Objective
Percutaneous transforaminal endoscopic discectomy (PTED) is minimally invasive and has been widely used to treat patients with lumbar disk herniation(LDH) due to its safety and efficiency. However, due to the unique anatomy of the L5-S1 level, the PTED procedure is often difficult to perform in the region. ZESSYS, a targeted and quantificational foraminoplasty device, may help to overcome these anatomical limitations. In this study, we assessed the efficiency and the short-term effects of PTED with ZESSY at the L5-S1 level.
Methods
Between January to August of 2018, Fifty-six patients with lumbar disk herniation at the single level of L5–S1 and underwent percutaneous transforaminal endoscopic discectomy were enrolled in this retrospective cohort study. They were segregated into the the TESSYS group and the ZESSYS group. The puncture time, foraminoplasty time, decompression time and fluoroscopy time evaluated for the operation efficiency. Clinical outcomes were assessed by visual analog scale(VAS) score and Oswestry Disability Index(ODI) score. The MacNab criteria were used to evaluate patient subjective satisfaction at 12-month follow-up postoperatively.
Results
The average puncture time (5.29±2.05 minutes), foraminoplasty time (12.82±2.52 minutes) and fluoroscopy time (26.29±5.96 seconds) were all significantly shorter in the ZESSYS group than in the TESSYS group (average puncture time 8.07±3.13 minutes, p<0.01; foraminoplasty time, 17.18±2.92 minutes, p<0.01; fluoroscopy time, 34.73±6.86 seconds; p<0.01). No significant differences were observed between the 2 groups in the decompression time (p=0.057). The VAS score of low back pain and leg pain, as well as the ODI score, improved at all time points postoperatively compared with preoprative , in both the TESSYS group and the ZESSYS group (P < 0.05). There were no significant differences in the VAS score of low back pain, VAS score of leg pain, and ODI score between the TESSYS group and the ZESSYS group at the same time points (P > 0.05). According to the MacNab criteria, the excellent and good rate at 12-month follow-up postoperatively was 85.7% in the TESSYS group and 89.3% in the ZESSYS group (P > 0.05).
Conclusion
The targeted and quantificational foraminoplasty device named ZESSYS was more efficient in the puncture and foraminoplasty procedures, effectively protecting the exiting nerve and minimizing the level of radiation exposure. The device is efficient and safe for PTED in treating lumbar disk herniation at the L5-S1 level.
Lumbar disc herniation (LDH) is a frequently occurring disease of the spine and leads to economic and medical burdens on families and society[1, 2]. Conservative treatment for LDH is effective, but a considerable number of patients will ultimately receive surgical treatment[3, 4]. Open microdiscectomy has been considered as the standard for surgical treatment for LDH until now. However in recent decades, percutaneous transforaminal endoscopic discectomy (PTED), via a posterolateral approach,has gained popularity in clinical practice due to advantages such as little trauma, few scares, a rapid recovery and a short hospital stay[5–9]. For this minimally invasive surgery, a working cannula with a diameter of 7.5 mm needs to be introduced safely to the space in front of the spinal dural sac via the posterolateral direction, and foraminoplasty often needs to be performed to enlarge the intervertebral foramen. The superior articular process (SAP) is the main obstacle for foraminoplasty. TESSYS is a classic technique that was invented by Hoogland[10] in 2006. Graded isocentric trephine and accompanying instruments are used to cut the upper and ventral aspects of the SAP. The intervertebral foramen is widened gradually, so that the working cannula and rod-shaped endoscope can enter the epidural space. Extruded and sequestered discs can be resected and the region of foraminal stenosis can be decompressed directly. However, in this procedure, the isocentric trephine makes contact with the exiting nerve root, traversing nerve root and para-foramen soft tissue, which is risk and arising concerns of damage to nerves[11]. In addition, at the L5-S1 level, the unique anatomy involving a high iliac crest, sacral ala, large facet joint, large L5 transverse process, narrowed disc space, and narrowed foramen can complicate the foraminoplasty process. These anatomical obstacles and inter-individual variability in the anatomy make PTED technically challenging [12]. PTED is more difficult to performed at this level in particular[13], even for skilled and experienced surgeons.
ZESSYS, is a novel targeted and quantificational foraminoplasty device originated from a modified version of the traditional TESSYS technique. The novel effective foraminoplasty tool was designed by Yue Zhou et al from the Second Affiliated Xinqiao Hospital of Army Medical University in Chongqing, China. The hypothesis is that it would be easier to implement and may help overcome the anatomical limitations at the L5-S1 level. In this study, the novel targeted and quantificational foraminoplasty device were used to perform PTED at the L5-S1 level. Meanwhile its foraminoplasty efficiency and the short-term clinical outcomes were assessed.
from January to August of 2018, a totle of Fifty-six patients with lumbar disk herniation who were treated in the Department of Orthopedics in Xinqiao Hospital at the Army Medical University were enrolled in the study. The inclusion criteria were as follows: (1) lumbar disk herniation at the single level of L5-S1 confirmed by preoperative magnetic resonance imaging (MRI) and computed tomography (CT) scans; (2) low back pain accompanied by sciatica and corresponding radiculopathy signs such as motor weakness, sensory deficiency and the presence of abnormal reflex (more than one of the above signs); (3) a lack of improvement after conservative treatment for at least 12 weeks; and (4) an age of 18–70 years. The exclusion criteria were as follows: (1) central stenosis (less than 10 mm) or lateral recess stenosis (less than 3 mm) confirmed by MRI and CT scans; (2) spinal instability confirmed by dynamic radiographs; (3) calcified disc herniation; (4) far lateral disc herniation; (5) a history of surgery in the same segment; (6) cauda equina syndrome; (7) BMI > 28 kg/m2; and (8) the unwillingness or inability to participate in the treatment. All 56 patients were averaged to the TESSYS and the ZESSYS group. All the patients completed the follow-up at 12 months postoperatively. All the study procedures were approved by the Ethics Committee of Xinqiao Hospital, at the Army Medical University. The procedures were performed by a single skilled and experienced surgeon.
Surgical instruments
The spine transforaminal endoscope system (TESSYS instrument system: Joimax, Inc., Irvine, California, USA) and tip-flexible electrode bipolar radiofrequency system (Elliquence LLC, Baldwin, New York, USA) were used in the two groups. In the ZESSYS group, the patented ZESSYS instrument
instead of the conventional grade guide rod and trephine was used for targeted and quantificational foraminoplasty. ZESSYS is a dual-cannula instrument that has 4 graded sizes. The thinner cannula containing Kirschner wire for fixation and the other larger cannula for foraminoplasty by using trephine. The detailed characteristics of ZESSYS including the external form, initial application and advantages were first detailed by Ao et al [14]. C-arm fluoroscopy was used for radiograph imaging in both groups.
Surgical procedures
The procedure was performed with local anesthesia combined with analgesic drugs. The patient laid on the radiolucent table in the prone position, and a soft cushion was placed beneath the abdomen to reduce the degree of lumbar lordosis. The puncture point was 12–14 cm from the middle line and a short distance from the iliac crest if the iliac crest obstacle to the predetermined trajectory. Local infiltration anesthesia was induced by 0.5% lidocaine around the puncture point. An 18-gauge needle was inserted toward the upper and ventral surface of the SAP for stepwise and additional local anesthesia. After the cranial part of the SAP was crossed, the tip of the 18-gauge needle was pushed gently to the ideal epidural position toward the middle line of the vertebral canal under anteroposterior imaging and toward the vertebral posterior superior margin under lateral imaging to the greatest extent possible. If needed an adjustment toward to the herniated disk was performed. At that point, the 18-gauge needle was withdrawn and replaced by a guidewire. Over the guidewire, a stab incision of measuring approximately 1 cm was made, and stepwise guiding and dilatation rods were introduced. Subsequently, foraminoplasty was performed after the ventral side of SAP was partially removed with the expansion tube, guide rod, and trephine in sequence. After surgical access was established, a working channel slope was placed close to the intervertebral disc. In all cases, the herniated disc and the extruding or sequestrated disc fragment could be observed. Then, discectomy and nerve root decompression were performed in the same manner as was the routine PTED procedure. After decompression, the surgical instruments were pulled out, and the wound was sutured.
This procedure was also performed with local anesthesia combined with analgesic drugs. The protocol used before the puncture step was identical to that used in the TESSYS group. Then an 18-gauge needle was inserted into the foramen and across the safe triangle stepwise. During this step, the needle was inserted along the ventral aspect of the SAP to prevent exiting nerve root injury. After traditional guidewire replacement and graded dilation, a 1-mm Kirschner wire was slightly hammered across the safe triangle and fixed on the the superior-lateral part of the posterior aspect of the distal vertebra. Following the Kirschner wire the double-cannula device with a properly predetermined diameter was inserted until the larger cannula reached the SAP through a 1 cm incision. Afterwards, the double-cannula device was rotated and centered on the Kirschner needle to determine the ideal trajectory. Foraminoplasty was then performed by a trephine, and part of the ventral side of the SAP was cut and removed. If needed, the foramen was widened further by a larger graded cannula or another rotational motion. Then, discectomy and nerve root decompression were performed in the same manner as in the routine PTED procedure. After decompression, the surgical instruments were pulled out, and the wound was sutured.
Statistical analyses were performed using the SPSS 17.0 statistical package (SPSS, Inc., Chicago, Illinois, USA). The two independent samples t-test, paired t-test, Chi-square test and Mann-Whitney U test were used for data analyses. The measurement data are presented as the means ± standard deviations. For all analyses, a p value less than 0.05 was considered statistically significant.
Fifty-six patients (TESSYS, 28 patients; ZESSYS, 28 patients) who underwent PTED surgery from January to August of 2018 were enrolled in this study. The baseline demographics for both groups are shown in Table 1. The grade of foraminal stenosis on the operative side at L5–S1 was classified according to the report of Lee et al[15]. The grade of f Iliac height at L5–S1 was classified according to the report of Choi et al[16]. No significant differences were observed between the 2 groups (Table 1).
TESSYS | ZESSYS | p value | |
---|---|---|---|
Age,years | 52.54 ± 9.10 | 49.21 ± 10.38 | 0.208 |
Sex,male:female | 18:10 | 15:13 | 0.415 |
ASA criteria, I:II | 20:8 | 18:10 | 0.571 |
BMI (kg/ m2 ) | 22.28 ± 2.69 | 21.99 ± 2.22 | 0.659 |
Smoker, yes:no | 16:12 | 17:11 | 0.786 |
Duration of symptoms, weeks | 15.21 ± 2.54 | 16.04 ± 2.96 | 0.271 |
Herniation type, contained:uncontained | 9:19 | 11:17 | 0.577 |
Foraminal Stenosis grade(L5-S1) 0:1:2:3 | 19:9:0:0 | 17:10:1:0 | 0.523 |
Iliac height grade 1:2:3:4:5:6 | 0:0:3:8:10:7 | 0:0:1:9:11:7 | 0.730 |
ASA, american society of anesthesiologists; BMI, body mass index. |
As shown in Table 2, the average puncture time(5.29 ± 2.05 minutes), foraminoplasty time(12.82 ± 2.52 minutes), fluoroscopy time (26.29 ± 5.96 seconds) and total surgery time(63.14 ± 7.76 minutes) in the ZESSYS group were significantly shorter than those in the TESSYS group (average puncture time 8.07 ± 3.13 minutes, p < 0.01; foraminoplasty time, 17.18 ± 2.92 minutes, p < 0.01; fluoroscopy time, 34.73 ± 6.86 seconds; p < 0.01; total surgery time 74.21 ± 12.16 minutes, p < 0.01 ). No significant differences were observed between the 2 groups in the decompression time (p = 0.057).
TESSYS | ZESSYS | p value | |
---|---|---|---|
Puncture time (min) | 8.07 ± 3.13 | 5.29 ± 2.05 | < 0.01 |
Foraminoplasty time (min) | 17.18 ± 2.92 | 12.82 ± 2.52 | < 0.01 |
Fluoroscopy time (s) | 34.73 ± 6.86 | 26.29 ± 5.96 | < 0.01 |
Decompression time (min) | 48.96 ± 9.02 | 45.04 ± 5.65 | 0.057 |
Total surgery time (min) | 74.21 ± 12.16 | 63.14 ± 7.76 | < 0.01 |
As shown in Table 3, the VAS scores for low back pain and leg pain, as well as the ODI score, improved at all time points postoperatively, both in the TESSYS group and the ZESSYS group (p < 0.05). There were no significant differences in the VAS score for low back pain, VAS score for leg pain, ODI score or MacNab criteria between the TESSYS group and the ZESSYS group at the same time points (p > 0.05). The excellent and good rates according to the MacNab criteria at 12 months postoperatively was 85.7% in the TESSYS group and 89.3% in the ZESSYS group. |
TESSYS | ZESSYS | p value | |
---|---|---|---|
VAS score for low back pain | |||
Preoperative | 3.60 ± 0.95 | 3.38 ± 0.69 | 0.310 |
Postoperative | 2.20 ± 0.92* | 2.36 ± 1.03* | 0.541 |
12 months | 1.48 ± 1.02* | 1.30 ± 0.84* | 0.467 |
VAS score for leg pain | |||
Preoperative | 5.91 ± 1.23 | 6.17 ± 1.61 | 0.505 |
Postoperative | 2.10 ± 0.97* | 2.25 ± 0.86* | 0.532 |
12 months | 1.22 ± 0.70* | 1.10 ± 0.84* | 0.572 |
ODI score | |||
Preoperative | 59.36 ± 8.33 | 61.57 ± 7.89 | 0.312 |
Postoperative | 15.50 ± 4.48* | 16.07 ± 4.82* | 0.648 |
12 months | 12.50 ± 3.06* | 12.07 ± 3.67* | 0.637 |
MacNab criteria (12 months) | |||
Excellent | 17 | 19 | 0.569 |
Good | 7 | 6 | |
Fair | 4 | 3 | |
Poor | 0 | 0 | |
VAS visual analog scale; ODI Oswestry Disability Index. | |||
*Compared with preoperatively, P < 0.05. |
All the patients in the two groups were successfully decompressed with immediate nerve relief. No cases required conversion to an open procedure, and no major complications such as dural tears, infections or vascular injuries were observed. One patient in the TESSYS group developed postoperative dysesthesia (POD), which was considered to be caused by irritation to the dorsal root ganglion. With temporary conservative treatment, the symptoms disappeared.
PTED technic has been an effective and safe procedure for LDH[17–20]. Hoogland et al firstly invented the TESSYS technique, in which a graded trephine was used to enlarge the foramen gradually and the working channel slope is placed close to the intervertebral disc. However, the extruded and sequestered disc fragments are accessible only when the operating instruments are placed in the optimal trajectory. Accuracy was highly demanded,especially for the puncture and foraminoplasty procedure. At the L5-S1 level, the unique anatomy includes a high iliac crest, sacral ala, large L5 transverse process, large facet joint, narrowed disc space, and narrow foramen hindering safe and effective access through a narrow channel. The puncture[21], foraminoplasty and channel indwelling processes are complicated. An optimal trajectory may be hard to achieve or inaccessible, which leads to a steeper learning curve[22] and challenges, even for skilled surgeons. In addition, the gradual trephine is isocentric, which increase the risk of damage to nerve roots. Most POD cases due to dorsal root ganglia stimulation or injury have been reported to occur in L5-S1 during the PTED procedure[23, 24]. To address the above problem, the ZESSYS, a targeted and quantificational foraminoplasty device with a double-cannula is conducive to overcoming the anatomic limitations at the L5-S1 level.
Firstly, it can reduce the difficulties of acupuncture. In our study, the average puncture time(5.29 ± 2.05 minutes) in the ZESSYS group were significantly shorter than those in the TESSYS group (average puncture time 8.07 ± 3.13 minutes, p < 0.01). In the conventional TESSYS technique, an18-gauge needle is inserted toward the middle line of the vertebral canal under AP imaging and toward the vertebral posterior superior margin under lateral imaging, and subsequently, foraminoplasty is performed to achieve a sufficient level of surgical access for PTED. If needed the targeted point can be adjusted for special extruded and sequestered disc fragments. When the high iliac crest, large L5 transverse process, and large facet joint obstruct to the trajectory at the L5-S1 level,the ideal or targeted puncture point is time-consuming, difficult or even inaccessible to reach, and the foraminoplasty time and fluoroscopy time increase correspondingly. ZESSYS is a dual-cannula adjustment instrument with a thin cannula containing a Kirschner wire for orientation and a larger cannula for bony abrasion by a trephine(Fig. 1). Primary puncture with an 18-gauge needle was only demanded to across the safe triangle area stepwise, after the guidewire was introduction and graded dilation was performed, the traditional guidewire was replaced with a 1-mm Kirschner wire and slightly hammered to fix on the posterior aspect of the distal vertebra, which could guide the following introduction of the dual-cannula device for foraminoplasty. The trajectory of the Kirschner wire did not need to strive for the accurate requirement of TESSYS technique; it was necessary to insert the wire between the exiting nerve root and the SAP, and it could be fixed at any position near the superior-lateral part of the posterior aspect of the distal vertebra. The double-cannula system take advantages to rotation and can be easily adjusted to find a proper and targeted entry point on the SAP, which can compensate for the Kirschner wire primary puncture point(Fig. 2). In addition, depending on the last trajectory of the Kirschner wire, size 4 of the graded dual-cannula size may be selected for selective resection of the SAP for quantificational foraminoplasty(Fig. 1, Fig. 2). However, as a rule, an extruding or sequestrated disc fragment is within the scope of the working channel’s coverage. In addition, if predetermined effective access to enter or across the safe triangle area is hard to achieve or inaccessible by using an 18-gauge needle in the case of a large L5 transverse process, large facet joint, narrowed disc space, or narrow foramen impeding the predefined trajectory, after replacement, the Kirschner wire can take place of the needle for the puncture process as it is more rigid and easier to adjust. In particular the use of a Kirschner wire with a 1 mm diameter for trajectory adjustment and fixation in the safe triangle area increases the risk of nerve root damage, so when the Kirschner wire advances into the foramen, it should cling to the ventral aspect of the SAP, and the patient’s responses should be taken into consideration to avoid possible nerve root and adjacent soft tissue injury. If possible, the tip of the Kirschner wire should be fixed between the inner side of the SAP and the middle line in the anteroposterior view, thereby allowing a small amount abrasion with the upper articular process.
Secondly, the ZESSYS device can improve the efficiency and safety of foraminoplasty. In this study, the average foraminoplasty time(12.82 ± 2.52 minutes) in the ZESSYS group were significantly shorter than those in the TESSYS group (average foraminoplasty time, 17.18 ± 2.92 minutes, p < 0.01). Among the anatomical limitations at the L5-S1 level, the iliac crest is considered a major obstacle[25, 26]. When the iliac crest is located above the middle of the L5 pedicle in the lateral radiograph, foraminoplasty may be required[27]. In addition, if the working spaces are limited by the other anatomical limitations, foraminoplasty is also necessary. In the conventional TESSYS technique the foramen is widened gradually by an isocentric trephine. During the process of foraminoplasty, the trephine blade makes close contact with foramen soft tissue and nerve roots, leading to a risk of damage, espically for the L5-S1 level with a narrowed disc space and intervertebral foramen than the other levels. During the PTED process with ZESSYS, a 1 mm Kirschner wire is passed between the SAP and exiting nerve root and fixed on the the posterior aspect of the distal vertebra, which acts as a steady pivot for the double-cannula device. When the predefined cannula is inserted to dock on the SAP at posterior orientation, it can be easily rotated to find the proper trajectory and achieve quantificationally decompression. In addition, the beveled design cannula is placed next to the ventral side of the SAP, excluding the exiting nerve root from the working zone of the trephine and protecting the exiting nerve root from damage. If needed, the foramen can be enlarged by a second cannula rotation. The double-cannulas greatly reduce the difficulty of foraminoplasty and enable foraminoplasty to be performed more precisely.
According to our study, the ZESSYS device is more efficient than the TESSYS device in the puncture and foraminoplasty procedures. The difference in the puncture time and foraminoplasty time between the two groups were significant. Not only that, the average fluoroscopy time (26.29 ± 5.96 seconds) and total surgery time(63.14 ± 7.76 minutes) in the ZESSYS group were significantly shorter than those in the TESSYS group (average fluoroscopy time, 34.73 ± 6.86 seconds; p < 0.01; total surgery time 74.21 ± 12.16 minutes, p < 0.01 ). The fluoroscopy time decreased with the application of ZESSYS, which was beneficial for both the patients and surgeons. The preliminary postoperative outcomes seemed to be equal between the ZESSYS group and TESSYS group. There were no significant differences in the VAS score for low back pain, VAS score for leg pain, ODI score or MacNab criteria between the two groups at the same time point. No cases of recurrent herniation were observed at the 12-month follow-up. However 1 patient in the TESSYS group experienced slight dysesthesia after surgery. Although his symptoms disappeared within 2 weeks after temporary conservative treatment. It is an alert of exiting nerve damage which is a common complication of PTED[18, 23, 28]. There were no cases of severe complications in the ZESSYS group. One of the reasons for this result might be that the new instrument not only effectively widened the foramen but also effectively protected the nerve root.
Nevertheless, our study has some limitations. Firstly, the sample sizes of the two groups were small, and the follow-up time was short. The clinical outcome was preliminary, and a larger study needs to be conducted in the future to verify the the reliability of the ZESSYS system. In addition, at the L5-S1 level, the unique anatomical limitations were triaxial and complicated, and there were no specific protocols for the evaluation of the anatomical limitations. Therefor, in this study, we analyzed two major anatomical limitations between the two groups, which may not be enough and may affect the statistical correctness.
The double-cannula device named ZESSYS is effective and safe for PTED for L5-S1 disc herniation. It was more efficient in the puncture and foraminoplasty procedure, as well as protected the exiting nerve effectively and minimized the degree of radiation exposure.
PTED: Percutaneous transforaminal endoscopic discectomy, LDH: Lumbar disk herniation, VAS: Visual analog scale, ODI: Oswestry disability index, SAP: Superior articular process, POD: Postoperative dysesthesia, MRI: Magnetic resonance imaging, CT: Computed tomography.
Acknowledgements
Not Applicable.
Authors’ contributions
Jinlong Liu and Junlong Wu: Conceptualization, Methodology , Data curation, Writing - Original Draft. Honglei Zhang: Data curation, Formal analysis. Rui Zuo and Jiabin Liu: Investigation, Data Curation. Chao Zhang: Conceptualization, Methodology , Supervision, Writing - Review & Editing.
Funding
None.
Availability of data and materials
All data generated or analyzed during this study are included in the article
Ethics approval and consent to participate
Ethical approval was obtained by the Ethics Committee of Xinqiao Hospital, at the Army Medical University. All consents to
participate were obtained from the participants
Consent for publication
We have consent for publication for every patient in our series.
Competing interests
The authors declare that they have no competing interests.
Author details
1Department of Spine Surgery, People’s hospital of Liaocheng, Liaocheng 252000, Shandong, China. 2Department of Spine Surgery, The 941th Hospital of Joint Logistics Support Force of Chinese PLA, Xining 810000, Qinghai, China. 3Department of Spine Surgery, Xinqiao Hospital, Army Medical University, Chongqing,400037, China