STS contains indole ring produces a low inherent FI in aquatic solution, which can be detected at 350 (λex 286 nm)[2]. The FI of STS was greatly increased using aqueous methanol at acidic pH. This may be attributed to the formation of protonated species of the cited drug in acidic medium which have enhanced fluorescence more than neutral species[20]. Its native fluorescence intensity, in aqueous methanol, was dramatically increased by about 20-folds in the presence of 0.05 N H2SO4 as shown in Fig. 2. Therefore, the emission band of STS at 350nm was used to develop a novel procedure intended for its analysis in tablets formulation and human plasma.
3.1. Optimization of experimental conditions
The experimental parameters influencing the FI of STS were studied and optimized. The general procedure was applied where the parameter under study was changed one at a time but the others were kept the same.
Various acids such as HCl, HNO3, H2SO4, and H3PO4 were tried to predict their effect on FI of drug under study. Highest FI with lowest blank reading were observed upon using sulphuric acid (Fig. 3). Therefore, H2SO4 was chosen as the appropriate acid for the drug under study.
The influence of the volume of 0.05 N H2SO4 on the FI was investigated using various volumes of reagent solution varied from 0.5 to 3.0 ml. It was observed that a steady increase in FI occurred upon using volumes of acid reached to 1 ml, after that no more raise in FI was achieved. Consequently, 1.5 ml of 0.05 N H2SO4 was selected as the suitable volume for estimation of cited drug.
The effect of several diluting solvents such as distilled water, ethanol, methanol, acetonitrile, acetone, and ethanol on the FI of the studied drugs was investigated. According to this study, highest FI and the lowest blank were observed upon using methanol (Figure S1). Thus, it was used as diluting solvent for in the general procedure.
The FI of STS in several organized media was examined. Examples of these media used throughout this study are anionic surfactant (SDS), cationic surfactant (CTAB), non-ionic surfactant (tween-80) and macromolecules (CMC). 1.5 ml from each media was used (0.05 N). Results are shown in (Figure S2). It was observed that, FI of the studied drug in all cases was decreased compared to treated control sample of drug freed from the organized media used. This finding can be explained based on the fact that the micelle formation was hindered by addition of methanol which in turn lead to partial or full damage of the micelle-enhancement effect of these surfactants on the FI of cited drug [21].
Under the described experimental condition provided under 2.4., the FI was developed instantly after dilution with methanol and remained stable for about 1 hr, time selected for study (Figure S3). Consequently, FI was measured directly after dilution of all the studied concentrations.
FI was checked against variation of temperature after dilution with methanol, over 25–55oC range, utilizing temperature controlled water bath. The observed FI was decreased upon increasing the temperature. This finding can be explained based on the higher internal conversion path at higher temperature, thus leads to suppression of the excited singlet state by other way [20]. Thus, all FI measurements were done at ambient temperature.
3.2. Validation of the suggested method.
To confirm that the performance features of the proposed procedure fulfill the demands for the designed analytic implementations, validation of the proposed method should be approved through ICH guidelines[18]. The entire group of validation steps were accomplished within the particular range of the suggested procedure to guarantee its validation.
3.2.1. Linearity and range
A sequence of 6 concentrations of studied drug were analyzed and the results obtained were utilized to construct standard calibration graphs where the FI reading were plotted against concentrations within the calibration range. The test results were analyzed by least squares method and the regression equation which correlate FI to its corresponding concentration was computed [22]. The analytical parameters such as effective range, intercept (Sa), slope (Sb), correlation coefficient (r) and determination coefficient (r2) are illustrated in Table 1. In addition detection Limit (LOD) and quantification limits (LOQ) were also calculated as reported[18].
Table 1
Analytical parameters for the analysis of Sumatriptan by proposed fluorimetric method
Item | Sumatriptan (STS) |
λex(nm) | 286 |
λem(nm) | 350 |
linear range (µg /ml) | (0.1-2.0) |
LOD (µg /ml) | 0.019 |
LOQ (µg /ml) | 0.057 |
Correlation coefficient (r) | 0.999 |
Determination coefficient (r2) | 0.999 |
Slop(b) | 0.437 |
Intercept(a) | 27.230 |
SD of the intercept (Sa) | 2.495 |
SD of slope (Sb) | 0.002 |
SD of residual ( Sy.x) | 4.771 |
LOD: Limit of detection (µg /ml). |
LOQ: Limit of quantitation (µg/ml). |
3.2.2. Accuracy and precision
The accuracy of the suggested procedure was checked out by analysis for six various concentrations of authentic sample of drug by the suggested procedure. As illustrated in Table 2, it was clear that there is a good coincidence between the found and real values, which reflect the perfect accuracy of the suggested procedure. Precision at both Intraday and interday were also evaluated using 3 concentrations (high, medium and low) where each concentration were analyzed six times. Data of such study (Table 3) reflect perfect repeatability and reliability of the suggested procedure as proved by values of relative standard deviation values where the values obtained with all concentrations studied were below 2%.
Table 2
Evaluation of the accuracy of the proposed fluorimetric method
Sample no. | Sumatriptan | |
Added (µg/ml) | found* (µg /ml) | % Recovery* |
1 | 0.3 | 0.29 | 97.50 | |
2 | 0.5 | 0.49 | 98.00 | |
3 | 0.9 | 0.89 | 98.50 | |
4 | 1.0 | 0.99 | 99.00 | |
5 | 1.5 | 1.47 | 98.30 | |
6 | 2.0 | 1.97 | 98.70 | |
Mean | | 98.30 | |
SD | 0.53 | |
RSD | 0.54 | |
*Average of three replicate measurements |
SD: Standard deviation. |
RSD: Relative standard deviation. |
Table 3
Evaluation of Interday and Intraday precision of the proposed fluorimetric method
Parameter | Sumatriptan (% found)* | |
0.3 µg/ml | 0.6 µg/ml | 1.0 µg/ml | | |
Intraday assay | 1 | 97.40 | 98.90 | 98.20 | |
2 | 98.70 | 98.60 | 97.50 |
3 | 98.00 | 99.30 | 99.00 |
Mean | 98.00 | 98.90 | 98.20 |
SD | 0.65 | 0.35 | 0.75 |
Interday Assay | RSD | 0.66 | 0.35 | 0.76 |
1 | 98.10 | 98.60 | 99.00 |
2 | 98.70 | 98.00 | 98.50 |
3 | 98.40 | 99.30 | 98.00 |
Mean | 98.40 | 98.60 | 98.50 |
SD | 0.30 | 0.65 | 0.50 |
RSD | 0.31 | 0.66 | 0.51 |
* Average of six determinations. |
3.2.3. Robustness of the suggested procedure
The robustness of the suggested procedure was checked versus little intentional variations in the experimental conditions such as variation in the volume of 0.05N H2SO4 (1.5 ± 0.5 ml), the excitation wavelengths (286 ± 5 nm) and emission wavelengths (350 ± 5 nm). The good percentage recoveries and standard deviations observed proved that the method is robust since small variations in volume of 0.05 N H2SO4, excitation or emission wavelengths have no considerable effect on the results of the proposed procedure. Consequently, the current procedure is reliable and appropriate for the analysis of the cited drug in either pure form or in different matrices.