Polymethylmethacrylate-augmented Cannulated Pedicle Screw Fixation for High-energy Vertebral Compression Fracture with Osteoporosis

DOI: https://doi.org/10.21203/rs.3.rs-482277/v2

Abstract

Background: High-energy vertebral compression fractures (HVCFs) with osteoporosis puts forward higher requirement for the stability of pedicle screw internal fixation system. However, few studies have concentrated on the clinical outcomes of cannulated pedicle screw to augment fixation with polymethylmethacrylate (PMMA) under this condition. This study aims to investigate the mid-term efficacy of bone cement-injectable cannulated pedicle screw (CICPS) in HVCFs with osteoporosis.

Methods: Eighteen patients with HVCFs (T < -2.5) were underwent spinal fixation by CICPS from 2012 to 2017. The operation time, blood loss, and hospitalization time were recorded. Pain and functional recovery were evaluated by Visual Analog Scale VAS) and Oswestry Disability Index (ODI), respectively. X-ray films were taken to evaluate the loss of vertebral height, kyphosis angle, pedicle screw loosening, and bony fusion. Surgical related complications were also recorded.

Results: The average follow-up time was 18.5 ± 8.7 months (range, 6-54 months). VAS and ODI scores indicated significant improvements in postoperative pain and activity function (P < 0.05). Comparing the last follow-up with the postoperative imaging findings, the loss of vertebral height and kyphosis angle was 3.4 ± 1.2 mm and 7.8 ± 3.8°, respectively. The rate of bone graft fusion was 100% while no case was involved in screw loosening or extraction. 1 case of superficial infection, 2 cases of PMMA leakage, and 1 case of cerebrospinal fluid leakage were found.

Conclusions: CICPS fixation using PMMA augmentation may be suggested as a feasible surgical technique in osteoporotic patients with HVCFs.

1. background

The incidence of osteoporotic fractures has been increasing rapidly among the ageing population [1], especially among postmenopausal women, the occurrence of which up to 50% [2]. The most common condition, vertebral compression fractures (VCFs) [3, 4], imposes a heavy socio-economic burden for elderly populations. Up to now, percutaneous vertebroplasty (PVP) and kyphoplasty (PKP), the two minimally invasive surgical methods have been widely acknowledged to have an excellent clinical effect on VCF[5, 6].

However, VCF due to high-energy trauma (HVCFs) defined as VCF combined with severe clinical manifestations, such as spinal nerve compression, spinal stenosis, instability, delayed kyphosis and neurological impairment (Kümmell’s disease), should be treated more complexly. In this situation, restoring spinal stability by internal fixation and decompressing spinal canal to relieve the neurological symptoms brings the challenge to surgeons as osteoporosis tends to cause screw loosening and trabecular fracture [7]. Ultimately, inadequate holding power of internal fixation may lead to chronic lower back pain, kyphosis and dysneuria.

Studies revealed that cannulated pedicle screw augmented by polymethylmethacrylate (PMMA) can significantly improve the screw holding power [810]. However, the clinical outcome of cannulated pedicle screw to augment fixation in HVCF with osteoporosis need to be further investigated. In this study, we improved the design and created a new type of PMMA-augmented pedicle screw named bone cement-injectable cannulated pedicle screw (CICPS). To evaluate the curative effect and feasibility of this technique in HVCFs with osteoporosis, we retrospectively reviewed the outcomes of eighteen consecutive patients treated with CICPS, and summarized the incidence of complications and the clinical efficacy of PMMA augmentation.

2. materials And Methods

2.1 Study design

The study was approved by the Daping Hospital ethics committee (IRB, 2019149); and informed consent was obtained from each participant. This study was a retrospective investigation of osteoporotic patients who underwent surgery at our clinic for HVCFs. Considering that using conventional pedicle screws in cases with a severely osteoporotic vertebral body may lead to screw loosening or pulling out, it would be unethical to include a control group for comparison. Instead, we evaluated the clinical outcomes of this technology by comparing the baseline with post-treatment measurement parameters.

2.2 Patients

From October 2012 to May 2017, a total of 30 cases with HVCFs were enrolled. Inclusion criteria were as follows: (1) Diagnosed as osteoporosis by dual-energy X-ray absorptiometry (T-value < -2.5) [10]; (2) HVCFs with acute or chronic impairment of spinal cord function, spinal stenosis or spinal structural instability; (3) Used CICPS for internal fixation; (4) consented the treatment requirement and schedule. Exclusion criteria were as follows: (1) Simple fragility fracture; (2) Old fracture; (3) Not used CICPS for internal fixation; (4) Allergy to any implant or having surgical contraindications. Figure 1 shows the flow chart of patient recruitment and study procedures.

2.3 Materials

The preliminary clinical application of CICPS had achieved satisfactory results in the previous studies performed by authors [8-10]. In references 8 to 10, we had described the design features of CICPS in detail and Figure 2 is a schematic diagram showing the assembly sequence of the system. CICPS is designed as a hollow rod with three side holes at the screw tip so that PMMA can outflow into the centrum through this route. Fixation ability is augmented because of anchoring the screw into cancellous bone with PMMA. In this study, the diameter and length of CICPS were determined according to the measurement results during actual intraoperative conditions. PMMA was a high-viscosity bone cement supplied by Heraeus Holding, Germany. 

2.4 Surgical techniques and postoperative treatment

The screws were manually placed according to the anatomical landmarks. PMMA powder and water agent were mixed, and then, injected by a special device during a dough-like mass viscosity. No more than 2ml of PMMA mixture was determined by intraoperative monitoring of intravertebral PMMA dispersion. The following measures were taken to ensure the augmentation effect and further to reduce the PMMA leakage risk: 1. CICPS suggested to keep parallel to the upper and lower vertebral endplates; 2. The probe was used to directly detect the required screw length so that the outflow side hole of CICPS was located in the anterior 2/5th of the vertebral body; 3. CICPS was not allowed to penetrate the upper, lower, inner and outer sidewalls of the pedicle and the lateral cortex of the vertebral body; and 4. The entry angle of CICPS was slightly larger than that of a conventional pedicle screw [9]. The quantity of CICPSs was determined by the bone quality, the stability requirement, and manual feeling during the tunnel preparation. In principle, a minimum quantity of CICPSs was recommended due to the difficulty in removing and repairing. In this series, most of the cases had instrumented 1-level above and below the fracture. The remaining screws adopted conventional ones. Decompression, fixation and fusion was performed as usual, if necessary.

Antibiotics were routinely used to prevent surgical site infection in the first 48 hours, and the wound drainage tube was removed after 48-72 hours after the amount of drainage fluid less than 50ml. Three days after the surgery, patients were encouraged to perform rehabilitation exercises wearing a thoracolumbar brace.   Anti-osteoporosis treatment should be implemented as soon as possible after operation. Specifically, the vitamin D and calcium intake was estimated at 800 IU/day and 1000 mg/day, respectively. And the intravenous zoledronic acid was supplemented at 5mg/year. 

2.5 Safety and effectiveness evaluation

Operation time, blood loss, and hospitalization time were recorded to evaluate the basic condition of the surgery. A review was carried out at 3 months, 6 months, 1 year, and every half-year after surgery. Moreover, lumbar X-ray films were obtained to evaluate bony fusion, loss of vertebral height, loss of the Cobb angle, and screw loosening or pull-out. The image data were measured by two experienced spine surgeons and the average values were taken. Visual Analog Scale (VAS) and Oswestry Disability Index (ODI) scoring systems were used to evaluate pain and functional recovery in lower limbs, respectively. Complications such as wound infection, cerebrospinal fluid leakage, and PMMA leakage were recorded. 

The method of measuring the loss of vertebral height: 

The vertebral body height measured in sagittal X-film was used to evaluate fracture reduction. Distances were measured between the following points: the most dorsal superior and inferior endplate margins, the most anterior superior and inferior margins, and the height of the centrums from the midpoint of the two lines. The average values were calculated, and vertebral height was determined after taking into consideration the change in vertebral height between the postoperative review and the last follow-up that was regarded as the value of vertebral height loss [10].

The method of measuring the Cobb angle: 

The Cobb angle was used to assess sagittal deformity on a plain radiograph. The upper and lower vertebral bodies near the fractured part with the largest inclination angle were considered, and the angle between the endplate was measured as the kyphosis Cobb angle. when the vertebral bodies that need to be measured had wedge-shaped changes, the adjacent normal endplate with the largest inclination angle is used for measurement  as an alternative. The angle change between the postoperative review and the last follow-up was regarded as the value of Cobb angle loss [10].

Evaluation criteria for spinal fusion on X-ray films  [10,11]: 

(1) Passage of trabecular bone through the bone graft area; (2) Vertebral movement between flexion and extension X-ray film < 3 mm or change in the intervertebral space angle < 5 degrees; and (3) Bone growth through the intervertebral disc space.

Evaluation criteria for screw loosening on X-ray films: 

(1) Screw displacement on a lateral X-ray greater than 1 mm [12,13]; (2) Double-circle sign around the screw on an anterior-posterior X-ray film [14].

2.6 Statistical analysis

The experimental data were calculated as mean ± standard deviation (x ± s) and analyzed by using the SPSS 13.0 statistical software. The paired test was conducted to compare variables, including VAS and ODI scoring and imaging indicators at the pre-operation, post-operation, and at the last follow-up. The change was considered statistically significant if P value was less than 0.05.

3. results

Twenty cases were enrolled and the follow-up rate at each time point was 90% because 2 cases were lost at three months after operation. Finally, 18 cases were completely recorded for an average follow-up time of 18.5 ± 8.7 months (range, 6-54 months). These cases comprised 6 males and 12 females with an average age of 72.3 ± 6.5 years ranging from 56 to 80 years. The average duration from injury to admission was 20 ± 13.5 days (range, 3 hours to 3 months). 

The operation time was 165 ± 23 min (range, 95-462 min), the bleeding amount was 206 ± 75 ml (range, 50-400 ml) and the length of the hospitalization time was 12.5 ± 3.5 d (range, 9-22 d). There were no death, surgical related neurological dysfunction and paraplegia in all cases at each follow-up time point. One case of superficial surgical site infection was noted, which was under control after an intravenous drip of antibiotics. Cerebrospinal fluid leakage occurred in one case, which healed completely after the drainage tube removed following two week of bed rest. With respect to PMMA leakage, there was one case leakage into the paravertebral vein, and another case into the intervertebral disc space during the operation, but no serious complication such as nerve or blood vessel injury or pulmonary embolism was observed (Table 1).

Table-1 

Baseline characteristics and clinical parameters of the 18 patients with high-energy vertebral compression fracture (HVCFs)  

Variable

Value


Average age (Y)

72.3 ± 6.5 (56-80)


Height (cm)

162.3 ± 5.9 (149-175)


Weight (kg)

50.3 ± 12.3 (44-78)


Follow-up time (M)

18.5 ± 8.7 (6-54)


Pre-hospital time (d)

20 ± 13.5 (3h-3M)


Gender (male: female)

6:12


Average BMD (T-value)

-3.20 ± 0.57 (-2.5~-5.2)


Diagnosis

 


HVCFs with spinal stenosis

9/18 (50%)



HVCFs with kyphosis

4/18 (22.2%)



HVCFs with delayed kyphosis and neurological impairment

5/18 (27.8%)


Operation time (min)

165 ± 23 (95-462)


Blood loss (ml)

206 ± 75(50-400)


Hospitalization time (d)

12.5 ± 3.5 (9-22)


Number of loosening CICPSs / total number of CICPSs used

0/40


Complication

 


Superficial infection

1/4 (25%)


PMMA leakage

2/4 (50%)


Cerebrospinal fluid leakage

1/4 (25%)


Fusion rate

100%


 

3.1 Imaging evaluation results

The height of the compressed vertebral body and the kyphosis angle were significantly improved (P < 0.05). The vertebral height changed from 16.3 ± 5.6 mm before operation to 22.7 ± 4.1 mm after operation, and to 19.3 ± 3.9 mm at the last follow-up; the kyphosis angle changed from 35.6 ± 10.4 degree to 18.5 ± 8.9 degree, and to 26.3 ± 9.3 degree at these three time points. At the last follow-up, the average values of vertebral height decreased and kyphosis angle increased at the same time. However, comparing these indexes at the date of post-operation with that of last follow-up, there was no significant difference in vertebral height (P > 0.05), while the loss of kyphosis angle changed significantly (P < 0.05) (Table-2). No screw loosening and fusion failure were observed in all cases according to the criteria. 

Table-2 

Height of the compressed vertebral body, Cobb angle, VAS and ODI score in patients before and after treatment.

Index

Pre-operation

Post-operation

Last follow-up

Vertebral height (mm)

16.3 ± 5.6

22.7 ± 4.1

19.3 ± 3.9b

Cobb angle (degree)

35.6 ± 10.4

18.5 ± 8.9

26.3 ± 9.3a

VAS score

8.3 ± 2.6

2.2 ± 0.8

1.8 ± 0.3b

ODI score (%)

75.5 ± 12.2

35.2 ± 7.2

30.5 ± 13.1b

Note: All indexes at the post-operation and last follow-up showed significant differences from those at the pre-operation, p<0.05. a, p<0.05, a significant difference was observed in Cobb angle between the post-operation and the last follow-up. b, p>0.05, no significant difference was observed in the vertebral height, VAS score, and ODI score between the post-operation and the last follow-up.

3.2 Pain relief and functional recovery

The VAS score improved from 8.3 ± 2.6 before operation to 2.2 ± 0.8 after operation, and to 1.8 ± 0.3 at last follow-up. The ODI score improved from 75.5 ± 12.2% to 35.2 ± 7.2%, and to 30.5 ± 13.1% at these three time points. These indexes were significantly improved after operation (P < 0.05), but no significant difference was observed when comparing post-operation with last follow-up (P > 0.05) (Table-2).

3.3 Typical cases

Typical case 1 (Figure 3): A 68-year-old female patient (T-value = -3.0), who had suffered from traumatic fracture of the T12 vertebral body 2 months ago, was diagnosed as HVCF and kyphosis, and she did not receive any bed rest and brace treatment. Imaging examinations showed wedge-shaped deformation of the T12 vertebral body and thoracolumbar kyphosis (Figure 3A-D). Thus, CICPS was used to strengthen the internal fixation, and the kyphosis was corrected after distraction and reduction (Figures 3E and F). Pouring PMMA with PVP device was performed to maintain vertebral height. 6 and 12 months after operation, bony fusion was achieved with maintenance of the spinal curvature and vertebral height, and there was no PMMA displacement or screw loosening (Figure 3G-J).

Typical case 2 (Figure 4): A 73-year-old female patient (T-value = -3.5) suffered from L1 HVCF, kyphosis, and cauda equina injuries due to a fall one month ago (Figure 4A-D). She was treated with spinal canal decompression, posterolateral arthrodesis, and CICPS-augmented internal fixation of upper and lower vertebral bodies (Figure 4E and F). The symptoms were then relieved after operation and CICPS loosening was not detected at the 18-month follow up. Although the vertebral height was reduced, the kyphosis angle was maintained and bony fusion was successfully achieved (Figure 4G and H). PMMA was leaked into the anterior venous plexus at L1 without any symptom.

4. Discussion

Known as microinvasive technologies, PVP and PKP, is considered as the "gold standard" for the treatment of osteoporotic VCFs [5, 6, 15]. However, PVP and PKP may no longer be applicable when VCFs accompanied with severe high-energy trauma. Some special reasons are as follows: 1. High-energy trauma may cause spinal canal stenosis due to the vertebral posterior wall fracture. Symptoms of neurological impairment usually require spinal decompression [16, 17]. 2. Internal fixation must be performed to restore spinal stability because of the three-column fracture. 3. Delayed neurological symptoms caused by the injured vertebral wedging may require anterior column resection and reconstruction [18, 19]. However, osteoporosis significantly reduces the internal fixation stability in HVCFs patients with a failure rate up to 12% [20].

Recently, PMMA-augmented cannulated pedicle screw has been considered as an optional technique for broad application prospects [21, 22]. Some researchers recommended this type of pedicle screw for lumbar spondylolisthesis with osteoporosis [23, 24]. Meanwhile, Sawakami et al. [25], Park et al. [26] and Cho et al. [27] have achieved satisfactorily clinical efficacy in pseudoarthrosis following VCFs and Kummell’s disease. But few studies concentrated on the clinical outcome of PMMA-augmented cannulated pedicle screw in HVCFs. Previously, in vitro study showed that CICPS could significantly improve the biomechanical stability of pedicle screws, and preliminary clinical application suggested that CICPS had satisfactory reliability and safety in osteoporotic patients with lumbar spondylolisthesis and thoracolumbar VCF (AO Type A) [810, 28]. This clinical study confirmed that CICPS can also obtain satisfactory effect in patients with HVCFs.

In current study, the VAS and ODI score significantly improved after operation immediately and at the last follow-up compared with those before operation. Additionally, there are no significant differences between VAS and ODI scores immediately after operation and at the final follow-up more than 6 months. The results indicate that satisfactory mid-term clinical outcomes can be achieved in this group. Singh V et al. [29] did a systematic analysis for PMMA-augmented cannulated pedicle screw. The published studies showed the average VAS score before operation was 8.4 (range 8-9.2) compared to 2.3 (range 1.42-4.8) at the last follow up. The average improvement ODI for assessment of functional recovery was 42.1. In this study, VAS score before operation was 8.3 compared with 2.2 at last follow-up and ODI improved from 30.5 to 75.5. Therefore, the results obtained in our series are similar to those of previous studies.

PMMA leakage is a focus point to surgeons because neurological or spinal cord damage may cause by the situation [12, 18, 30]. Traditionally, PMMA is injected into the preset nail tunnel using a PVP device during laddering period, and then the screw is inserted into the nail tunnel [25]. The challenges of this procedure lie in uncontrollable diffusion of PMMA during the screw insertion process. Even if PMMA is removed immediately when the spinal canal leakage occurs, these surgical procedures will lead to the time window of PMMA injection being exceeded. Therefore, we chose to further investigate CICPS instead of traditional injection method. The advantage is that screws are left in situ and monitored by intraoperative X-ray during PMMA injection, which can reduce the leakage risk due to pressurised cement injection. Moreover, compared with common screws, the design of side holes located at the screw rod can also play an important role to avoid PMMA leakage [810].

CICPSs were used in 18 patients with HVCFs and osteoporosis. 3 screws in 2 patients (3/40, 7.5%) were observed PMMA leakage, but without any neurological symptom. Wuisman et al. [31] reported four out of 49 screws (8.2%) had PMMA leakage into spinal canal using traditional technique and removed PMMA by laminectomy and dura manipulation. Janssen et al. [32] reported that the incidence of asymptomatic PMMA leakage was 66.7% and that of symptoms was 5.5%, in 1.2% of which needed to remove the PMMA and screws. Martı´n-Ferna´ndez et al. [33] observed a 62.3% incidence of PMMA leakage in 313 patients. 1.55% of the cases had symptoms such as radicular pain of lower limbs and transient sensorimotor deficit. In brief, PMMA leakage presented no symptoms or only mild symptoms in most studies and the leakage rate showed a big difference, the reason of which might be related to the small sample size or different screw design in different studies.

In our series, the loss of kyphosis cobb angle increased significantly comparing these indexes at the date of post-operation with that of last follow-up. While there is no significant differences in the vertebral height and no screw loosening was observed. The reason for this phenomenon can be explained by physiological progress. Studies have shown that kyphosis progressed an average 7 degree over 15 years in healthy women, especially between age of 50 and 59 [34, 35]. In thoracolumbar burst fracture following internal fixation without PMMA augmentation, progressive kyphosis was also observed [36]. Therefore, trauma-related disc apoptosis or osteoporosis process [3739] may play important roles in kyphosis progression spite of PMMA augmentation or not. Besides high-energy fractures are often accompanied by loss of sagittal balance of the spine, which may be another reason for the continued loss of Cobb angle of the spine compared to the postoperative correction angle. From a biomechanical perspective, postoperative stress concentration in the the upper vertebra of the uppermost fixed vertebra may also lead to progressive loss of Cobb angle and even fracture.

In one study of Huang et al [30], they observed the clinical outcomes of traditional PMMA augmentation technology in 28 cases of Kummell’s disease. The operation time, blood loss and hospitalization time were 115.9±10.0 minutes, 214.3±17.5 ml and 7.6±1.3 days respectively. While in this study, the corresponding data were 165 ± 23 minutes, 206 ± 75 ml and 12.5 ± 3.5 days. Furthermore, 3 patients had asymptomatic PMMA leakage, one developed a urinary tract infection, one contracted pneumonia, and one experienced delayed wound healing. There was no screw loosening at the final follow-up. Comparing these data with our findings, we cannot draw any better conclusions by comparing these data with our findings. However, the absolute values of the data between CICPS and the traditional PMMA augmentation technology only have small differences, which may be related to the technical proficiency of surgeons and postoperative management procedures.

This study also has several limitations. Firstly, this study is not a randomized controlled study, contributing to a low level of evidence. Secondly, the sample size is small and the follow-up time is short, thus a study with more patients and longer follow-up duration is required to further confirm our results. Thirdly, since the safety and effectiveness of CICPS have been biomechanically evaluated before [8], only clinical evaluation was performed in this study. Finally, standard application protocol of CICPS should be developed to avoid the catastrophic consequences from excessive application, and it will be our future research focus.

5. conclusion

In conclusion, CICPS using PMMA augmentation maintains their functional outcomes after internal fixation at latest follow-up in spite of kyphosis progression. The incidence of asymptomatic PMMA leakage, screw loosening and complications are not higher than that of previous reports. CICPS may be a clinically safe and effective technology which can be used to augment pedicle screws in HVCFs with osteoporosis. This study provides an evidence-based basis for developing the guidelines of the cannulated pedicle screw application, especially in HVCFs with osteoporosis.

Abbreviations

HVCF: High-energy vertebral compression fracture

CICPS: Cement-injectable cannulated pedicle screw

PMMA: Polymethylmethacrylate

VAS: Visual analog scale

ODI: Oswestry disability index

PVP: Percutaneous vertebroplasty

PKP: Percutaneous kyphoplasty

VCF: Vertebral compression fracture

Declarations

Ethics approval and consent to participate

The study was approved by the Daping Hospital ethics committee (IRB, 2019149).

Consent for publication

All data published here are under the consent for publication. Written informed consent was obtained from all individual participants included in the study.

Data availability

The datasets generated and analyzed during the present study are available from the corresponding author on reasonable request. 

Conflict of Interest

The authors declare that they have no conflict of interest.

Funding

This work was supported by grants from the National Natural Science Foundation of China (No. 81902257) and our University Project of Medical Elite Training (2018XLC2024 and 2019CXLCB015).

Author contributions

YYL and JX conducted the experiments and wrote the manuscript, XY, MYL, ZW and LH collected and analyzed the data. PL and FD supervised the project. All authors read and approved the final manuscript.

Acknowledgements

Not applica

References

  1. Sozen T, Ozisik L, Basaran NC. An overview and management of osteoporosis. European journal of rheumatology. 2017;4(1):46–56.
  2. Porter JL, Bhimji SS. Osteoporosis. Treasure Island.FL: StatPearls Publishing; 2017. Available at:https://www.ncbi.nlm.nih.gov/books/NBK441901/.Accessed October 20, 2017.
  3. Lee YK,Jang S,Jang S,Lee H J,Park C,Ha YC, et al. Mortality after vertebral fracture in Korea: analysis of the National Claim Registry. Osteoporosis international: a journal established as result of cooperation between the European Foundation for Osteoporosis and the National Osteoporosis Foundation of the USA. 2012;23(7):1859–1865.
  4. Mayumi S,Naoto E,Takeo O, et al. Incidence of osteoporotic fractures in Sado, Japan in 2010. Journal of bone and mineral metabolism. 2014;32(2):200–205.
  5. Kong LD, Wang P, Wang LF, et al. Comparison of vertebroplasty and kyphoplasty in the treatment of osteoporotic vertebral compression fractures with intravertebral clefts. European journal of orthopaedic surgery & traumatology: orthopedie traumatologie. 2014;24 Suppl 1:S201-208.
  6. Hartmann F, Gercek E, Leiner L, et al. Kyphoplasty as an alternative treatment of traumatic thoracolumbar burst fractures Magerl type A3. Injury. 2012;43(4):409–415.
  7. Patel V, Issever AS, Burghardt A, et al. MicroCT evaluation of normal and osteoarthritic bone structure in human knee specimens. Journal of orthopaedic research: official publication of the Orthopaedic Research Society. 2003;21(1):6–13.
  8. Liu Y, Xu J, Sun D,et al. Biomechanical and finite element analyses of bone cement-Injectable cannulated pedicle screw fixation in osteoporotic bone. Journal of biomedical materials research Part B, Applied biomaterials. 2016;104(5):960–967.
  9. Dai F, Liu Y, Zhang F, et al. Surgical treatment of the osteoporotic spine with bone cement-injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients. Clinics. 2015;70(2):114–119.
  10. Rong Z, Zhang F, Xiao J, et al. Application of Cement-Injectable Cannulated Pedicle Screw in Treatment of Osteoporotic Thoracolumbar Vertebral Compression Fracture (AO Type A): A Retrospective Study of 28 Cases. World neurosurgery. 2018;120:e247-e258.
  11. Lowe TG, Tahernia AD, O'Brien MF, et al. Unilateral transforaminal posterior lumbar interbody fusion (TLIF): indications, technique, and 2-year results. Journal of spinal disorders & techniques. 2002;15(1):31–38.
  12. Wu JC, Huang WC, Tsai HW, et al. Pedicle screw loosening in dynamic stabilization: incidence, risk, and outcome in 126 patients. Neurosurgical focus. 2011;31(4):E9.
  13. Ko CC, Tsai HW, Huang WC,et al. Screw loosening in the Dynesys stabilization system: radiographic evidence and effect on outcomes. Neurosurgical focus. 2010;28(6):E10.
  14. Dakhil-Jerew F, Jadeja H, Cohen A, et al. Inter-observer reliability of detecting Dynesys pedicle screw using plain X-rays: a study on 50 post-operative patients. European spine journal: official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2009;18(10):1486–1493.
  15. Hsieh MK, Chen LH, Chen WJ. Current concepts of percutaneous balloon kyphoplasty for the treatment of osteoporotic vertebral compression fractures: evidence-based review. Biomedical journal. 2013;36(4):154–161.
  16. Tang H, Zhao JD, Li Y, et al. Efficacy of percutaneous kyphoplasty in treating osteoporotic multithoracolumbar vertebral compression fractures. Orthopedics. 2010;33(12):885.
  17. Gan M, Yang H, Zhou F, et al. Kyphoplasty for the treatment of painful osteoporotic thoracolumbar burst fractures. Orthopedics. 2010;33(2):88–92.
  18. Yang H, Pan J, Sun Z, et al. Percutaneous augmented instrumentation of unstable thoracolumbar burst fractures: our experience in preventing cement leakage. European spine journal: official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2012;21(7):1410–1412; author reply 1413.
  19. McLain RF, Burkus JK, Benson DR. Segmental instrumentation for thoracic and thoracolumbar fractures: prospective analysis of construct survival and five-year follow-up. The spine journal: official journal of the North American Spine Society. 2001;1(5):310–323.
  20. Lonstein JE, Denis F, Perra JH, et al. Complications associated with pedicle screws. The Journal of bone and joint surgery American volume. 1999;81(11):1519–1528.
  21. Frankel BM, Jones T, Wang C. Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery. 2007;61:531–537; discussion 537-538.
  22. Fransen P. Increasing pedicle screw anchoring in the osteoporotic spine by cement injection through the implant. Technical note and report of three cases. J Neurosurg Spine. 2007;7:366–369.
  23. Jang SH, Lee JH, Cho JY, et al. The efficacy of hydroxyapatite for screw augmentation in osteoporotic patients. Neurol Med Chir. 2013;53:875–881.
  24. Chandra Vemula VR, Prasad BC, Jagadeesh MA, et al. Minimally invasive transforaminal lumbar interbody fusion using bone cement-augmented pedicle screws for lumbar spondylolisthesis in patients with osteoporosis. Case series and review of literature. Neurol India. 2018;66:118–125.
  25. Sawakami K, Yamazaki A, Ishikawa S, et al. Polymethylmethacrylate augmentation of pedicle screws increases the initial fixation in osteoporotic spine patients. J Spinal Disord Tech. 2012;25:E28-E35.
  26. Park SJ, Kim HS, Lee SK, et al. Bone cement-augmented percutaneous short segment fixation: an effective treatment for Kummell’s disease? J Korean Neurosurg Soc. 2015;58:54–59.
  27. Cho Y. Posterior vertebrectomy and circumferential fusion for the treatment of advanced thoracolumbar Kummell disease with neurologic deficit. Asian Spine J. 2017;11:634–640.
  28. Liu Y, Xiao J, Yin X,Liu M;Zhao J;Liu P,et al. Clinical efficacy of bone cement-injectable cannulated pedicle screw short segment fixation for lumbar spondylolisthesis with osteoporosise. Sci Rep. 2020;10(1):3929
  29. Singh V, Mahajan R, Das K, et al. Surgical trend analysis for use of cement augmented pedicle screws in osteoporosis of spine: a systematic review (2000-2017). Global Spine J. 2019;9:783–795.
  30. Rohith Kumar R, Ashok Kumar S, Periyasami Bhuvana K. Preparation and evaluation of bioactivity of porous bioglass tablets for bone tissue regeneration. Sci Medicine J. 2019;1(3):112–123.
  31. Wuisman PI, Van Dijk M, Staal H, et al. Augmentation of (pedicle) screws with calcium apatite cement in patients with severe progressive osteoporotic spinal deformities: an innovative technique. Eur Spine J. 2000;9:528–533.
  32. Janssen I, Ryang YM, Gempt J, et al. Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine. Spine J. 2017;2:837–844.
  33. Martı´n-Ferna´ndez M, López-Herradón A, Piñera AR, Tomé-Bermejo F, et al. Potential risks of using cement-augmented screws for spinal fusion in patients with low bone quality. Spine J. 2017;17:1192–1199.
  34. Kado DM, Miller-Martinez D, Lui LY, et al. Hyperkyphosis, kyphosis progression, and risk of non-spine fractures in older community dwelling women: the study of osteoporotic fractures (SOF). J Bone Miner Res. 2014;29:2210–2216.
  35. Ball JM, Cagle P, Johnson BE, Lucasey C, et al.. Spinal extension exercises prevent natural progression of kyphosis. Osteoporos Int. 2009;20:481–489.
  36. Lakshmanan P, Jones A, Mehta J, et al. Recurrence of kyphosis and its functional implications after surgical stabilization of dorsolumbar unstable burst fractures. Spine J. 2009;9:1003–1009.
  37. Lin HH, Chang MC, Wang ST, et al. The fates of pedicle screws and functional outcomes in a geriatric population following polymethylmethacrylate augmentation fixation for the osteoporotic thoracolumbar and lumbar burst fractures with mean ninety five month follow-up. International Orthopaedics. 2018;42(6):1313–1320.
  38. Zhang SJ. Clark M, Liu XL. The effects of bio-inspired electromagnetic fields on healthy enhancement with case studies. Emerging Science J. 2019;3(6):369–381.
  39. Swati P, Nikita S. Effect of body positions on quadriceps angle measurement. Sci Medicine J. 2019;1(1):20–2.