First phase of PROUD study
PROUD (pulse PhotoplethysmogRaphy as an early tool for the diagnosis of sepsis thrOUgh a two-stage Development approach) was a clinical study that was conducted in four study sites (two departments of Internal Medicine and two Intensive Care Units) participating in the network of the Hellenic Sepsis Study Group (HSSG) (www.sepsis.gr). The study protocol (CIV-19-06-028824) was approved by the Ethics Committees of the participating study sites, by the National Ethics Committee of Greece (approval MD 3/19) and by the National Organization for Medicine of Greece. (approval MD 3/19; ClinicalTrials.gov; NCT04149132). The enrolment of patients took place between November 2019 and February 2020. Once the analysis of the data was available in March 2020 when the COVID-19 pandemic was prominent in Greece, it was considered appropriate to ask for an extension of the study to validate the results in patients with infection by SARS-CoV-2. This extension was approved by the National Organization for Medicines on March 30th 2020. Written informed consent was provided by the patients or by first-degree relatives in case of patients not able to consent. The patients analysed here have not been reported in any other submission by our group or anyone else.
Participants were adults of both genders at high risk for the development of sepsis. High risk for the development of sepsis was considered as the presence of any two of the following situations: a) any infection in a patient with total SOFA score equal to 0 or 1; and b) patients with Charlson’s Comorbidity Index (CCI) more than 2 irrespective the reason of admission based on previous findings showing that CCI more than 2 is an independent predisposing factor for sepsis (7).
Main exclusion criteria were age less than 18 years; any stage IV malignancy; do not resuscitate decision; active tuberculosis; and pregnancy or lactation. Enrolled patients will be under follow-up by two groups of investigators, namely groups A and B, each being blind to the results of the other group. Group A investigators performed OB PPG recordings every two hours for three consecutive days. The OB was placed on the forefinger and each recording lasted for five minutes. The OB is a patented oximeter-like device that has been developed by Sanmina (Huntsville, AL) and works by measuring optical absorptions using reflectance techniques in five wavelengths i.e. 940nm (IR), 660nm (red color), 530nm (green color), 465nm (blue color) and 395 ± 10 nm (ultraviolet). The ratio of these wavelengths associates with vasoconstriction and vasodilation so as to provide information on the endothelial state. The recorded information was transmitted from the OB to a smartphone and from there to a cloud for data analysis. Group B investigators recorded the following information for three consecutive days: a) vital signs; b) type of infection; c) SOFA score; d) complete blood cell count and differential; e) biochemistry, PCT, CRP and blood gases; and f) microbiology. An amount of 3ml of whole blood was sampled after venipuncture of one antecubital vein under aseptic conditions on the same days. Blood was immediately poured into one sterile and pyrogen-free tube the first three days that was placed on ice. The tube was transported immediately to the lab and centrifuged in 40C at 1,500g. NO was measured in the supernatant by the Griess reaction (Enzo Life Sciences Inc, Farmingdale, NY).
Based on the collected information, enrolled patients were classified into those who eventually developed sepsis during the 3-day intense follow-up and into those who did not develop sepsis. Classification into sepsis required both of the following (2): a) presence of an infection; and b) increase of admission total SOFA score by at least two points.
The primary study endpoint of the first phase of PROUD was the accuracy of the OB for the diagnosis of sepsis at the timepoint of clinical diagnosis using the SOFA score. In order to achieve so, an algorithm that can provide the likelihood for sepsis at each time-point of sampling was developed. The working principle of the OB is emitting light into the local tissue using reflectance PPG techniques for 5 Wavelengths (940, 660, 530, 465, and 395 ±10nm). The OB device samples each wavelength absorbance approximately at 150Hz and then recreates the arterial pulse pressure responses for each wavelength independently. Next, the device analyses the individual PPG wavelengths for each cardiac stroke synchronized to the systolic pulse pressure peak in order to calculate the a/c and d/c components encompassing the systolic and diastolic periods in the sampling window using the related volumetric changes of arterial blood at the specified wavelength dependent tissue depths. The information is subsequently used to calculate a series of parameters to compare the information from a blood analytical and vascular response point-of-view. For the blood analytical series of parameters Logarithmic (L) values, are calculated for each wavelength. Subsequently R values are also calculated using the optical AC amplitude (pulsating PPG arterial signal) compared to the optical “DC” amplitude (non-pulsating arterial, venous, and tissue signals) using the equation R = Iac(lambda 1) / Idc(lambda 1)/ Iac(lambda 2) / Idc(lambda 2). The risk of developing sepsis is aggregated by using a combination of calculations for the algorithm currently proposed including heart rate, relative vessel diameter, metabolites and a combination of L and R values related to NO and to oxygenated hemoglobin. This information generates the optical signatures via compiled Neural network (NN) training vectors. The output of this NN contains two algorithms; one on the confidence percentage the positive likelihood for sepsis; and another on the negative likelihood for sepsis using 30-second sample windows of the optical biosensor data. Both algorithms have values ranging from 0 to 100. For the purpose of analysis, the means of all time readings of each patient were taken into used.
The correlation of the two algorithms was done by the Spearman’s rank of order. In order to evaluate the diagnostic performance of the algorithm, one Receiver Operator Characteristics (ROC) curve analysis was done using the Youden index to identify the best cut-off point for discrimination. Comparisons of quantitative data were done by the Student’s t-test for parametrical data and by the Mann-Whitney U test for non-parametrical data. Comparisons were done by the Fisher exact test for qualitative data. ORs and 95% CIs were calculated by Mantel-Haenszel statistics. Any p value less than 0.05 was considered significant.
The first phase of the study was powered for 139 patients. This was calculated in order to define a cut-off that can discriminate sepsis with 90% specificity with 90% power at the 5% level of significance. To adjust for possible missing values, 150 patients were enrolled.
Second phase of the PROUD study
This phase started after the analysis of the data of the first phase. During this phase, participants were adults of both genders admitted at the emergencies with symptoms compatible with upper or lower respiratory tract infection. Main exclusion criteria were age less than 18 years; any stage IV malignancy; do not resuscitate decision; active tuberculosis; and pregnancy or lactation. All patients were subject to the following interventions: sampling of one nasopharyngeal swab; one single testing with the forefinger OB PPG point-of- care device for five minutes as described above; and one blood draw as described above. The recorded information was stored on a microSD card contained inside the OB device. The local time synchronization and length of test was controlled by the smartphone, the microSD cards were individually retrieved, sterilized and the data was transferred to a storage device for data analysis. Each OB and smartphone was discarded following recording as safety precaution. Sampled swabs were subject to molecular detection of SARS-CoV-2. All patients with COVID-19 were subject to chest X-ray and/or chest computed tomography for the diagnosis of lower respiratory tract infection. Patients negative for SARS-CoV-2 were considered to have “flu-like” symptoms. NO was measured in the blood by the Griess reaction, as described above.
The diagnostic performance of the algorithm developed during the first phase was applied firstly to discriminate between COVID-19 and flu-like symptoms among all participants. It was then used to discriminate between severe and non-severe cases among all COVID-19 cases. Severe COVID-19 was diagnosed according to the WHO classification.