Pathogen | This study – total number of stool samples tested by TAC = 109 | Reported literature |
Median storage duration (IQR) and no. of samples: | PPA and NPA | Study Design | PPA and NPA |
ETEC | No of samples: 59 11.5 (7.2–12.0) years ≤ 3 years: 13 samples 5–13 years: 46 samples | PPA:62.7 (49.2–75.0) NPA:70.0 (55.4–82.1) Stratified by duration: ≤ 3 years: PPA: 69.2(38.6–90.9) NPA: 66.7(47.2–82.7) 5–13 years: PPA: 60.8(45.4–74.9) NPA: 75.0(50.9–91.3) | Sample size: 109 (10 ETEC positive) Comparator: Culture + PCR of 5 colonies PCR platform: TAC Storage duration: 2–4 years (8) | PPA:100.0 (69.2–100.0) NPA (LT): 55.0 (41.6–67.9) NPA (ST): 69.7 (57.1–80.4) |
Campylobacter | No of samples:21 2.6 (2.6–11.4) years ≤ 3 years: 11 5–13 years: 10 | PPA:66.6(43.0-85.4) NPA:69.3(58.6–78.7) Stratified by duration: ≤ 3 years: PPA: 63.6(30.8–89.0) NPA: 68.8(50.0-83.8) 5–13 years: PPA: 70.0(34.8–93.3) NPA: 69.6(56.0-81.2) | Sample size: 1557 (35 Campylobacter positive) Comparator: Culture PCR platform: BioFire FilmArray GI panel Storage duration: days-weeks (1) | PPA:97.1 (85.1–99.9) NPA:98.4 (97.7–99.0) |
Shigella/EIEC | No of samples:10 7.6 (6.7–10) years 5–13 years: 10 | PPA:100.0(69.2–100.0) NPA:82.8(74.0-89.6) | Sample size: 1557 (49 Shigella/EIEC positive) Comparator: Culture (PCR for EIEC) PCR platform: BioFire FilmArray GI panel Storage duration: days-weeks (1) | PPA:95.9 (86.0-99.5) NPA:98.5 (99.5–130.0) |
Norovirus | No of samples:19 2.6 (2.6–2.6) years ≤ 3 years: 19 | PPA:36.8(16.2–61.6) NPA:80.0(70.2–87.6) | Sample size: 1557 (55 Norovirus positive) Comparator: RT-PCR PCR platform: BioFire FilmArray GI panel Storage duration: days-weeks (1) | PPA:94.5 (84.9–98.9) NPA:98.8 (98.1–99.3) |
TAC on FTA card vs. standard microbiology (benchmark) This study – total number of FTA cards tested by TAC = 261 |
Pathogen | Median storage duration (IQR) and no. of samples: | PPA and NPA | Stratified by duration: |
ETEC | No of samples:70; 10.5(6.9–12.0) years ≤ 3 years: 14 5–13 years: 56 | PPA:55.7 (43.3–67.6) NPA:85.3 (80.0–90.0) | ≤ 3 years: PPA: 57.1(28.8–82.3) NPA: 86.0(75.0-93.3) 5–13 years: PPA: 55.4(41.4–68.6) NPA: 85.0(77.6–90.8) |
Campylobacter | No of samples: 59; 10.4(2.6–11.8) years ≤ 3 years: 28 5–13 years: 31 | PPA:62.7 (49.2–75.0) NPA:82.2 (76.2–87.2) | ≤ 3 years: PPA:71.4(51.3–86.8) NPA: 84.0(70.8–92.8) 5–13 years: PPA: 54.8(36.0-72.6) NPA: 81.6(74.4–87.4) |
Shigella | No of samples: 62; 6.0 (2.5–7.6) years ≤ 3 years: 6 5–13 years: 56 | PPA:88.7 (78.1–95.3) NPA:86.4(81.0–91.0) | ≤ 3 years: PPA: 16.6(1.0-64.1) NPA: 94.4(86.4–98.4) 5–13 years: PPA: 96.4(87.6–99.6) NPA: 81.8(74.0-88.2) |
Norovirus | No. of samples:27; 2.6 (2.6–2.6) years | PPA:51.8(32.0-71.3) NPA:97.8(95.0-99.3) | N/A (all samples stored ≤ 3 years) |
TAC on FTA card vs. TAC on stool (benchmark) |
Pathogen | This study – total number of paired FTA cards and stool samples tested by TAC = 109 | Reported Literature |
Median storage duration (IQR) and no. of samples: | PPA and NPA | Study Design | PPA and NPA |
ETEC | No. of paired samples:52; median duration of FTA card storage: 8.8 (2.6–11.8) years ≤ 3 years: 19 5–13 years: 33 | PPA: 75.0(63.2–86.8) NPA: 91.2(83.8–98.6) Stratify by duration: ≤ 3 years: PPA: 73.6(48.8–90.8) NPA: 91.6(73.0-98.9) 5–10 years: PPA: 75.8(57.7–88.9) NPA: 90.9(75.6–98.0) | Total sample size: 187 ETEC positive TAC on stool: 85 Storage duration: 2 years | PPA:90.6(82.2–96.0) NPA:97.6(93.3–99.5)(3) |
Campylobacter | No. of paired samples: 41; median storage duration: 7.0 (2.6–11.8) years ≤ 3 years: 17 5–10 years: 24 | PPA:58.5(42.1–73.6) NPA:88.2(78.1–94.8) Stratify by duration: ≤ 3 years: PPA: 47.0(22.9–72.2) NPA: 92.3(74.8–99.0) 5–13 years: PPA: 66.6(44.6–84.4) NPA: 85.7(71.4–94.6) | Total sample size: 187 Campylobacter positive TAC on stool: 11 Storage duration: 2 years | PPA:90.9(58.7–99.8) NPA:100.0(98.0-100.0)(3) |
Shigella | Shigella positive TAC on stool: 27; median storage duration: 7.7(6.0-11.3) years ≤ 3 years: 5 5–10 years: 22 | PPA: 66.6(46.0-83.4) NPA: 92.6(84.8–97.2) Stratify by duration: ≤ 3 years: PPA: 40.0(5.2–85.3) NPA: 94.7(82.2–99.4) 5–13 years: PPA: 72.7(49.8–89.2) NPA: 90.9(78.3–97.4) | Total sample size: 187 Shigella positive TAC on stool: 18 Storage duration: 2 years | PPA:88.9(65.2–98.6) NPA:99.4(96.8–99.9)(3) |
Norovirus | No of samples:19; median duration of FTA card storage: 2.6 (2.6–2.6) years | PPA:28.0(12.0-49.4) NPA:92.8(85.1–97.3) Stratify by duration: ≤ 3 years: PPA: 36.4(10.9–69.2) NPA: 81.2(636 − 92.8) 5–13 years: PPA: 21.4(4.6–50.8) NPA: 100.0(93.2–100.0) | Total sample size: 187 Norovirus positive TAC on stool: 24 Storage duration: 2 years | PPA:38.0(18.8–60.0) NPA:100.0(98.9–100.0)(3) |