Automated Bag Breathing Unit for COVID-19 Ventilator Shortages

DOI: https://doi.org/10.21203/rs.3.rs-488517/v1

Abstract

Background: The COVID-19 pandemic has caused a global mechanical ventilator shortage for treatment of severe acute respiratory failure. Development of novel breathing devices has been proposed as a low cost, rapid solution when full-featured ventilators are unavailable. Here we report the design, bench testing and preclinical results for an 'Automated Bag Breathing Unit' (ABBU). Output parameters were validated with mechanical test lungs followed by animal model testing.

Results:  The ABBU design uses a programmable motor-driven wheel assembled for adult resuscitation bag-valve compression. ABBU can control tidal volume (200-800 ml), respiratory rate (10-40 bpm), inspiratory time (0.5-1.5 sec), assist pressure sensing (-1 to -20 cm H2O), manual PEEP valve (0- 20 cm H2O).   All set values are displayed on an LCD screen. Bench testing with lung simulators (Michigan 1600, SmartLung 2000) yielded consistent tidal volume delivery at compliances of 20, 40 and 70 (mL/cm H2O). The delivered fraction of inspired oxygen (FiO2) decreased with increasing minute ventilation (VE), from 98% to 47% when VE was increased from 4-16 L/min using a fixed oxygen flow source of 5 L/min.

ABBU was tested in Berkshire pigs (n=6, weight of 112±5.8 lb) utilizing normal lung model and saline lavage induced lung injury. Arterial blood gases were measured following changes in tidal volume (200-800 ml), respiratory rate (10-40 bpm), and PEEP (5-20 cm H2O) at baseline and after lung lavage. Physiological levels of PaCO2 (≤40 mm Hg [5.3 kPa]) were achieved in all animals at baseline and following lavage injury. PaO2 increased in lavage injured lungs in response to incremental PEEP (5-20 cm H2O) (p<0.01). At fixed low oxygen flow rates (5 L/min), delivered FiO2 decreased with increased VE.

Conclusions: ABBU provides oxygenation and ventilation across a range of parameter settings that may potentially provide a low-cost solution to ventilator shortages. A clinical trial is necessary to establish safety and efficacy in adult patients with diverse etiologies of respiratory failure.

Background

On January 31, 2020, the US Department of Health & Human Services announced a public health emergency related to a novel coronavirus, SARS-CoV-2, and the disease it causes, COVID-19 (1). The early rapid spread of the COVID-19 pandemic resulted in a shortage of mechanical ventilators and accessory components (e.g., humidifiers, circuits, etc.) in many regions throughout the world (2-5). In response to these shortages, a global surge in development and production occurred, including repurposing non-medical device assembly lines to manufacture quickly designed ventilators (e.g., FORD, GM, Virgin, etc.) (6-10).

As of March 2021, over 150 million COVID-19 cases have been identified leading to over 3.0 million deaths worldwide (11). Among hospitalized patients, 30% require care at intensive care unit (ICU) and 29% or more of those require mechanical ventilation (12).

In response to the shortage of mechanical ventilators to treat COVID-19 patients, resuscitation bag-valve breathing devices were conceived as a potential solution for short-term emergency use. The FDA has classified these devices as "emergency resuscitators" to distinguish them from mechanical ventilators (13-18). Our design uses a self-inflating resuscitation bag-valve, an automobile windshield motor, and lever arm to mimic manual hand bag-valve ventilation - along with essential operator controllable parameters:  tidal volume (VT), respiratory rate (RR), inspiratory time (TI), positive end-expiratory pressure (PEEP) and patient-initiated breath pressure sensing. ABBU uses readily available components, low flow O2 sources, standard electrical power, and can be rapidly mass produced at low cost.

The purpose of this study was to determine if ABBU can provide oxygenation and ventilation in a mechanical test lung and preclinical porcine model across a range of clinically relevant parameter settings.

Methods

  1. Design: mechanical, electrical, software, safety

ABBU was designed to replace manual ventilation of a bag valve resuscitator when a conventional ventilator device is unavailable (Fig. 1A-B). ABBU features include VT (200-800 mL), RR (10-40 bpm), TI (0.5-1.5 s), and adjustable patient-initiated breath sensing (-1 to -20 cm H2O). ABBU can use low flow oxygen (5-15 L/min) from widely available sources (e.g. concentrators, hospital wall-source, tanks, and liquid oxygen reservoirs).

ABBU senses the patient inspiratory effort below a software-calculated pressure threshold to trigger a breath. Patient-ABBU synchrony is facilitated by clinician titration of the triggering threshold. Auto-cycling can be reduced by increasing the triggering threshold (e.g. more negative). Conversely, ineffective triggering is reduced by decreasing the sensing threshold (e.g. less negative). Patients unable to trigger breaths (e.g., weakness, neuromuscular blocker-induced paralysis, central apneas) receive mandatory breaths at the set VT, RR, and TI. Total RR will be determined by the patient-triggered rate and the set rate.

ABBU provides visual and auditory alarms for circuit blockage, air-leaks, low pressure (e.g. disconnection), high airway pressure (50-70 cm H2O), motor, and electric failure. The audible power loss alarm has a backup battery. A high visibility enclosure facilitates rapid troubleshooting of the circuit and motor-bag interface (Fig. 1B). If ABBU fails, clinicians can quickly open the enclosure to access the bag and provide manual ventilation (Fig. 1C). This capability is a key safety feature of the ABBU design.

Figure 1D shows the breathing circuit components used in animal experiments. The patient exhalation valve (CPR-2 bag, Mercury Medical, Clearwater, FL) includes a manual adjustable PEEP valve. The resuscitator bag (adult Ambu® Spur® II bag, AMBU Inc., Columbia, MD) is centered in a cradle and secured on both ends by an elastic cord inside the unit. The bag PEEP valve (Ambu Disposable PEEP Valve, 0-20 cm H2O size) is set to 0 cm H2O, and PEEP is adjusted manually on a second PEEP valve interfaced to the patient exhalation valve. Two pressure sensing lines (3/16” ID, 22mm OD AirLife connector) are used for circuit pressure monitoring and breath triggering assist. An FiO2 analyzer (MaxO2+AE, Maxtec, Salt Lake City, UT) was interfaced into the breathing circuit for all animal experiments.

Bench testing

A Michigan adult dual-lung simulator (Model 1600, Michigan Instruments, USA) and Ventilator Validation System (VVK100-SYS, BIOPAC Systems, Goleta, CA) was used to validate ventilator parameters. ABBU was tested at compliances of 20, 40 and 70 mL/cm H2O with resistances 5-50 cm H2O/(L/s). VT of 200, 400, 600 and 800 mL were tested across a range of compliances and resistance levels. For performance experiments, RR was set at 15 bpm and the PEEP was set at 15 cm H2O. Twenty breath cycles were collected for each measurement and processed to validate measures of VT, pressure, TI, RR, and confirm PEEP value.

Durability and VT stability of different brand resuscitation bags:  AMBU (SPUR II, Ambu, Columbia, MD), HUDSON (RCI 5387, Teleflex, Morrisville, NC), MEDLINE (CPRM1116, Medline, Northfield, IL), Mercury (CPR-2, Mercury Medical, Clearwater, Florida) were evaluated on mechanical test lungs (SmartLung 2000, IMT Analytics, Buchs, Switzerland) at maximum RR (50 bpm) and TI of 0.5s continuously over 7 days. Cardone electric motors (Model 85-3024, Cardone Industries, Ontario, CA) were operated continuously for > 30 days to assess durability.

Animal testing

Animal procedures were approved by the Institutional Animal Care and Use Committee at the University of Texas Health Science Center at San Antonio. Studies were performed on 6 healthy female pigs (Berkshire, 112±5.8lb). Pigs were sedated via Telazol (4–8 mg/kg IM), Xylazine (1–2.2 mg/kg IM), and 3–4% Isoflurane, followed by endotracheal intubation and maintenance on 0.5–3% Isoflurane. Body temperature was kept in the normal range (38-39oC) by heated pad. Arterial pulse pressure was monitored by a micromanometer pressure sensor in the descending thoracic aorta. After collection of baseline blood samples, ABBU was connected to the proximal end of the endotracheal tube by a 90-degree adapter plugged into the breathing circuit, which included the FiO2 analyzer and side-stream ETCO2 analyzer (Fig.1D).

ABBU settings were changed in accordance with the experimental protocol. TI was kept constant (1s) during the entire experiment. Baseline testing was performed on healthy lungs, followed by testing on saline injured lungs. Neuromuscular paralysis was used as needed (vecuronium, IV, 0.1-0.2 mg/kg). Heart rate (HR), blood pressure (BP) and body temperature (rectal) were monitored continuously. Pigs were euthanized (Euthasol, IV, 100 mg/kg) following completion of experiments (6-8 hours).  

Saline lung lavage was performed as previously described (19). In brief, warmed saline (30 mL/kg) was poured into the lungs through a funnel.  As arterial pressure fell below 50 mm Hg, lavage fluid was drained passively. The animal was reconnected to ABBU with an O2 flow rate of 15 L/m and RR was adjusted to maintain arterial pH>7.25. Lavages were repeated until partial pressure of oxygen (PaO2) was < 100 mm Hg [13.3 kPa] for 30 min.

Arterial blood samples were analyzed by CG4+ cartridges (iSTAT analyzer, Abbott, IL, USA). Blood gas responses for different VT, RR and PEEP were compared with their respective baselines for normal and lung injury model. Parameters: FiO2 (%), HR (bpm), RR (bpm), ETCO2 (mm Hg), SpO2 (%), and BP (mm Hg), were recorded concurrent with blood sample collection.

Statistical analysis

Data in graphs is shown as mean ± SE. Two-tail T test and one-way ANOVA were used for all comparisons. A value of p < 0.05 was considered statistically significant. Pearson Correlation Coefficient (PCC) was computed to test correlation between two variables.

Results

1. Bench testing

1.1 Durability testing of bags and motors. Three Cardone motors were continuously operated at constant VT (800 mL) and RR (50 bpm) for 32 days. The motor temperature was monitored. The testing was discontinued after Motor 1 overheated (up to 73oC) and stopped operating (Fig. 2A).

Additionally, the performance of four brands of bags (AMBU Spur II, Hudson, Medline, Mercury) was tested with ABBU running continuously for 7 days at VT 800 ml and RR 50 bpm. Data for each bag was collected at VT of 400 mL. The correlation slope and standard deviation were calculated to indicate when bag performance started to decline. A slope correlation closest to 1 indicates bag is able to achieve the targeted VT for all parameters in the test matrix. The AMBU Spur II bag performed the best (correlation slope between 0.93-1.03, data not provided). Figure 2B shows experiments with seven Spur II AMBU bags. Values computed from the BIOPAC data were consistent with the set controls on the ABBU instrument. The data demonstrates the consistent performance of AMBU Spur II bag (400 mL VT, 15 bpm RP, 1 sec TI) over seven days of continuous ABBU operation. At day 4 there was a significant variation in VT resulting from a shift of bag position in the cradle. AMBU bag degradation (scratches, loss of elasticity, loss of compliance) was observed after continuous operation (RR 50 bpm, TI-0.5 sec, VT 800 mL) starting at day 4, however performance was still adequate. Performance declined after 7 days, after which replacing the bag is recommended.

1.2 Accuracy of controls of instruments. Targeted VT of 400 mL was consistently delivered (SD ≤ 50 mL) to the Michigan test lung at varying compliances (20, 40, 70 mL/cm H2O) and TI settings (0.5, 1 s) (Fig. 2C, D).

PEEP and PIP showed minimal variation during continuous operation of ABBU for 7 days at VT of 800 mL and RR of 50 bpm (Fig. 2E).

FiO2 decreased significantly with increasing RR at a constant VT (400 mL) and fixed oxygen flow rate of 5 L/m (Fig. 2F).

2. Animal testing

ABBU was tested in a porcine model at baseline and following saline lavage lung injury. Data was obtained sequentially on the same animal over 6-8 hours. HR, BP and body temperature were maintained within physiological levels throughout experimentation.

2.1 Normal lung. Before experiments, the blood gas and hemodynamic responses from switching to ABBU were compared with the veterinary ventilator baseline (Narkomed 2B, Drager, Germany).  Switching from the veterinary ventilator (FiO2 100%) to ABBU (FiO2 73%) caused a decrease in PaO2 from 467.5 ± 25.8 mm Hg [62.3±3.4 kPa] to 307.3±51.9 mm Hg [41±6.9 kPa] attributed to the difference in FiO2 between the two devices (Table 1). The blood gas responses at different VT and RR were compared with their respective baseline values (Fig. 3) at constant TI (1 s), PEEP (5 cm H2O), and flow rate (5 L/min). Mean PaCO2 in arterial blood at baseline VT (200 mL) was 74.8±3.8 mm Hg [10 ±0.5 kPa]. Figure 3A demonstrates that increasing VT and with fixed RR at 20 bpm, lowered PaCO2 to physiological level (≤40 mm Hg [5.3 kPa]) in all animals (p<0.001). Changes from high to low PaCO2 (e.g., low to high VE) was associated with significant decreases in FiO2 (0.998 PCC) similar to test lung data using a fixed oxygen flow rate. Increases in VE decreased delivered FiO2 at all combinations of VT and RR.  The response of PaCO2 to changes in RR while keeping VT constant is shown in Figure 3B.

2.2 Saline lavage lung injury. Hypoxemia following saline lavage was confirmed in all six pigs. The average arterial PaO2, PaCO2 and pH at baseline was 69.5±8.6 mm Hg [9.3±1.1 kPa], 43.1±2.2 mm Hg [57.5±0.3 kPa] and 7.4±0.02, respectively.

Similar to pre-lavage baseline, an increase in VT while maintaining a constant RR (20 bpm) led to a significant decrease in arterial PaCO2 level (p<0.005) and FiO2. Mean PaCO2 (mm Hg [kPa]) at 200 mL, 400 mL and 600 mL was 72.7±7.1 [9.7±0.9], 63.7±4.1 [8.5±0.5] and 45.3±3.8 [6.0±0.5], respectively. Increasing RR, demonstrated a similar decline in PaCO2 (p<0.001) and decrease in FiO2 (0.984 PCC).

In 5 out 6 pigs, increasing PEEP effectively improved oxygenation of the saline injured lung (Fig. 3E). Incremental PEEP steps from 5-20 cm H2O led to significant increase in PaO2 from 62.3±7.4 mm Hg [8.3±0.99 kPa] to 287.5 ±14.2 [mm Hg 38.3±1.9 kPa] (p<0.01).

2.3 ABBU synchronous operation testing. A breath-triggering software algorithm was developed during preliminary pig studies (data not included). A “true positive is defined as the patient initiating a breath (e.g. triggered-assisted breath) and ABBU delivering an assist within 160 ms. A “true negative” (e.g. control breath) is defined as the patient not initiating a breath and ABBU delivering a breath according to the rate setting. A “false positive” is defined as ABBU delivering a breath at a time the patient did not initiate (e.g. false triggering or auto-cycling). A “false negative (e.g., ineffective triggering) is defined as the patient initiating a breath and ABBU not delivering an assist. The final version of the ABBU triggering algorithm (e.g. assist-control mode) was tested in 2 pigs (Figure 4).

The sensing threshold was set at -5 cm H2O, and data was collected as the animal went from light anesthesia (spontaneous breathing) to deep anesthesia (paralysis). Tidal volume was adjusted from 200-800 mL. PEEP valve was set at 5 cm H2O. Because PEEP decay during the exhalation phase was found to cause false triggering, the algorithm was modified to  include a two respiratory cycle pause after a triggered breath before delivering a control breath. The algorithm effectively calculated PEEP for pressure threshold sensing unless there was applied suction, circuit leaks, or rapid respiratory rates exceeding 50 bpm (Fig 4 Panel B, Subject #6).

Discussion

The shortage of mechanical ventilators due to the COVID-19 pandemic has led to attempts to repurpose hand-operated AMBU bags into automated bag-compression devices (6-10, 13-18). In 2020, our group developed and tested the Automated Bag Breathing Unit (ABBU), to assist with the shortage of conventional ventilators (1-2).  ABBU is designed for training healthcare providers with limited respiratory care expertise, which may be critical in pandemics. Importantly, ABBU uses widely available resuscitation bags and circuit components and can be quickly mass-produced to potentially mitigate ventilator shortages.

ABBU is not a full-featured ICU ventilator, but a device that provides automated compression of a bag valve resuscitator.  The FDA classifies these devices as "emergency resuscitators" and they typically provide controlled ventilation with a fixed oxygen flow rate, adjustable RR and VT, manual PEEP valve, and basic alarms such as high airway pressure or power failure (21-22).

To our knowledge, ABBU is the only resuscitator providing a software-based pressure-sensing algorithm with adjustable triggering thresholds. This is an important feature of ventilation in patients with acute respiratory failure or when weaning patients from ventilation (23). Despite a growing number of approved FDA Emergency Use Authorization (EUA) resuscitators, few have published specifications or pre-clinical testing results and none have reported clinical trials in patients (24-28). Here we report ABBU is capable of providing physiological gas exchange in a short-term (6-8 hours) adult-size porcine model of normal and saline lavage lung injury. A saline lavage injury model was chosen for simplicity and reproducibility. Saline lavage causes surfactant washout with readily recruitable lung and  rapid recovery, but does not reflect the severity or heterogeneity of clinical acute respiratory distress syndrome (19, 28).  As expected, PaO2 increased with incremental PEEP and PaCO2 decreased with incremental minute ventilation by adjusting VT or RR.

It is important to understand that automated resuscitators, including ABBU, have significant limitations compared to fully functional ICU ventilators (10, 22, 29-30).  Due to the use of a fixed low flow oxygen source, delivered FiO2decreases with increases in VT or RR and may be a significant factor contributing to oxygen desaturation in patients. ABBU has no capacity for automated flow augmentation or leak compensation, such that PEEP decays during the exhalation phase. PEEP decay may be clinically significant in patients with long exhalation times, bronchopleural fistulas, or endotracheal cuff-leaks, resulting in loss of lung recruitment.

Additional limitations of ABBU and similar emergency resuscitators include a lack of measuring actual VT delivery (e.g. set bag VT plus spontaneous breath VT)  which may be significantly greater than the clinician set VT (or less in the setting of air leaks). There is no automated inspiratory or expiratory pause feature to assess inspiratory plateau pressure or auto PEEP, respectively. In contrast to ICU ventilators, there are no pressure, volume, or flow graphics to assess respiratory mechanics or patient-ventilator synchrony. Patient work of breathing and ABBU-patient synchrony assessment could not be readily reproduced in this anesthetized animal model and should be evaluated in clinical trials.

The ABBU design currently does not have an integrated battery backup for use as a transport device. However, in case of electrical or motor failure, the AMBU bag may be removed from the enclosure and used manually. This feature is an advantage over resuscitation devices that rely on a continuous source of compressed air. Durability of the ABBU device may be limited by the lifespan of the electric motor and AMBU bag (approximately 30 days and 7 days of continuous operation, respectively).

ABBU's limitations are inherent to the simplicity and low-cost design goal of achieving rapid mass production in a ventilator shortage scenario. These deficiencies are potentially addressable by close patient monitoring to include use of pulse oximetry, end-tidal CO2, FiO2 analyzer, and a VT respirometer. At the time of this writing, an application for FDA Emergency Use Authorization has been submitted and is pending review.

Conclusions

The ABBU emergency resuscitator supports short term oxygenation and ventilation in an animal model across a range of parameter settings that may potentially provide a low-cost solution to adult ventilator shortages. Clinical trials of ABBU (and similar emergency resuscitation bag devices) are necessary to establish safety and efficacy before use in patients with diverse etiologies of respiratory failure.

List Of Abbreviations

ABBU – Automated Bag Breathing Unit

BP – blood pressure

ETCO2 – end-tidal carbon dioxide

FiO2 – inspired oxygen

HR – heart rate

ICU – intensive care unit

PaCO2 – partial pressure of carbon dioxide

PaO2 – partial pressure of oxygen

PCC – Pearson correlation coefficient

PEEP – positive end-expiratory pressure

RR – respiratory rate

TI - inspiratory time

VE - minute ventilation

VT - tidal volume

Declarations

Ethics approval: Animal study was approved by the Institutional Animal Care and Use Committee at the University of Texas Health Science Center at San Antonio.

Consent for publication: Not applicable

Availability of data and materials: The datasets generated and analyzed during the current study are available from the corresponding author on reasonable request.

Competing interests: The authors declare that they have no competing interest.

Funding: The study was supported by UTHSCSA Clinical and Translational Science Award, Texas Innovation Center at the University of Texas at Austin.

Authors’ contribution:

Innovation team: A.Z., V.N.T., N.V., M.D.F., R.W., T.E.M., S.D.

Project team: A.B.G, N.K., A.G.C., S.F.J., J.W.V., T.B.P., A.B.M., R.K.L., A.Z., M.D.F., T.E.M., S.D.

Regulatory team: V.N.T., N.V.

Design and engineering: N.K., S.F.J., J.W.V., T.B.F., A.B.M. A.Z., T.E.M., S.D.

Bench testing: N.K., S.F.J., J.W.V., A.B.M., S.D.

Usability and pre-clinical studies: A.B.G., A.G.C., M.D.F., R.W., S.D.

Writing and revision of the manuscript: A.B.G., S.D., N.K., J.W.V, M.D.F, T.E.M

Study supervision: T.E.M., M.D.F., S.D.    

All authors read and approved the final manuscript.

Acknowledgements: Not applicable

References

  1. Azar II AM. Determination that a Public Health Emergency Exists.  Available from: https://www.phe.gov/emergency/news/healthactions/phe/Pages/2019-nCoV.aspx. Accessed 15 January 2021.
  2. Ranney ML, Griffeth V and Jha AK (2020) Critical Supply Shortages – The Need for Ventilators and Personal Protective Equipment during the COVID-19 Pandemic. N Engl J Med 382: e41.
  3. Rosenbaum L (2020) Facing COVID-19 in Italy – Ethics, Logistics, and Therapeutics on the Epidemic’s Front Line. N Engl J Med 382: 1873-1875.
  4. Meng L, Qiu H, Wan L, et al (2020) Intubation and Ventilation amid the COVID-19 Outbreak. Anesthesiology ZZZ:00-00.
  5. Vijayaraghavan BKT, Myatra SH, Mathew M, et al (2020) Challenges in the Delivery of Critical Care in India During the COVID-19 Pandemic.  J Intensive Care Society 0(0): 1-7.
  6. Gibney M, Panzino C. COVID-19 ventilator shortage kicks global industries into production frenzy. Available at: https://www.spglobal.com/marketintelligence/en/news-insights/latest-news-headlines/covid-19-ventilator-shortage-kicks-global-industries-into-production-frenzy-57839740. Accessed 17 January 2021.
  7. Bainbridge R. The global push to increase ventilator production. Available at: https://www.itij.com/latest/news/global-push-increase-ventilator-production. Accessed 17 January 2021.
  8. Vasich T. UCI team initiates effort to build 'bridge' ventilators. UCI News. 2020. Available at: https://news.uci.edu/2020/04/01/uci-team-initiates-effort-to-build-bridge-ventilators. Accessed 15 January 2021.
  9. Nacharaju D, Menzel W, Fontaine E, et al (2020) Three-Dimensional Printed Ventilators: A Rapid Solution to Coronavirus Disease 2019–Induced Supply-Chain Shortages. Critical Care Explorations 2(10).  
  10. Pearce JM (2020) A review of open-source ventilators for COVID-19 and future pandemics [version 2; peer review: 3 approved].  F1000Research 9:218.
  11. COVID-19 Map [Internet]. Johns Hopkins Coronavirus Resource Center. Available at: https://coronavirus.jhu.edu/map.html. Accessed 25 March 2021.
  12. Kleinpell R, Ferraro D, Maves RC, et al (2020) Coronavirus Disease 2019 Pandemic Measures: Reports From a national Survey of 9,120 ICU Clinicians. Critical Care Med 48 (10): e846-e855.
  13. Madani K. How a Small Team of Researchers, Designers, and Engineers at Fitbit Created an Emergency Ventilator to Help Save Lives. Available at: https://blog.fitbit.com/fitbit-flow. Accessed 17 January 2021.
  14. Lo J. Virgin Orbit Ventilators. Available at: https://virginorbit.com/ventilators. Accessed 17 January 17 2021.
  15. Cincinnati OH: Venti-Now. Available at: https://www.venti-now.org. Accessed 15 January 2021.
  16. Austin P51.  Available at: https://www.p51ventilators.com. Accessed 21 January 2021.
  17. Bahmer M. FDA approves enhanced version of Rice ventilator. Today's Medical Developments. Available at: https://www.todaysmedicaldevelopments.com/article/fda-approves-rice-ventilator-apollo-abvm-emergency-use. Accessed 13 January 2021.
  18. LifeMech: Machines Saving Lives. Available at: https://lifemech.org. Accessed 2 February  2021.
  19. Russ M, Kronfeldt S, Boemke W, et al (2016) Lavage-induced Surfactant Depletion in Pig as a Model of the Acute Respiratory Distress Syndrome (ARDS). J of Visualized Experiments 115: 53610.
  20. Griffiths MJD, McAuley DF, Perkins GD, et al (2019) Guidelines on the management of acute respiratory distress syndrome. BMJ Open Respiratory Research 6(1); e000420.
  21. Brown AFT and Lipman J (2008) Chapter 28- Prehospital Care. In: Mechanical Ventilation. Papadakos PJ, Lachmann B (Eds.). Elsevier, p: 307-319.
  22. Branson R, Dichter JR, Feldman H, et al (2021) The US Strategic National Stockpile Ventilators in Coronavirus Disease 2019: A Comparison of Functionality and Analysis Regarding the Emergency Purchase of 200,000.  Devices. Chest 159(2):634-652.
  23. Mora Carpio AL, Mora JI (2020) Ventilator Management.  In: StatPearls. Treasure Island (FL): StatPearls Publishing. 
  24. Dhanani J, Pang G, Pincus J, et al (2020) Increasing ventilator surge capacity in COVID-19 pandemic: design, manufacture and in vitro-in vivo testing in anaesthetized healthy pig of a rapid prototype mechanical ventilator. BMC Res Notes 13:42.
  25. Fang Z, Li AI, Wang H, et al (2020) AmbuBox: A fast deployable low-cost ventilator for COVID-19 Emergent Care. SLAS Technology 25(6):573-584.
  26. Petsiuk A, Tanikella NG, Dertinger S, et al (2020) Partially reprapable automated open source bag valve mask-based ventilator. HardwareX 8:e00131.
  27.  SVASTA, PRANA and VaU: Three Novel Ventilators from Space Community (2020) Trans Indian Natl Acad Eng 5(2):355-364.
  28. Matute-Bello G, Frevert CW, Martin TR (2008) Animal models of acute lung injury. Am J Physiol Lung Cell Mol Physiol 295: L379-L399.
  29. Badnjevic A, Gurbeta L, Jimenez ER, et al (2017) Testing of mechanical ventilators and infant incubators in healthcare institutions. Technol Heal Care 25(2):237-250.
  30. Kakkar K and Gupta B (2020) Is the use of automated artificial manual breathing unit resuscitators justified during a pandemic mechanical ventilator crisis? Acute and Critical Care 35(3): 220-222.

Tables

Table 1. Gas Exchange and hemodynamics 

Ventilator

FiO2, %

CO2, mm Hg

[kPa]

SpO2, %

pH

PaCO2,

mm Hg

[kPa]

PaO2,

mm Hg

[kPa]

SaO2, %

Veterinary ventilator

100 ± 0

41.5 ± 0.8

[5.5 ± 0.1]

97.8±0.8

[13.0 ± 0.1]

7.5 ± 0.01

45.5 ± 1.9

[6.1±0.3]

467.5 ± 25.8

[62.3±3.4]

100 ± 0

ABBU

72.7 ± 10.1 a

43.8 ± 2.6 b

[5.8 ± 0.3]

95.8 ± 0.9 b

[12.8 ± 0.1]

7.4 ± 0.1 b

51.3 ± 8.2 b

[6.8±1.1]

307.3 ± 51.9 a

[41.0±6.9]

100 ± 0

     Summary of O2 and CO2 exchange with the veterinary ventilator and ABBU. Two-tail T-test was used for comparisons. All results are mean ±SE.

     a p<0.05 vs veterinary ventilator; b p>0.05 vs veterinary ventilator