Diabetes is a condition characterized by abnormal protein, carbohydrate, fat, and metabolism caused by either a lack of insulin production or a lack in tissue sensitivity to insulin. Fatigue, excess thirst, frequent urination, weight loss are some of the symptoms commonly seen in diabetic patients. Many citizens are suffering from Type 2 Diabetes Melitus (T2DM). It is characterized by inadequate insulin production and secretion as a result of insulin resistance. T2DM is a progressive disease. The prevalence and incidence of T2DM rise with age. DM2 accounts for 80% to 90% of all DM cases. Most people with Type 2 diabetes have intraabdominal obesity, which is linked to the prevalence of insulin resistance. This is the most prevalent type of diabetes mellitus and is strongly linked to a family history of diabetes, advanced age, obesity, and a lack of exercise. For treatment of T2DM alpha glucoside inhibitors, dipeptidyl peptidase 4 inhibitors, SGLT2 inhibitors are available.[1-3]
The market is flooded with many drugs for the management of diabetes and it is also prime requirement to have quality medicines a patient have to get relief from increase glucose level. Several methods available to estimate Vildagliptin (VIL) and Dapagliflogin (DAP) by using UV Spectroscopy [1-5], UPLC [6], RP- HPLC[7-13], UPLC and LC-MS/MS [14-16] in individual dosage form and combination with Metformin and Saxagliptin. But, no single method was available for simultaneous determination and stability indicating assay method of VIL and DAP by RP HPLC. In the current scenario, regulatory authority also recommend manufacturer to have quantitative and qualitative estimation by sophisticated instrument HPLC. Therefore, it is advantage that analyst apply developed and validated, sensitive method for simultaneous estimation of vildagliptin and dapagliflozin by liquid chromatography in pharmaceutical dosage form.
Chemicals and Materials:
There were several chemicals utilised include acetonitrile, methanol, triple distilled water, trifluoro acetic acid, potassium dihydrogen phosphate, acetic acid and sodium hydroxide. All the chemicals were of HPLC grade.
Both the drugs were procured from Acron Pharmaceuticals, Ahmedabad as a gift sample and identified by melting point test, solubility test and IR spectra.
The analysis was carried out by HPLC 2030 model, Shimadzu corporation, Japan. UV spectroscopy 1800, Shimadzu corporation, Japan, was utilised for wavelength determination through UV spectra.
Chromatographic Condition:
The estimation was performed by using following chromatographic conditions. Separation was achieved by C18 columns with 250 mm x 4.6 mm, 0.5 micron particle size. The column oven temperature was set at 25 °C. Mobile phase was selected after many trials as Methanol : Water Containing 0.01% TFA in the ration of 95:05 % v/v (pH: 2.78 ± 0.02).The injection volume was kept at 10 µL and flow rate was adjusted at 0.8 ml/min. Both the drugs were estimated at 210 nm wavelength.
Preparation of Mobile Phase
The mobile phase was prepared by mixing methanol and 0.01% trifluoro acetic acid (pH-2.78) in ratio 95:05 %v/v. The prepared mobile phase was sonicated and filtered through 0.45μm membrane filter.
Preparation of Standard Stock Solution
Weighed accurately 5 mg of dapagliflozin and vildagliptin drugs were transferred into a 10ml volumetric flask and diluted up the mark with methanol to produce 500 µg/ml concentration.
Preparation of Working Stock Solution
The aliquot of 0.5 ml from standard stock solution was transferred in to 10 ml volumetric flask and make up the volume up to the mark which produced 100 µg/ml concentration for each drug.
Experimental conditions:
Preparation of Calibration Curve (Linarity)
The working standard solution (1,2,3,4,5,6 ml) were transferred into 10 mal volumetric flask to make 10-60µg/ml for DAPA and VIL. These prepared solutions were diluted withmobile phase and were injected to system.
System suitability
The system suitability studies conceded as per ICH guideline. The parameters like tailing factor, number of theoretical plates, and resolution were calculated by injecting six replicates of standard concentration of 50 µg/ml into HPLC system.
Specificity
Specificity was examined by the overlay spectra of working standard and blank i.e., diluent, to exhibit for interference or hinderance of diluents and any impurities of synthetic process or formed during shelf life.
Limit of Detection and Limit of Quantitation
The linearity study was carried out in triplicate. The LOD and LOQ were calculated based up on following equation. The LOD and LOQ were calculated using the average of slope and standard deviation of intercept.
LOD = 3.3 × (σ / S)
LOQ = 10 × (σ / S)
Where, σ = The standard deviation of Y-intercept of 3 calibration curves
S = The mean slope of the 3 calibration curves
Accuracy
Accuracy of the method was confirmed by recovery studies. It was carried out by standard addition method. The difference between theoretical added amount and practically achieved amount was calculated. It was performed by spiking API to Tablet sample at specified levels. Accuracy was determined at three different levels 50%, 100% and 150% of the target concentration in triplicate.
Precision
Intraday Precision was done by performing the analysis at different time (10am, 1pm and 4pm) on the same day and the concentration and % RSD was calculated.
Interday Precision was examined by analyzing the standard of same concentration on different days (day1, 3 and 7) and %RSD was calculated.
Robustness
Robustness was performed by deliberately changing specific parameters like mobile phase ratio, wavelength, flow rate and % RSD was calculated.
Assay
Accurately weighed 20 tablets crushed and powder of the formulation was transfer to volumetric flask to have test solutions of 100 µg/ml (VIL) and 10 µg/ml (DAPA). Solutions prepared were injected after stabilization of baseline into HPLC system and chromatograms were recorded. The concentration of each test solution was determined by using slope and intercept values from the calibration graph.
Force Degradation Studies:
Hydrolysis Degradation
Acid and Alkaline Hydrolysis was carried out with the working standard concentration (50µg/mL) by exposing the drugs with 0.1N HCl and 0.1 N NaOH solution for the interval of 30mins, 1hour, 2hour and 24hour. Degradation pattern was calculated by comparing peak area with the standard chromatogram area of both the drugs.
Oxidative Degradation
Hydrogen peroxide as an oxidising agent was used to perform the degradation study.Drugs were solubilised into freshly prepared 3% Hydrogen peroxide and kept for different intervals Degradation achieved was calculated.
Thermal Degradation
Solid API was exposed at 45°C in hot air oven. Further analysis was carried out and % degradation was checked and calculated.
Photolytic Degradation
Solid API was exposed in UV light up to 4 hours, solutions were prepared and analysed. % degradation was noted and calculated.