to achieve the intracuff pressure of 60 cmH 2 O in Soft Seal ® , AuraOnce TM , LMA-Classic TM , LMA-ProSeal TM laryngeal mask size 3 were 11.80(1.88), 9.20(1.88), 8.95(1.50) and 13.50(2.48) ml, respectively, and these volumes in laryngeal mask size 4 were 14.45(4.12), 12.55(1.85), 11.30(1.95) and 18.20(3.47) ml, respectively. The maximum recommended volume resulted in high intracuff pressures (>60 cmH 2 O) in all laryngeal mask types and sizes studied. Conclusion : Pressure-volume curves of adult laryngeal masks are all in sigmoidal shape. Cuff designs and materials can effect pressure and volume correlation. Approximately half of the maximum recommended volume is required to achieve the intracuff pressure of 60 cmH 2 O except LMA-ProSeal TM which required two-thirds of the maximum recommended volume. Trial registration: Thai Clinical Trials Registry, TCTR20150602001, May 28, 2015

Figure 1

Figure 2

Figure 3

Figure 4
Loading...
On 27 Apr, 2020
On 27 Apr, 2020
Posted 11 Oct, 2019
On 20 Apr, 2020
Received 22 Jan, 2020
Invitations sent on 06 Nov, 2019
On 06 Nov, 2019
On 10 Oct, 2019
On 09 Oct, 2019
On 09 Oct, 2019
On 05 Jun, 2019
Received 20 Mar, 2019
Received 20 Mar, 2019
On 10 Mar, 2019
On 08 Mar, 2019
Invitations sent on 05 Mar, 2019
On 12 Feb, 2019
On 05 Feb, 2019
On 05 Feb, 2019
On 27 Apr, 2020
On 27 Apr, 2020
Posted 11 Oct, 2019
On 20 Apr, 2020
Received 22 Jan, 2020
Invitations sent on 06 Nov, 2019
On 06 Nov, 2019
On 10 Oct, 2019
On 09 Oct, 2019
On 09 Oct, 2019
On 05 Jun, 2019
Received 20 Mar, 2019
Received 20 Mar, 2019
On 10 Mar, 2019
On 08 Mar, 2019
Invitations sent on 05 Mar, 2019
On 12 Feb, 2019
On 05 Feb, 2019
On 05 Feb, 2019
Background : Hyperinflation of laryngeal mask cuffs may carry the risk of airway complications. The manufacturer recommends inflating cuff until the intracuff pressure reaches 60 cmH 2 O, or inflate with the volume of air to not exceed the maximum recommended volume. We prospectively assessed the correlation of cuff inflating volumes and pressures, and the appropriated the cuff inflating volumes to generate an intracuff pressure of 60 cmH 2 O in the adult laryngeal masks from different manufacturers. Methods : Two groups of eighty patients requiring laryngeal mask size 3 and 4 during general anesthesia were randomized into 4 subgroups for each size of the laryngeal mask: Soft Seal ® (Portex ® ), AuraOnce TM (Ambu ® ), LMA-Classic TM (Teleflex ® ) and LMA-ProSeal TM (Teleflex ® ). After insertion, the cuff was inflated with 5-ml increments of air up to the maximum recommended volume. After each 5-ml intracuff pressure was measured, the volume of air that generated the intracuff pressure of 60 cmH 2 O was recorded . Results : Mean(SD) volume of air required to achieve the intracuff pressure of 60 cmH 2 O in Soft Seal ® , AuraOnce TM , LMA-Classic TM , LMA-ProSeal TM laryngeal mask size 3 were 11.80(1.88), 9.20(1.88), 8.95(1.50) and 13.50(2.48) ml, respectively, and these volumes in laryngeal mask size 4 were 14.45(4.12), 12.55(1.85), 11.30(1.95) and 18.20(3.47) ml, respectively. The maximum recommended volume resulted in high intracuff pressures (>60 cmH 2 O) in all laryngeal mask types and sizes studied. Conclusion : Pressure-volume curves of adult laryngeal masks are all in sigmoidal shape. Cuff designs and materials can effect pressure and volume correlation. Approximately half of the maximum recommended volume is required to achieve the intracuff pressure of 60 cmH 2 O except LMA-ProSeal TM which required two-thirds of the maximum recommended volume. Trial registration: Thai Clinical Trials Registry, TCTR20150602001, May 28, 2015

Figure 1

Figure 2

Figure 3

Figure 4
Loading...